background image

 

INSTALLING THE COMBIped

 BRACE 

 

> The brace thickness “hard“ should be used for very thick and 
rigid nails only, therefore, please use “soft“ and “medium“ for 
most of the applications! 

 

CAUTION

:  There  is  danger  of  the  nail  to  come  off  when  using  a  brace 

with too  much  tension.  That  is  why  moderate  tension  should  be  applied 
only. If the patient feels too much tension, remove the brace and replace 
it by the next lower tension level. 

1. Preparations 

 
Preparing the nails and cleaning the nail fold 

Remove  any  non-adhesive  parts  from  the  nail 
surface.  Smoothen  extremely  rough  nails  or  slightly 
roughen extremely smooth nails. 
Clean  the  nail  fold  and  prepare  the  area  such  as  to 
enable  the  hook  to  be  fixed  under  the  nail  edge 
before installing the COMBIped

 braces. 

 

Disinfecting the brace  

Disinfect  the  wire  of  the  COMBIped

  brace  before 

bending  the  hook  and  also  each  time  before  you 
install it by wiping with the alcohol swab. 

 

Selecting the right brace  

 

Please choose the right brace size depending on the 
nail thickness (force needed). 
 

soft

:    

Wire thickness of 0.25 mm – for soft and normal 
nails 

 

medium:   

Wire thickness of 0.30 mm – for normal and thick 
nails 

 

hard:     

Wire thickness of 0.35 mm – for extremely rigid and 
thick nails 

 

Содержание COMBIped

Страница 1: ...SET OF INSTRUCTIONS COMBIped...

Страница 2: ...2...

Страница 3: ...S 5 INSTALLING THE COMBIPED BRACE 6 1 Preparations 6 2 Adjusting the brace 7 3 Cleaning nail and brace 10 4 Adhering the COMBIped brace 11 5 Cutting off the locking pin and smoothing 13 REMOVING THE C...

Страница 4: ...ply tension on both sides the brace should be as long as possible and should stretch accross the entire width of the nail You may also set two COMBIped braces one after the other also with different w...

Страница 5: ...products and instruments you need in order to use the COMBIped brace 1 COMBIped braces in three different sizes 2 Adhesive with brush 3 Alcohol swab for cleaning nail and brace 4 3TO combination plier...

Страница 6: ...esive parts from the nail surface Smoothen extremely rough nails or slightly roughen extremely smooth nails Clean the nail fold and prepare the area such as to enable the hook to be fixed under the na...

Страница 7: ...mbination pliers 2 Adjusting the brace Grab the brace Fix the brace on both pins using the COMBIped fixation pliers see photo A clearance between the pad and the COMBIped fixation pliears will keep th...

Страница 8: ...ition and then press the pliers together completely The hook should not protrude on the other side of the pliers Cut off the hook short with the side cutting nippers The flat side of the nippers will...

Страница 9: ...strong the wire may be pulled out of the pad thereby destroying the brace Check brace tension and optimse Do not bend the wire of the COMBIped brace leave it straight first Please agree the level of...

Страница 10: ...d with the alcohol swab Allow nail and brace to dry off well after cleaning to ensure that the adhesive will hold It is recommended to always use a lower brace tension rather than bending the brace Pl...

Страница 11: ...ce will not hold on the nail if too much adhesive is used Keep the edge and the thread of the bottle clean Clean with the swab if needed Slightly wipe off the brush on both sides Apply a thin layer of...

Страница 12: ...show too much upwards the adhesive surface will have no contact with the nail and the brace will not hold Very slightly pull the pliers to avoid the hook to jump off and then use the COMBIped pliers a...

Страница 13: ...g the side cutting nippers as short as possible one after the other one pin after the other 2 cuts Smoothen the pad with one of the sandpaper files 5 Cutting off the locking pin and smoothing CAUTION...

Страница 14: ...d newly every 4 8 weeks as needed until the treatment has been completed successfully Use the side cutting nippers to remove the COMBIped brace from the nail Remove any remaining adhesive and smoothen...

Страница 15: ...with the eyes or internal organs consult a doctor immediately Do not inhale its vapours It may also not come into contact with wounds cover up any wounds prior to using it with a coating of oil or gre...

Страница 16: ...16 3TO GmbH Birkenstra e 8 D 82041 Deisenhofen 49 89 20353444 Fax 49 89 20353445 info 3to gmbh de www 3to gmbh de 3TO GmbH March 2010...

Отзывы: