 16

7-2. Gravity acceleration correction   

 

When the balance is first used or has been moved to a different place, it should be 
calibrated using a calibration weight. 
But if the calibration weight cannot be prepared, the gravity acceleration correction will 
compensate the balance. Change the gravity acceleration value of the balance to the 
value of the area where the balance will be used. See the gravity acceleration map 
appended to the end of this manual. 
 

‰

 Note 

  Gravity acceleration correction is not required when the balance is 

calibrated using a calibration weight at the place where the balance 
is to be used.   

 
 
1. Press and hold the calibration (CAL) switch until   

Cal

         

appears, and release the switch. 

 
2. The balance displays 

 Cal 0

  . 

 
 
3. Press the    RE-ZERO    key. 

The display shows the gravity acceleration value 
stored in the balance. 
Use the following keys to change the value. 

 

 SAMPLE  key  To select the digit blinking to 

change. 

 

 RE-ZERO  key  To set the value of the digit 

selected. 

 

 PRINT  key 

To store the value and return to 
step 2. 

 

 MODE   key 

To cancel the value and return to 
step 2. 

 
4. After setting the value, press the    PRINT    key. 

 Cal 0  

  is displayed again. 

 
5. If it is necessary to calibrate the balance using a 

calibration weight, go to step 4 of 7-1. 
To finish the setting, press    MODE    key. 

 
6.  

end   

  appears and the balance returns to the 

weighing mode. 

 
 
 
 
 
 
 

 

  Cal 

  Cal 0 

  9.7980 

 Cal 0 

Returns to   

the weighing mode

To store

Release 

the CAL switch.

Set the value using

the relevant keys. 

Press and hold

the CAL switch.

Содержание EK-120i

Страница 1: ...x 1WMPD4000465B...

Страница 2: ...nt has been tested and has been found to comply with the limits of Class A computing device pursuant to Subpart J of Part 15 of FCC rules These rules are designed to provide reasonable protection agai...

Страница 3: ...unting mode pcs 10 5 8 Percent mode 11 6 COMPARATOR 12 6 1 Setting example 12 7 CALIBRATION 14 7 1 Calibration using a weight 14 7 2 Gravity acceleration correction 16 8 FUNCTIONS 17 8 1 Key operation...

Страница 4: ...sembly 31 11 3 OP 09 Rechargeable battery pack Ni MH 32 11 4 OP 12 Carrying case 32 12 MAINTENANCE 33 12 1 Notes on maintenance 33 12 2 Error codes 33 13 SPECIFICATIONS 35 13 1 EK i series 35 13 2 EW...

Страница 5: ...use in a dimly lighted place The standard RS 232C serial interface can be connected to a printer or personal computer Using the serial interface Good Laboratory Practice GLP data can be output The opt...

Страница 6: ...s the weighing range when rng 0 is selected pcs mode Enters the sample unit weight storing mode mode Enters the 100 weight storing mode Outputs the weight data to a printer or computer or enters a uni...

Страница 7: ...e where combustible or corrosive gases may exist Allow the balance to reach equilibrium with the ambient temperature before use Switch the power ON at least half an hour before use so that the balance...

Страница 8: ...re is a load beyond this range the balance will be tared to zero and the NET mark and the ZERO mark turn on 2 Pressing the ON OFF key again and the power will be switched OFF Auto power off function I...

Страница 9: ...are not available for higher or lower capacity models For details see 13 SPECIFICATIONS Note It is possible to store only the units that will be actually used from the units available It is also poss...

Страница 10: ...lance if the weight is within 2 of the weighing capacity around the power on zero point The ZERO mark turns on When the weight exceeds 2 of the weighing capacity it will be subtracted to zero as a tar...

Страница 11: ...be tared to zero and the weighing range changes to the lowest range rng 0 Manual range Press the SAMPLE key when the display shows a weight value neither counting nor display The weighing range chang...

Страница 12: ...ple The balance is set to count objects with this unit weight When a unit weight is too light to store the display shows lo for a moment and returns to the former display When the sample weight is lig...

Страница 13: ...amples as displayed 5 8 Percent mode Displays the weighing value in percentage compared to the reference 100 weight Selecting the percent mode 1 Press the MODE key to select percent Storing the refere...

Страница 14: ...our display divisions The upper limit and lower limit numerical values are common to each of the weighing counting and percent mode An example for the EK 1200i 2000i 3000i is as follows Upper limit va...

Страница 15: ...p 7 With Cp lo displayed press the PRINT key Enter the lower limit value using the following keys SAMPLE key To select the digit blinking to change RE ZERO key To set the value of the digit selected H...

Страница 16: ...s Cal 0 To change the calibration weight value proceed to step 4 To use the calibration weight value in the balance memory proceed to step 5 4 Press the SAMPLE key The display shows the calibration we...

Страница 17: ...Do not touch the pan during weighing 7 end appears Remove the weight from the pan and press the CAL switch or MODE key to return to the weighing mode Note The value set in step 4 is stored in memory...

Страница 18: ...ion CAL switch until Cal appears and release the switch 2 The balance displays Cal 0 3 Press the RE ZERO key The display shows the gravity acceleration value stored in the balance Use the following ke...

Страница 19: ...r selection See Function list Cancels the operation and turns off the balance Item key Selects a class and an item In the weighing mode press and hold the key to enter the function setting mode Enter...

Страница 20: ...y to display poff 1 4 Press the SAMPLE key several times to display aCai 1 5 Press the RE ZERO key to select aCai 0 6 Press the PRINT key to store the parameters Cp Hi appears after end 7 Press the MO...

Страница 21: ...key 4 Press the SAMPLE key to select a weighing unit 5 Press the RE ZERO key to store the weighing unit 6 Repeat steps 4 and 5 to store all weighing units to be used 7 Press the PRINT key id appears a...

Страница 22: ...data 3 Compares plus data 4d 4 Compare stable plus data 4d 5 Compares data 4d or 4d Cp Comparator mode 6 Compares stable data 4d or 4d Conditions to compare d the minimum display division 0 Buzzer do...

Страница 23: ...nds 2 Turns off after 10 seconds 3 Turns off after 30 seconds 4 Turns off after 60 seconds func ltUp LCD Backlight control 5 Always on To control how the LCD backlight turns off Weight change or key o...

Страница 24: ...ight type to connect with a computer Windows Communication Tools Software WinCT to communicate with a computer is provided as freeware Visit the A D website to download WinCT 9 1 Interface specificati...

Страница 25: ...c t Weight data carat mom Weight data momme d w t Weight data penny weight GN Weight data grain t l Weight data tael The terminator is always CRLF Example of output data Weight data gram S T 0 0 1 2...

Страница 26: ...print mode B Set the function prt 3 The balance transmits the weight data when the display is stable the stability mark is on and the data is greater than 4d of the lowest range for EW i series or le...

Страница 27: ...er using the following keys SAMPLE key To select the digit blinking to change RE ZERO key To set the character of the digit selected See the table below for the display character set PRINT key To stor...

Страница 28: ...he calibration has been completed 3 glp is displayed and calibration report is output 4 end appears again Remove the weight and press the MODE key to return to the weighing mode AD 8121 format info 1...

Страница 29: ...tion test weight according to the procedure step 4 of 7 1 Calibration using a weight 4 With nothing on the pan press the PRINT key The zero point is measured and the weighed value is displayed for a f...

Страница 30: ...the weight data by pressing the PRINT key or selecting the auto print mode A D MODEL EK 1200i S N Q1234567 ID ABCDEF DATE 2011 11 14 TIME 03 15 40 CAL TEST EXT ACTUAL 0 0 g 1000 1 g TARGET 1000 0 g SI...

Страница 31: ...011 11 17 START TIME 01 23 45 ST 000123 4 g ST 000234 5 g ST 000345 6 g ST 000456 7 g END TIME 01 25 43 SIGNATURE A D CRLF MODEL EK 1200i CRLF S N Q1234567 CRLF ID ABCDEF CRLF DATE CRLF CRLF START CRL...

Страница 32: ...to sound a buzzer according to the comparison result See the function bep to set which result will make a buzzer sound The comparator function on off the comparison mode and comparator buzzer output...

Страница 33: ...bottom of the balance large objects that are difficult to load on the weighing pan can be weighed in suspension and the specific gravity of objects may be measured OP 07 is available for use with the...

Страница 34: ...ly stop using it and recharge the battery pack or use the AC adapter Charge the battery pack at a temperature between 0 C 32 F and 40 C 104 F Do not charge too long Overcharging will reduce the life o...

Страница 35: ...he weight sensor receives strong force upward Check if there is anything sandwiched around the weighing pan There is a possibility that the weight sensor itself may have a failure Unit 100 weight erro...

Страница 36: ...rning to indicate that the output voltage of the AC adapter is too high Check if the AC adapter is correct Stability error Warning to indicate that the weight value is not stable and the balance canno...

Страница 37: ...kg 1 1 kg 1 3 kg Calibration weight factory setting 120 g 200 g 300 g 600 g MODEL EK 1200i EK 2000i EK 3000i EK 6000i EK 12Ki Weight capacity 1200 g 2000 g 3000 g 6000 g 12 kg Min display d 0 1 g 0 1...

Страница 38: ...ht factory setting 400 g 600 g 4000 g 6000 g In case of Umin 0 factory setting 13 2 EW i series MODEL EW 150i EW 1500i EW 12Ki Weight capacity 30 g 60 g 150 g 300 g 600 g 1500 g 3 kg 6 kg 12 kg Min di...

Страница 39: ...228 lb Min display 0 0001 0 0001 0 001 0 001 Capacity 12 860 19 290 128 60 192 90 ozt Min display 0 001 0 001 0 01 0 01 Capacity 2000 0 3000 0 20000 30000 ct Min display 0 1 0 1 1 1 Capacity 106 665 1...

Страница 40: ...38 13 4 Dimensions EK 120i EK 200i EK 300i EK 600i EK 1200i EK 2000i EK 3000i EK 6000i EK 12Ki EK 410i EK 610i EK 4100i EK 6100i EW 150i EW 1500i EW 12Ki...

Страница 41: ...s2 Ottawa Paris 9 807 m s2 9 810 m s2 Copenhagen 9 816 m s2 Rio de Janeiro 9 788 m s2 Cyprus 9 797 m s2 Rome 9 803 m s2 Frankfurt 9 811 m s2 San Francisco 9 800 m s2 Glasgow Havana 9 816 m s2 9 788 m...

Страница 42: ...40 World map...

Страница 43: ...NTS LIMITED German Sales Office Gro e Stra e 13 b 22926 Ahrensburg Deutschland Telefon 49 0 4102 459230 Telefax 49 0 4102 459231 A D Australasia Pty Ltd 32 Dew Street Thebarton South Australia 5031 AU...

Отзывы: