Bair Hugger Model 505 Service Manual
12
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document.
www.bairhugger.com reorder #202430A
Tools/Equipment
Phillips-head screwdriver
• Use a calibrated temperature meter that can accept a male
subminiature connector and read a K style thermocouple
(e.g., a Fluke Model 52 K/J Thermometer).
• If the connector does not fit your meter, you can remove
the provided connector and attach a connector that fits your
meter.
Over Heat Alarm Testing
WARNING:
Perform all testing and adjustments of warming unit
temperature with Temperature Test Unit Model 22110. 3M
Patient Warming assumes no responsibility for the reliability,
safety, or performance of the Bair Hugger system if warming unit
temperature testing or adjustments are made in any other manner
than that described here. Improper measurement or adjustment of
the warming unit’s over heat alarm setting could result in patient
exposure to temperatures outside of the indicated range, and may
lead to patient injury.
NOTE:
The Temperature Test Unit Model 22110 is designed to
simulate the operating characteristics of Bair Hugger blankets and
Bair Paws gowns when used with Bair Hugger warming units; use
of the temperature test device with any other warming system may
result in imprecise readings.
The Temperature Test Unit Model 22110 temperature readings are
taken from a calibrated temperature meter.
Service Frequency
12-month intervals or 500 hours of use