6 Press the ZERO switch to adjust the temperature
setting.
When the Display direction switch is set to
↑
, each
time the ZERO switch is pressed, the number
increases by 1.
When the Display direction switch is set to
↓
, each
time the ZERO switch is pressed, the number
decreases by 1.
7 Press the ON/OFF switch.
The scale stores the new temperature setting at
calibration and indicates [C-6].
8 Press the ON/OFF switch.
The display indicates [9798], which means the
gravitational acceleration setting of the calibration site
is 9.798 m/s
2
. Here, adjust the value as necessary.
9 Press the ZERO switch to adjust the gravitational
acceleration.
When the Display direction switch is set to , each
time the ZERO switch is pressed, the number
increases by 1.
When the Display direction switch is set to , each
time the ZERO switch is pressed, the number
decreases by 1.
Note
The gravity values for various locations are listed on page 19.
10 Press the ON/OFF switch.
The scale stores the new gravitational acceleration and
indicates [C-7].
11 Press the ZERO switch several times until the display
indicates [C-10].
Each time the ZERO switch is pressed, the number
increases by 1; [C-8] [C-9] [C-10] ….
12 Press the ON/OFF switch.
The display indicates the reference count for current
output of Thermo-sensitive resistor. Wait for the Stable
mark to turn on.
13 When Stable mark is displayed, press the ON/OFF
switch.
The scale stores the reference count for current output
The display value increments
Display direction switch
The display value decrements
The display value increments
Display direction switch
The display value decrements
Stable mark