7
Liqui-Cel
®
Membrane Contactors
Cleaning Guidelines
Table 2: Cleaning Solution Flow Rate and Backpressure Guidelines
Contactor Size
Cleaning Solution
Flow Rate on Shell Side
Cleaning Solution
Flow Rate on Lumen Side
MiniModule
®
*
≤ 0.08 gpm
(≤ 300 ml/min)
≤ 0.13 gpm
(≤ 500 ml/min)
2 x 6
≤ 0.26
(≤ 1 lit/min)
Not
Applicable
2.5 x 8
1
– 2 gpm
(0.23
– 0.45 m
3
/h)
≤0.5 gpm
(≤ 0 .11 m
3
/h)
4 x 13
5
– 15 gpm
(1.1
– 3.4 m
3
/h)
4 - 6 gpm
(0.9
– 1.4 m
3
/h)
4 x 28, 6 x 28
,
8 x 20
10
– 30 gpm
(2.3
– 6.8 m
3
/h)
3 - 7 gpm
(0.68
– 1.60 m
3
/h)
8 x 40
40 - 60 gpm
(9.0
– 13.6 m
3
/h)
5 - 10 gpm
(1.1
– 2.3 m
3
/h)
8 x 80
60 - 80 gpm
(13.6
– 18.2 m
3
/h)
5
– 10 gpm
(1.1
– 2.3 m
3
/h)
10 x 28
30
– 40 gpm
(4.5
– 9.0 m
3
/h)
10 - 20 gpm
(2.3
– 4.5 m
3
/h)
14 x 28
50
– 60gpm
(11.4
– 13.6 m
3
/h)
10 - 20 gpm
(2.3
– 4.5 m
3
/h)
14 x 40
100
– 150 gpm
(22.7
– 34.1 m
3
/h)
10
– 20 gpm
(2.3
– 4.5 m
3
/h)
* MiniModules
®
operate with liquid on the lumenside
.
To prepare a cleaning solution when using untreated raw water, it is important to know the water
chemistry. We recommend using water that has been filtered and de-chlorinated. We recommend
using de-ionized water for cleaning if possible. We also recommend paying attention to sparingly
soluble compounds of Ca, Mg, Fe, Al, etc. and insoluble silica, SiO
2
.These elements can
precipitate onto the membrane when there is a pH shift in the water.