3
3. PREPARATION
3-1. The DIP Switch Functions
The DIP switch assembly for the settings is located on the base of the printer.
The switch assembly is protected by a white plastic cover. Open the cover with a
screwdriver.
DIP switches (Factory setting)
Before changing the DIP switch settings, be sure to turn the power off.
DIP Switch Settings
Baud rate between the printer and the scale
Switch No.
Speed
1
Comments
2400 bps
OFF
Factory setting
9600 bps
ON
Label length
Switch No.
Length
2
3
Comments
34 mm
OFF
OFF
Factory setting
40 mm
ON
OFF
46 mm
OFF
ON
UFC setting
ON
ON
Barcode data selection (Type of barcode is Code 39 only.)
Switch No.
Barcode data
4
5
6
Comments
No barcode
OFF
OFF
OFF
Factory setting
ID code (Header is ‘ID’)
ON
OFF
OFF
Weight (Header is ‘WT’)
OFF
ON
OFF
Quantity (Header is ‘QT’)
ON
ON
OFF
Unit weight (Header is ‘UW’)
OFF
OFF
ON
Tare (Header is ‘TR’)
ON
OFF
ON
The barcode is printed in
the format that the FC-
i
/ S
i
scale can accept.
Not in use
OFF
ON
ON
Not in use
ON
ON
ON
Data format
Switch No.
Data bit / Parity
10
Comments
7 bits / Even parity
OFF
8 bits / Non parity
ON
Factory setting
Switch No.
Keep these switches as shown.
7
8
9
Comments
Internal use
OFF
OFF
ON
Factory setting
Please note that DIP switch No.10
is labeled as 0.