LED Indicator
•
Red - indicates dose has exceeded the alarm level.
• Amber - indicates stopped.
• Green - indicates measuring.
Calibrating
A calibration may be performed before and/or after a
dosimetry test.
1. Remove the windscreen while the Edge is turned on
and stopped.
2. Slide the Edge into the cavity of the
calibrator, ensuring the adapter
ring is inserted into the calibrator.
3. Turn on the calibrator. Set the
calibrator to 114 dB and 1000 Hz
(if applicable).
4. “CAL” will flash on the Edge display
automatically detecting the calibration
screen.
5. Press the key to trigger the calibration screen.
6. To adjust the calibration value on the display, repeatedly
press the up/down arrows to match 114.0 dB.
7. Press key to exit.
Viewing and/or navigating data
1. Press the the up/down arrow keys to navigate through the
measurements and parameters.
Setting up the Edge
All the Edge parameters are setup via 3M
TM
Detection
Management Software DMS in the Instrument Communication
panel or button from the DMS start screen.