7. TROUBLESHOOTING
Symptom
Description
appears.
Communication settings of the AD-8923-CC do not match with those
of the weighing instrument.
Check the settings such as baud rate and parity and correct them as
necessary.
For details, refer to “
3.1. Setting the weighing instrument and the AD-
-
(Bar display)
remains and the weight
value is not displayed.
Is the data output mode of the weighing instrument set to “stream mode”?
In a mode other than “stream mode”, the weight values are displayed
only when they are transmitted.
Check if the communication settings are correct.
Check if the cables are the correct type and are not damaged.
The display flickers.
Electrical noise may cause this symptom.
Ground the FG terminal located on the rear of the AD-8923-CC.