WINDOWS 2000 – Downloaded
soundtracks are stored in:
C:\Documents and
Settings\username\Desktop and
appear on the desktop.
7. The soundtrack filename format is:
’Trk*.e4k’ (* is the soundtrack
number on the stethoscope)
Example: soundtrack 2 = ’Trk2.e4k’
If the default soundtrack filename
already exists, a new name, ’Copy of
Trk*.e4k’ or ’Copy # of Trk*.e4k’ is
given, example: soundtrack 2 = ’Copy
of Trk2.e4k’ (# is the number of
duplicates from the same soundtrack)
It is advisable to rename the
soundtrack files once they are
transmitted to avoid confusion. It is
not necessary to maintain the file
extension *.e4k.
If
appears in the display of the
stethoscope, refer to the Error
Messages section of this manual to
determine a possible cause.
WINDOWS 95/98 - From the Start
menu select ‘Start – Programs –
Windows Explorer’. Open the folder
"My Received Files" and locate the
soundtrack file to be transmitted.
WINDOWS 2000 – From the desktop
(or within the folder that contains the
soundtrack file to be transmitted) locate
the soundtrack file to be transmitted.
1. Select the Model 4000 soundtrack
to receive the transmitted file
Caution: If the selected soundtrack
contains a recording, the transmitted
file will replace it. Erased soundtracks
cannot be recovered.
2. Position the two IR ports within
two feet.
3. Press the Model 4000 RECEIVE
button. The arrow will blink until the
infrared connection is made.
4. Immediately right click on the
soundtrack file to be transmitted and
select ‘Send To – Infrared Recipient’
5. Transmission is indicated on
the Model 4000 display by a rotating
. The computer status bar will
display two red lights flashing
towards one another.
6. Progress of the transmission is
shown on the Model 4000 display with
the
symbol.
Note: The soundtrack filename is not
transmitted to the stethoscope. If
appears in the display of the
stethoscope, refer to the Error
Messages section of this manual to
determine a possible cause.
Error Messages
If
shows while transmitting data to another Model 4000
or a computer:
• The path of infrared light may be blocked. Position the
infrared ports at each other and retry the operation.
• The infrared port may be dirty. Clean both infrared ports.
• The Model 4000 was removed before the soundtrack was
completely transferred. Do not move the stethoscopes until
the
is finished rotating.
• The Model 4000 may be exposed to excessive EMI/RFI,
electromagnetic noise. Remove the source of noise.
If none of these is the cause, refer to the Littmann
Stethoscope Service and Warranty section.
Headset Positioning
The Littmann Electronic Stethoscope Model 4000 is designed to
provide a comfortable, acoustically sealed ear fit. Notice that the
eartubes are permanently set at an angle to accommodate the
typical anatomy of the ear canal. The eartips should point in a
forward direction when inserted into the ear canals.
Correct Incorrect
To reduce spring tension in the
headset, hold each eartube at
the bend near the eartip and
gradually pull apart until fully
extended.
To increase spring tension,
grasp the headset with one hand
where the eartubes enter the
plastic tubing and squeeze until
the plastic tubing on one eartube
touches the other. Repeat as
necessary.
For maximum acoustic
performance, comfortable patented
3M™ Littmann® Soft-Sealing
Eartips are provided with the
stethoscope. The stethoscope
utilizes a unique design for
attaching the eartip to the eartube.
The eartips are pushed onto the end of the eartube and snapped
in place. To remove, pull firmly on the eartip.
Diaphragm Removal and Replacement
Under normal conditions, it is unnecessary to remove the rim
and diaphragm for cleaning. The diaphragm can easily be
cleaned with an alcohol wipe. If, however, it is necessary to
remove the rim and diaphragm, carefully follow instructions:
Rim Removal:
• With diaphragm side up, using
a thumbnail, pry the rim out of
its designated groove, and pull
towards you. If the diaphragm
does not come off along with
the rim, remove this separately.
Do not disturb the components located under the diaphragm
once it is removed.
Cleaning of the Rim and Diaphragm:
• Separate the diaphragm from the rim, and clean with mild
soapy water and/or alcohol. The surface of the chestpiece
can also be wiped with alcohol. Take care not to saturate
the chestpiece. Excess liquid used in the cleaning process
can result in moisture getting into the internal components.
Diaphragm and Rim
Reassembly:
• Once the rim and diaphragm
are completely dry, place the
rim on a firm surface. Insert the
diaphragm into the inside
Receiving a copy of
a soundtrack from
a computer using
Microsoft®
Windows®
95/98/2000
2
with an infrared
(IR) port
2. Prompts may vary between versions of Windows and are the responsibility of the Microsoft.
References to Windows as it pertains to use of the Littmann Electronic Stethoscope Model 4000 are
for clarification only. Please consult your Microsoft users manual for specific instructions.
Microsoft® and Windows® are trademarks of Microsoft Corporation