4
2.2.3 Communication Parameters and Connections
The digital interface operates via RS485 (optional RS-232 is available) and pro-
vides access to applicable internal data including: flow, temperature, pressure
reading, auto zero, totalizer and alarm settings, gas table, conversion factors and
engineering units selection, dynamic response compensation and linearization
table adjustment.
Communication Settings:
Baud rate:
-------- 9600 baud
Stop bit:
-------- 1
Data bits:
-------- 8
Parity:
-------- None
Flow Control:
-------- None
RS-485 communication interface connection:
The RS485 converter/adapter has to be configured for: multidrop, 2 wire, half
duplex mode. The transmitter circuit has to be enabled by TD or RTS (depending
on which is available on the converter/adapter). Settings for the receiver circuit
usually should follow the selection made for the transmitter circuit in order to
eliminate echo.
RS-485 T(-) or R(-)
-------- pin 11 of the 25 pin "D" connector (-)
RS-485 T(+) or R(+)
-------- pin 24 of the 25 pin "D" connector (+)
RS-485 GND (if available)
-------- pin 20 of the 25 pin "D" connector (GND)
RS-232 communication interface connection:
Crossover connection has to be established:
RS-232 RX
(pin 2 on the DB9 connector) -------- pin 11 of the 25 pin "D" connector (TX)
RS-232 TX
(pin 3 on the DB9 connector) -------- pin 24 of the 25 pin "D" connector (RX)
RS-232 GND
(pin 5 on the DB9 connector) -------- pin 20 of the 25 pin "D" connector (GND)
Содержание DFM 26
Страница 44: ...40...
Страница 45: ...41...
Страница 46: ...42...
Страница 47: ...43...
Страница 48: ...44...
Страница 62: ...58 APPENDIX IV COMPONENT DIAGRAM DFM Analog PC Board TOP...
Страница 63: ...59 COMPONENT DIAGRAM DFM Analog PC Board BOTTOM...
Страница 64: ...60 COMPONENT DIAGRAM DFM Digital PC Board TOP...
Страница 65: ...61 COMPONENT DIAGRAM DFM Digital PC Board BOTTOM...