2. Specifications
AD-4329A-07 Analog
output
Current output
4mA to 20mA
Outputs 4mA or less when the display is turned off or the indicator
is in a mode other than the weighing data displaying mode, such
as calibration.
Outputs 20mA or more at overload display.
Non-linearity ±0.1%FS
max.
Load resistance
0
to 520
Zero temperature coefficient ±0.02 %FS/°C max.
Span temperature coefficient ±0.02 %FS/°C max.
Resolution Approx.
1/3000
Output adjustment
SPAN control and ZERO control located on the option board
Terminal
M3 compression terminal
Note: Analog output can not be used at the same time as RS-232C output or other options
except OP-02.
Pin Connection
Related settings
Settings related to analog output are performed by the function settings,
f-60
,
f-61
, and
f-62
.
For details on the setting procedure, refer to the AD-4329A instruction manual.
Item Parameter
Description
f-60
0
Displayed value
f-60
1
Gross data
Output data
f-60
2
Net data
Value at 4mA output
f-61
-999999 to 999999 (Factory setting: 0)
The decimal point position depends on the
Cf-00
setting.
Value at 20mA output
f-62
-999999 to 999999 (Factory setting: 10000)
The decimal point position depends on the
Cf-00
setting.
Note: To confirm the F-61 and F-62 settings, press the key and preceed to the next item.
Do not press the MODE key.