63
Calibrate
Performing a calibration
Models SE and DL
Figure 5-3: Setting pre-cal screen
6. Allow the measurement to stabilize and then press
Arrows
to set 114.0
dB level.
7. Press
Enter
to store the new calibration. The Calibration History screen will
appear with the new calibration values in the display.
Figure 5-4: Calibration history screen
8. Press
On/Off/Esc
key to exit the Cal screen.
9.
For a
Post Calibration
, ensure the SoundPro is in “Pause” mode. (Essentially,
run a measurement and press the run/pause key.) The Pause indicator will
appear at the top of the display.) Then repeat the Calibration steps above and
press Enter to store the Post Calibration information. A sample screen is
displayed below with post calibration history.
Measurement
value
Measurement value
displayed graphically
Pre-Cal
Displays pre-
calibration
information
Press up/down softkey to
increase/decrease cal value
Содержание SP-SE-2-1/1
Страница 1: ...3M Personal Safety Division 3MTM SoundPro SE DL SeriesSound Level Meters SoundProUser Manual ...
Страница 13: ...xii Table of Contents Models SE and DL This page left blank intentionally ...
Страница 23: ...10 Detection Management Software DMS DMS Start Page Models SE and DL This page left intentionally blank ...
Страница 35: ...22 Turning off Navigating Models SE and DL This page left blank intentionally ...
Страница 69: ...56 Communications GPS Models SE and DL This page left blank intentionally ...
Страница 89: ...76 Back erasing Stopping a session Models SE and DL This page left blank intentionally ...
Страница 123: ...110 Speech Intelligibility STI PA results screens Models SE and DL This page left blank intentionally ...
Страница 137: ...AC DC output DC output 124 Models SE and DL This page left blank intentionally ...
Страница 149: ...Appendix A 136 Models SE and DL Block Diagram for SoundPro ...
Страница 161: ...Appendix B 148 Models SE and DL This page left blank intentionally ...