16
7-2. Gravity acceleration correction
When the balance is first used or has been moved to a different place, it should be
calibrated using a calibration weight.
But if the calibration weight cannot be prepared, the gravity acceleration correction will
compensate the balance. Change the gravity acceleration value of the balance to the
value of the area where the balance will be used. See the gravity acceleration map
appended to the end of this manual.
Note
Gravity acceleration correction is not required when the balance is
calibrated using a calibration weight at the place where the balance
is to be used.
1. Press and hold the calibration (CAL) switch until
Cal
appears, and release the switch.
2. The balance displays
Cal 0
.
3. Press the RE-ZERO key.
The display shows the gravity acceleration value
stored in the balance.
Use the following keys to change the value.
SAMPLE key To select the digit blinking to
change.
RE-ZERO key To set the value of the digit
selected.
PRINT key
To store the value and return to
step 2.
MODE key
To cancel the value and return to
step 2.
4. After setting the value, press the PRINT key.
Cal 0
is displayed again.
5. If it is necessary to calibrate the balance using a
calibration weight, go to step 4 of 7-1.
To finish the setting, press MODE key.
6.
end
appears and the balance returns to the
weighing mode.
Cal
Cal 0
9.7980
Cal 0
Returns to
the weighing mode
To store
Release
the CAL switch.
Set the value using
the relevant keys.
Press and hold
the CAL switch.
Содержание EK-120i
Страница 1: ...x 1WMPD4000465B...
Страница 42: ...40 World map...