43
Minimum weight measurement result
The [Minimum weight measurement result] screen is displayed when the
1
Measurement method
2
Measurement tolerance
3
Repeatability (standard deviation)
4
Minimum weight
5
PRINT button
6
REGISTER button
Name
Description
1
Measurement method Displays the method used for the minimum weight measurement.
2
Measurement tolerance Displays the minimum weight measurement tolerance.
3
Repeatability (standard
deviation)
Displays the standard deviation for this measurement.
4
Minimum weight
Displays the minimum weight.
5
PRINT button
Outputs the minimum weight measurement result to the device connected to
the balance.
6
Register button
Registers the minimum weight.
*
If the repeatability (standard deviation) is 0.41d or less, the minimum weight is 2000 times the 0.41d in
compliance with USP.
"d" is a unit of readability.
Содержание BA-125DTE
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315 ...
Страница 74: ...74 Output example for daily check output results Daily check output results Daily check PDF results ...
Страница 84: ...84 Output example for periodic check output results Periodic check output results ...
Страница 85: ...85 Periodic check results PDF ...
Страница 166: ...166 Unit mm Display unit 138 182 73 Ionizer AX ION 25 162 68 129 ...
Страница 170: ...MEMO ...
Страница 171: ......