11
Sensitivity adjustment procedure
Performs sensitivity adjustment with the AD-4212C using the external weight.
Operation
1. Warm up the AD-4212C for 30 minutes or more
with nothing on the pan.
2. Press the CAL key.
is displayed.
If you want to cancel sensitivity adjustment, press the CAL key.
The display will return to the weighing mode.
If you want to change the weight value, press the SAMPLE key.
Press the RE-ZERO key to select the weight value, and press
the PRINT key to store it.
is displayed.
3. Confirm that there is nothing on the pan and press
the PRINT key. The AD-4212C measures the zero point.
Do not allow vibration or drafts to affect the AD-4212C.
The weight value is displayed.
4. Place a weight, of the weight value displayed, on the pan
and press the PRINT key.
The AD-4212C measures the weight.
Do not allow vibration or drafts to affect the AD-4212C.
5.
is displayed. Remove the weight from the pan.
6. The display will automatically return to the weighing mode.
7. Place the weight on the pan and confirm that adjustment
was performed correctly. If not, check the ambient conditions
such as drafts or vibration, and repeat steps 2 through 7.
Press
Weight
Press
Press
STABLE FAST MID.
RESPONSE
SLOW
CAL
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW