![3TECH TURBINE Ti-940L Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/3tech/turbine-ti-940l/turbine-ti-940l_manual_3202560014.webp)
27
26
TURBINE Ti-940L can off er the patented drive air pressure regulating system (Patent No.ZL2013 20060022.8) to maintain a stable supply of the
pressure to the turbine to reach consistent rotation and hence better cutting eff ect.
Coupling system available in the Ti-940L turbines:
Type
Compatible with other brands speed coupling
Notice
KaVo
®
LED
KaVo
®
MULTIfl ex
®
LUX
K2 with spray ring
Sirona
®
Sirona
®
LED coupler
With spray ring
W&H
®
W & H
®
Roto Quick
With spray ring
Bien Air
®
Bien-Air
®
Unifi x
®
LED Coupler
B2 with spray ring
NSK
®
NSK
®
PTL-CL coupler
With spray ring
5. Specifi cations and type
Head type
Torque head
Speed range
320,000 rpm - 350,000 rpm
Power
≥ 22 W
Retention power
≥ 30 N
Nozzle spray mode
4-port spray
Bur removal
Push button
Bur specifi cations
As per ISO1797-1
Coupling type
LED coupling
Voltage of LED Bulb
3 V
Illumination
≥ 25 000 LUX
Drive air pressure
250 ± 10% kPa (2.5 ± 10% kgf/cm²)
Air consumption
40 Nl/min ± 10%
Recommended water amount
> 50 ml/min
6. Assembly and removing the medical device
6.1 When LED coupling KAVO
®
Multifl ex
®
is assembled
6.1.1 Assemble the LED coupling and attach the medical device
– Screw the LED coupling onto the turbine hose (Fig.6-1①).
– Tighten the screw cap of the turbine hose with the wrench (delivered with the medical device) (Fig.6-2②③).
– Assemble the medical device accurately on the LED coupling until it audibly locks in the medical device (Fig.6-2④).
6.1.2 Removing the medical device
Hold the LED coupling tightly; pull the medical device off while twisting slightly (Fig.6-3⑤).