GlucoMen
®
LX - 38973 - 09/07
UK
GlucoMen
®
LX - 38973 - 09/07
diS�playS�, MeaninGS�, aCTiOnS�
display
What it Means
What to do
Meter does
not begin test
countdown
after applying a
blood sample.
• Not enough blood sample.
• Sample applied after meter
automatically turned off.
• Meter may not be working properly.
• Test strip may be damaged.
Repeat the test with a new test strip
again. After 3 attempts, call A. Menarini
Diagnostics for assistance:
(UK) 0800 243667 -
(Ireland) 1800 709903.
diS�playS�, MeaninGS�, aCTiOnS�
display
What it Means
What to do
Used Strip Error: Used or damaged
strip.
Retest with a new strip.
Blood Sample Error.
Incorrect application of blood sample or control
solution onto the test strip, or the test strip
may be damaged. Review your sampling
technique. The blood sample quantity may
be incorrect or hematocrit may be out of the
allowed range for glucose measurement.
No test results in the last 30 days.
No action required.
Meter does not
turn on
after inserting
a test strip.
• Battery is dead.
• Battery is installed incorrectly or there
is no battery in the meter.
• Test strip is inserted upside down or
not completely in.
Replace the battery. Check that the battery
is correctly installed with the “+” sign facing
you. Insert the test strip correctly with
the GlucoMen name and white tip facing
up and out. Call A. Menarini Diagnostics for
assistance: (UK) 0800 243667 - (Ireland)
1800 709903.