![Aalborg XFM Series Скачать руководство пользователя страница 7](http://html1.mh-extra.com/html/aalborg/xfm-series/xfm-series_operating-manual_3238069007.webp)
2.2.3 Communication Parameters and Connections
The digital interface operates via RS485 (optional RS-232 or Profibus DP avail-
able) and provides access to applicable internal data including: flow, CPU tem-
perature reading, auto zero, totalizer and alarm settings, gas table, conversion
factors and engineering units selection, dynamic response compensation and lin-
earization table adjustment.
Communication Settings for RS-485/RS-232 communication interface:
Baud rate:
......................
9600 baud
Stop bit:
......................
1
Data bits:
......................
8
Parity:
......................
None
Flow Control: ......................
None
RS-485 communication interface connection:
The RS485 converter/adapter must be configured for: multidrop, 2 wire, half
duplex mode. The transmitter circuit must be enabled by TD or RTS (depending
on which is available on the converter/adapter). Settings for the receiver circuit
should follow the selection made for the transmitter circuit in order to eliminate
echo.
RS-485 T(-) or R(-)
...................... pin 8 of the 15 pin "D" connector (TX-)
RS-485 T(+) or R(+)
...................... pin 15 of the 15 pin "D" connector (RX+)
RS-485 GND (if available) ...................... pin 9 of the 15 pin "D" connector (GND)
RS-232 communication interface connection:
Crossover connection has to be established:
RS-232 RX (pin 2 on the DB9 connector) ..... pin 8 of the 15 pin "D" connector (TX)
RS-232 TX (pin 3 on the DB9 connector) ..... pin 15 of the 15 pin "D" connector (RX)
RS-232 GND (pin 5 on the DB9 connector) ..... pin 9 of the 15 pin "D" connector (GND)
4
Содержание XFM Series
Страница 51: ...48 APPENDIX IV COMPONENT DIAGRAM TOP COMPONENT SIDE Aug 09 2007...
Страница 52: ...49 BOTTOM COMPONENT SIDE Aug 09 2007...
Страница 53: ...APPENDIX V DIMENSIONAL DRAWINGS XFM WITHOUT READOUT 50...
Страница 54: ...XFM WITH PROFIBUS CAPABILITY 51...
Страница 55: ...XFM WITH READOUT OPTION 52...