8. Function Settings 

In the function setting mode, settings of the AD-8527 

performance and communication can be changed. 

Each function has parameters. By changing the parameter, 

specify the performance of the function. 

The set parameters are maintained in memory even if the 

AD-8527 is disconnected from the weighing instrument or 

personal computer. 

 

8.1. Setting Procedure 

1. 

At the standby display, press and 
hold the SET key for 2 seconds or 
more.

 

 

2.  The function display appears. (Refer to “8.2. Function List”.) 

Press the (+) key to select a 
function, and press the SET key to 
display the parameters. 

 

  When the CANCEL key is pressed, 

or after 10 seconds of non operation, 
the AD-8527 returns to the standby 
display. 

 
3. The parameter of the selected 

function appears. The current 
setting has the   mark at the 
lower right. 
The flashing parameter can be 
changed. 

 
 
 

4.  Using the following keys to change the parameter. 

  (+) key:   

Display the parameters one by one 

CANCEL key:   Go to the next function without changes. 
SET key:  

Confirm the setting and go to the next 
function. 

 
 

After changes, a buzzer sounds, 
the parameter and the    mark 
appear for 1 second, and then the 
next function appears.

 

 
 

5.  Press the CANCEL key to return to 

the standby display. 

 

 

 

 

 

 

 

 8.2. Function List 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

(Note 1) Set this function according to the setting of the personal 

computer used. 

  *: Factory setting

 

 

About the number of data on a line, interval mode, data to 

transmit, date and time data, refer to the following sections. 

 

8.3. Data to Transmit 

Adding the date and time 

The date and time when the AD-8527 receives the data can be 

transmitted to a personal computer. The date, time and weight 

are separated by tabs. 

 

e.g.: 6/10/2009<Tab>10:41:24<Tab>+00123.45<Enter> 

 

Example below is when the data is transmitted to Excel 

 

 

 

 

 

 

 
 
 
 

 

 

Selecting data to transmit 

Two ways of transmitting data are available; weight values only 

or all data. Transmitting weight values only is convenient when 

performing statistical calculations using Excel. When performing 

GLP output, transmitting all data is required. 

For example, the received data is “ST,+00123.45  g”. When 

transmitting weight values only is selected, “+00123.45” will be 

transmitted. When transmitting all data is selected, 

“ST,+00123.45  g” will be transmitted.

 

Number of data on a line 

 This function is useful for storing data with different contents or 

items at the same time, for example, ID and number of samples 

in a counting scale. When the number is set, <CR>s of up to the 

setting are converted into commas. Once the number of <CR>s 

is reached the setting, a new line starts. 

For example, the ID, number of samples and weight are sent 

individually. Using this function, these three data items can be 

placed on one line as shown in (B).

 

 

   

Example data received to transmit to Excel 

       

ID,000123<CR><LF> 

       

QT,+00000056 PC<CR><LF>   

       

ST,+0003.102   g<CR><LF> 

 

(A)  When the number of data on a line is 1, the ID, number 

of samples and weight are input on separate lines. 

 

 

 

 

 

 

 

 

 

(B)  When the number of data on a line is 3, the ID, number 

of samples and weight are input on one line. 

 

 

 

 

 

 

 

 

 

 

 

 

 
 

9. Interval Mode 

  In the interval mode, the weighing data is transmitted at a certain interval. 

  Available intervals are: 

1, 2, 5, 10, 15, 20, 30 seconds / 1, 2, 5, 10, 15, 30 minutes / 1 hour 

 

1.  Connect the AD-8527 to a personal computer using USB. 

2. Connect the AD-8527 to a weighing instrument using the 

RS-232C cable. 

3. Select the interval mode (

int

) for the 

data transmission mode (

out

) of the 

function setting and set the interval 
time (120 seconds in the example). 
 

  Press the CANCEL key to return to the 

standby display.

 

 In the interval mode, the   mark turns 
on at the lower right. 

4. Set the data output mode of the weighing instrument to the 

stream mode. 

 

For details on the setting procedure, refer to the instruction 
manual of the weighing instrument. 

5.  Press and hold the (+) key for 2 seconds or 

more. The AD-8527 starts to transmit the data 
at a set interval. “

out

” is displayed and the   

mark flashes while the data is transmitted. 

6.  To stop transmitting the data, press and hold 

the (+) key. 
The    mark stops flashing and turns on. 

 

 

10. Errors 

10.1. Error Display 

Error Description 

Remedy 

 

The supplied voltage 
is low.

 

Confirm that the USB is 
connected correctly. 

 

The baud rate and 
parity of the received 
data is different from 
the set values.

 

Set the same values 
for both the AD-8527 
and the weighing 
instrument.

 

 

Internal clock error 
such as low backup 
battery

 

Ask for repair.

 

 

10.2. If the Data Fails to be Input into Excel Correctly 

If data is not input into Excel correctly when using 
Windows XP, do as follows. 

1. Open “Control Panel”, and then “Regional and 

Language Options”. 

 

 

 

 

 

 

 
 

2. Select the Languages tab 

and click on the Details 
button. 

 

 

 

3. Select the Advanced tab 

and check the Turn off 
advanced text servises 
box. 

 
 
 
 

11. Specifications 

Devices   

Balances and scales with an 

RS-232C connector manufactured 

by A&D 

Applicable OS 

Windows 2000 / XP / Vista 

Clock error 

±

1 minute / month 

Dedicated power supply  Not required 

Power is supplied from a 

weighing instrument or a personal 

computer. 

Operating temperature 

5°C to 40°C 

Operating humidity 

85%RH or less, no condensation 

Dimensions 55x103x16.5 

mm 

Mass Approximately 

60 

g

 

 

Windows Explorer, Excel, Word and Notepad are registered 
trademarks of Microsoft Corporation.

 

Press and hold 

Press 

Current setting 

 

to be connected 

OK

OK

(Note1)

Function

Description

Data bit, parity

Parameter

 

7bit-Even*, 7bit-Odd, 8bit-None

Baud rate

Parameter

 

600, 1200, 2400*, 4800, 9600, 19200 bps

Number of data on a line

Parameter

 

1*,2, 3, 4, 5, 6, 7

Data transmission mode

Parameter

 

Standard mode (All data)*, Interval mode

Buzzer

Parameter

 

Sounds the buzzer*, Does not sound

When the interval mode is selected

Interval time

Parameter

 

1*,2,5,10,15,20,30,60,120,300,600,900,1200,1800,3600

Data to transmit

Parameter

  

Weight values only*,  All data (with a header and unit)

Date and time data

Parameter   Adds the date and time*, Does not add

Date format

Parameter

 

YMD*, MDY, DMY

DATE

DATE

DATE

DATE

DATE

SEC

SEC

SEC

DATE

SEC

DATE

SEC

SEC

OK

OK

OK

 

ID

Number of  Weight

  

samples 

 

From the front 