![3M Steri-Vac GS Series Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/3m/steri-vac-gs-series/steri-vac-gs-series_site-planning-and-installation-manual_3051960040.webp)
40
70-2011-5643-0 Issue Date: 10/2015
3M
™
Steri-Vac
™
Sterilizer/Aerator GS Series
Site Planning and Installation Guide
Note:
Figure 28 shows a 3M
™
Steri-Vac
™
Sterilizer/Aerator GS Series, Model GS5 on
the top shelf and a 3M
™
Steri-Vac
™
XL Aerator
on the bottom. Service hookups will vary
based on configurations. Refer to the
3M
™
Steri-Vac
™
Sterilizer/Aerator GS Series
and 3M
™
Steri-Vac
™
XL Aerator Installation
Guides for specific requirements.
Figure 28. Installation Print No. 8: 3M
™
In-Wall Mounting Unit Model 1238 Unit Rear Hookups
200-240 Volt
Supply for
Sterilizer as
required
Boost
Transformer
(If Needed)
Vent Duct
for Aerator
Compressed
Air Line
Damper
200-240
Volt
Supply for
Aerator
Vent Line
for
Sterilizer
Vent opening for
local exhaust
installation
(hospital-supplied
exhaust system
required)