NOTE
To test your meter and test strips only ever use the GlucoMen® areo Control control solutions (provided
separately).
CAUTION
●
Do not
use control solutions after their expiry date. The expiry date is indicated on the control solution
vial next to symbol.
●
For accurate test results, allow the meter, test strips and control solution to adjust to the surroundings
for at least 30 minutes before running the control test:
- temperature: 5 to 45 °C (41 to 113 °F);
- humidity: 20 to 90% RH.
●
Do not use the control solutions if 3 months or more have passed since first opening the vial.
●
Do not
drink the control solution. It is not intended for human consumption.
●
Avoid contact of the solution with the skin and the eyes as this could cause inflammation.
To perform a control solution test, proceed as follows:
a
. Insert a new test strip into the test strip port.
The drop icon starts blinking on the screen
. If nothing
appears on the screen, remove the test strip, insert it into the test strip port again and wait for the drop
icon to start blinking.
b
.
Enable the CTL mode
(§2.1).