![A&D EK-15KL Скачать руководство пользователя страница 24](http://html.mh-extra.com/html/aandd/ek-15kl/ek-15kl_instruction-manual_446383024.webp)
22
8-3. Setting example
To set the auto power-off function to “Enabled” and the ACAI function to “Disabled.”
1. With the power turned OFF, press and hold the
RE-ZERO key and press the ON/OFF key to turn
the power ON and enter the function setting mode.
func
appears.
2. Press the PRINT key to display
poff 0
.
3. Press the RE-ZERO key to display
poff 1
.
4. Press the MODE key several times to display
aCai 1
.
5. Press the RE-ZERO key to display
aCai 0
.
6. Press the PRINT key to store the parameters.
Cp Hi
appears after
end
.
7. Press the ON/OFF key to turn the power OFF or
press the RE-ZERO key.
Switches the parameter
To store
Switches the parameter
Press several times