Specifications for Remote Scale
The scale has the ability of driving up to 4 load cells (350
Ω
) in a remote scale
(platform). Set for the capacity range 0.5kg to 5,000kg (1 lb to 10,000 lb).
The scale has the ability of driving up to 4 load cells (350
Ω
) in a remote scale
(platform).
Capacity range
for the remote scale is 0.5kg to 5,000kg (1 lb to 10,000 lb).
Minimum output
at zero point is 1mV.
Maximum output
at full load is 14mV.
Excitation Voltage of the scale is 5V.
Cable length
should be kept under 5m (16.5 ft.) for higher accuracy.
Pin connection
(JM:NJC-207-PF):
The Load Cell and Input Sensitivity
The relationship between load cell and input sensitivity (
X
) for the scale is follows:
Example
Load Cell Capacity
100kg
“
A
”
Rated
Output
3mV/V
“
B
”
Min., Division of Display
0.01kg
“
D
”
5,000 x x
B
D
A
“
X
”= µV
When a single Load Cell is used,
the following formula should apply:
System design will be satisfactory if “
X
” is
greater than 0.5µV. In the example above
“
X
” = 1.5µV.
Capacity and Resolution
The resolution of the remote scale is automatically determined during the calibration
procedure. The following is to enable you to calculate the resolution for a given capacity.
1. Decide the capacity value and assign it to “Ws”. Maximum 5 digits.
2. Get the maximum count “Ns” for the capacity. Ignore the decimal point and add “0”
to Ws until it is 5 digits.
3. Calculate: d’=Ns/10,000.
4. Decide the minimum division “d”.
d’=1
Æ
d=1
2
≥
d’>1
Æ
d=2
5
≥
d’>2
Æ
d=5
d’>5
Æ
d=10 This should be changed to d=1, dividing Ns by 10.
5. Now “Ns x d” and the resolution 1/No = d/Ns can be determined.