background image

A.H. Systems Model PAM-0118P Pre-Amplifier

INTENDED PURPOSES

This equipment is intended for general laboratory use in a wide variety of 
industrial and scientific applications and designed to be used in the process of 
generating, controlling and measuring high levels of electromagnetic Radio 
Frequency (RF) energy. It is the responsibility of the user to assure that the 
device is operated in a location which will control the radiated energy such that 
it will not cause injury and will not violate regulatory levels of electromagnetic 
interference.

HAZARDOUS RF VOLTAGES

The PAM-0118P preamplifier is reliable when operated under specified 
conditions; however the use of sensitive, high-performance semiconductors 
causes the preamplifier to be susceptible to ESD and EOS. The risk of damage 
due to ESD is small, but possible, because the charged object would have to 
come in contact with one of the center conductors. It is more likely that EOS 
damage will cause the failure of the preamplifier. 

ESD – (Electrostatic Discharge)
A discharge of static electricity can degrade or even worse, seriously damage 
the preamplifier. There are two kinds of ESD damage that can occur. The worst 
being severe resulting in complete failure of the preamplifier. This damage is 
relatively easy to create although often expensive and time consuming to fix. 
The second kind of damage is less severe and can result in degradation or 
weakening of components which will not result in evident failure. This kind of 
damage is difficult to detect and often results in faulty or degraded performance.

EOS – (Electrical Overstress)
Precautions against EOS should always be maintained while operating the 
preamplifier. It must be emphasized that excessive power applied to the RF 
input of the preamplifier will damage the input stage. EOS damage can even 
occur when storing the preamplifier in areas of strong EMI or RFI fields. 
Improper connection sequence is another possible source of EOS damage 
such as application of RF signals to the preamplifier before powering up. When 
connecting alternate sources of DC power caution should be made that 
excessive Input DC voltage is not applied as this is another source of EOS 
damage. Loose connections causing intermittent events can also damage the 
preamplifier.

 A.H. Systems inc. – May 2014

REV D

4

Содержание PAM-0118P

Страница 1: ...A H Systems Model PAM 0118P Pre Amplifier PAM 0118P Preamplifier Operation Manual A H Systems inc May 2014 REV D 1 ...

Страница 2: ...es 4 Range of Environmental Conditions 5 OPERATING INSTRUCTIONS Assembly Instructions 6 Mounting Instructions 6 Operating Instructions 6 GENERAL INSTRUCTIONS Specifications 7 Included Equipment 7 Optional Equipment 7 Typical Data 8 Conversion Formulas 9 MAINTENANCE AND WARRANTY Maintenance 12 Warranty 13 A H Systems inc May 2014 REV D 2 ...

Страница 3: ...sitivity by at least 35 dB and is ideal for low amplitude test measurements When measuring in the EMC lab or out in the field for general RFI testing this preamplifier has a rugged compact design BEFORE APPLYING POWER Review this manual and become familiar with this preamplifier and testing instructions Verify that the equipment line voltage is compatible with the main power source A H Systems inc...

Страница 4: ...city can degrade or even worse seriously damage the preamplifier There are two kinds of ESD damage that can occur The worst being severe resulting in complete failure of the preamplifier This damage is relatively easy to create although often expensive and time consuming to fix The second kind of damage is less severe and can result in degradation or weakening of components which will not result i...

Страница 5: ...A H Systems and will void the warranty To connect 1 Apply DC Voltage 2 Connect output load 3 Apply RF input To disconnect 1 Remove RF input 2 Remove load 3 Remove DC voltage RANGE OF ENVIRONMENTAL CONDITIONS This equipment is designed to be safe under the following environmental conditions Indoor use Altitude up to 2 km Temperature 5 C to 40 C Maximum relative humidity 80 for temperatures up to 31...

Страница 6: ...M 0118P preamplifier is quite simple The unit is turned on by first plugging the phone type plug from an external power supply into the power input connector The power supply is then plugged into the appropriate line voltage supply Ensure that the preamplifier has sufficient time 15 minutes to warm up and stabilize Pushing the switch to the on position turns on the preamplifier and the power indic...

Страница 7: ...nd uses a 2 5mm coax output connector The preamplifier has a low voltage indicator that allows you to power the preamplifier with an external battery supply OPTIONAL EQUIPMENT The following is a recommend accessory list for the Model PAM 0118P preamplifier SAC 18G 0 5 0 5 meter high frequency low loss cable to interconnect the preamplifier to the receiver or analyzer This high performance cable is...

Страница 8: ...A H Systems Model PAM 0118P Pre Amplifier TYPICAL DATA The following is typical data of the Model PAM 0118P preamplifier A H Systems inc May 2014 REV D 8 ...

Страница 9: ...enna factor plus cable loss to receiver reading in dB V to convert to field strength in dB V m Field Strength dBuV m SA dBuV AF dB m cable loss dB SA Spectrum Analyzer or Receiver voltage reading AF Antenna Correction Factor CL Cable Loss in dB IMMUNITY TESTING For Immunity measurements the generated electric field strength can be calculated by FS Approximate Field Strength in V m P Power in watts...

Страница 10: ...logic First consider the pointing vector which relates the power density W m2 to the electric field strength V m by the following equation P E 2 Where is the free space characteristic impedance equal to 120 Transforming this equation to decibels and using the appropriate conversion factor to convert dBW m2 to dBmW m2 for power density and dBV m to dB V m for the electric field the constant becomes...

Страница 11: ...s T is related to the magnetic field strength H in A m by the permeability of the medium in Henrys per meter H m For free space the permeability is given as o 4 x 10 7 H m Converting from T to pT and from A m to A m and deriving the Log the constant becomes 240 120 20Log10 4 x 10 7 2 0 dBpT dB V dBpT V Cable Loss dB V m dBpT 49 5 dB A H Systems inc May 2014 REV D 11 ...

Страница 12: ... is always a concern To ensure reliable and repeatable long term performance annual re calibration of your Model PAM 0118P preamplifier by A H Systems experienced technicians is recommended Our staff can recalibrate almost any type or brand of antenna For more information about our calibration services or to place an order for antenna calibration visit our website www AHSystems com or call 818 998...

Страница 13: ...ng with the delivery of the product by A H Systems The remedy set forth herein shall be the only remedy available to the buyer and in no event shall A H Systems be liable for direct indirect incidental or consequential damages This warranty shall not apply to any part of the product which without fault of A H Systems has been subject to alteration failure caused by a part not supplied by A H Syste...

Отзывы: