5.5 Eye safety

WARNING

The scanner emits bright light from its tip during operation . 
The scanner complies with EN (IEC) 62471 (Photobiological 
safety of lamps and lamp systems) . However, we recommend  
caution when handling the scanner .

A brief glimpse of the light into the eye is not dangerous, 
but do not stare into the beam or view directly with optical 
instruments, and do not aim the beam towards other 
people’s eyes . In general, viewing bright light sources 
(e .g . headlights), which have a high secondary exposure 
potential due to their blinding effect, may cause temporary 
reduction in visual acuity and afterimages, leading to irri-
tation, annoyance, visual impairment, and even accidents, 
depending on the situation .

5.6 Hygiene

WARNING

To maintain a clean work environment and the safety  
of the patient, wear clean examination gloves when:
•  Handling the scanner tip – changing the tip, setting  

the tip for upper/lower scans or handling the mirror .

•  Handling the protection tip .
•  Handling the color calibration kit .
•  Applying TRIOS for scanning patients .
•  Touching the TRIOS IOS system .

WARNING

Always clean the scanner unit and its optical window before 
attaching the protection tip . Do not allow patient contact 
with the protection tip .

WARNING

Always have the scanner tip mounted when you insert 
the scanner into the patient’s mouth or when the scanner 
comes in contact with the patient in any way . Mounting 
of the scanner tip is always required for maximal electrical 
safety and due to the risk of contamination .

WARNING

Always have a cleaned and sterilized scanner tip mounted 
before attaching the color calibration adaptor . Do not 
allow patient contact with the color calibration kit .

WARNING

Before each patient, you must:
•  disinfect the system,
•  apply a new sterilized or high-level disinfected tip to the 

scanner .

See the detailed instruction in the chapter below .

System S3P (pod) only

WARNING

Before each patient, make sure any PC connected to the 
system and its peripherals are disinfected . It is recom-
mended to use medical-grade peripherals (e .g ., keyboards 
and mice) that can be disinfected easily .

5.7 Safety when scanning with TRIOS

CAUTION

The handheld scanner is a highly precise optical instrument 
and must be handled carefully . Do not drop, shake or bump  
the handheld scanner, as these actions could cause it to break  
or malfunction . Follow these guidelines:
•  Always return the handheld scanner to its mount when 

not in use .

•  Do not submerge the handheld scanner in any liquids .
•  Do not place the handheld scanner on heated or wet 

surfaces .

•  Grip the handheld scanner firmly during scanning  

and when removing/returning it to its mount .

IMPORTANT NOTICE

The scanner will get warm during use . This is normal  
but not part of any patient treatment .

CAUTION

The handheld scanner tip is fragile and contains a delicate 
mirror . To avoid damage to the tip and mirror, take care not 
to bump the tip onto the patient’s teeth or restorations .

CAUTION

When the TRIOS system is used for pediatric patients, 
please pay extra attention while scanning the patient .

5.8 Protection from overheating

WARNING

The ventilation openings in the rear of the handheld  
scanner must never be blocked or obstructed .

If overheated, the scanner may become uncomfortably 
warm and eventually shut down .

System S3A (MOVE/MOVE+ cart) only

WARNING

Ventilation openings must never be blocked or obstructed . 
There are openings at the back of the MOVE/MOVE+ cart 
(behind the touch screen) and on the top and the bottom 
of the base of the MOVE/MOVE+ cart . If overheated, the 
system may stop functioning .

6. Cleaning, Disinfection,  
and Sterilization

6.1 Cleaning, high-level disinfecting,  
and sterilizing the scanner tip

The TRIOS scanner tip is the part of the scanner that is inserted into the 
patient’s mouth during scanning . The tip can be reused a limited number 
of times, but it must be cleaned and sterilized, or cleaned and high-level 
disinfected, between patients to avoid cross contamination .

Содержание TRIOS S3P-2

Страница 1: ...04 2021 3Shape TRIOS Safety and Setup Guide S3P 2 S3P 8 S3A 10 S3A 14 S3A 18 S3A 22...

Страница 2: ...rmity EU 3ShapeTRIOSA SherebydeclaresthattheTRIOSobservestheprovisionsof Medical Device Regulation EU 2017 745 MDR Restriction of Hazardous Substances in Electrical and Electronic Equipment Directive...

Страница 3: ...Electronic Equipment Directive 2012 19 EU Radio Equipment Directive 2014 53 EU CAUTION EU only CE marking for connected products Further products which are connected to this unit must also bear the CE...

Страница 4: ...t in PC 3 scanner batteries and a scanner battery charger You can find the figures in the final section of this guide All system variants comprise Handheld TRIOS intraoral scanner S3 2 S1 and S2 Wired...

Страница 5: ...our service provider if there is visible damage or malfunctioning System S3A MOVE MOVE cart only Please follow the steps below for assembling TRIOS IOS system Figure 5 Step 1 Place the MOVE MOVE cart...

Страница 6: ...is due Other messages include a reminder to mount the scanning tip Please make sure the scanning tip is mounted on the scanner Calibration A calibration of the handheld TRIOS scanner is required befo...

Страница 7: ...pendent on local reg ulations and the configuration of the product 5 Safety and Warnings 5 1 Prerequisites CAUTION Read all instructions carefully including all warnings and caution notices You must c...

Страница 8: ...it should immediately be calibrated before further use If calibration fails please contact your technical service provider See this guide s instructions on calibrating the handheld scanner WARNING If...

Страница 9: ...wer cable network cable and the smart cord must never be subjected to pulling stress Spilled liquids WARNING Do not bring liquids such as beverages near the TRIOS IOS system Do not spill liquids on th...

Страница 10: ...S3P pod only WARNING Before each patient make sure any PC connected to the system and its peripherals are disinfected It is recom mended to use medical grade peripherals e g keyboards and mice that ca...

Страница 11: ...manual for more information CAUTION New tips must be cleaned and sterilized high level disinfected before using them for the first time CAUTION A class B Vacuum autoclave according to EN 13060 is requ...

Страница 12: ...iod of use IMPORTANT NOTICE All surfaces that the protection tip comes in contact with particularly the scanner tube and its optical window must always be cleaned and disinfected before attaching the...

Страница 13: ...the correct rating and type 250 V T 3 15 A H must be used To replace the fuses first mount them in the drawer then insert the drawer back into the slot and push all the way in until the tabs click ba...

Страница 14: ...3Shape TRIOS A S Holmens Kanal 7 1060 Copenhagen Denmark 10030970A April 2021 Copyright 3Shape TRIOS A S Let s change dentistry together More information at 3Shape com...

Отзывы: