3M © 2003  78-8078-8977-5                    Page 5 of 9                                           Attest 290 Auto-reader 

 

  INSTRUCTIONS FOR USE 

 
 

Wells are color coded to match the cap of  
each type of biological indicator

; i.e. the 1291 

RRBI with a blue cap goes into a blue well and the 
1292 RRBI with a brown cap goes into a brown 
well.

 

 
A yellow light will come on to indicate the RRBI is 
incubating in the reader well and should 

not

 be 

touched. A yellow light indicates the results are not 
available and incubation is in progress. 

 

Do not remove RRBIs until the red (+) or green 
(-) light indicates the test is complete. 

 

Note:

 

If RRBIs are inadvertently removed

 before 

the incubation period is complete, an alarm will 
sound and all three lights (+, •, and  -), for that 
incubation well will light. C2 Caution code will also 
be displayed. You 

must

 

return the RRBI to the 

reader well within 10 seconds or the results will 
be lost.

 

 
 
 
 
 
 
 
 
 

 
 
 
 
 

 
All red, yellow and green lights will flash for one 
second when the cover is opened demonstrating 
they are functional. 

 

 
The cover should 

only

 be opened when placing 

additional RRBIs into incubation wells or when 
removing RRBIs at the completion of the 
designated incubation period. 
 

 

A positive result (+) will automatically be shown as 
soon as it is detected. The full incubation period is 
required before a negative result (-) is shown. The 
Attest 1291 RRBI requires 1 hour of incubation and 
the Attest 1292 RRBI requires 3 hours of incubation 
before a negative (-) result will be shown. The 
negative result will be shown automatically after 
completion of the required incubation time. 
 
The processed indicator and the positive control 
may also be further incubated for a visual pH color 
change. The final negative reading (media remains 
purple) for a visual color change is made at 24 hours 
(1291) or 48 hours (1292). The positive control 
should show a yellow color change of the growth 
media.  For any incubation beyond 24 hours (1291) 
or 48 hours (1292), RRBIs should be moved to a 
humidified environment. 

 

Each facility must establish final readout time in its 
policy and procedures. The policy should be based 
on manufacturer’s instructions for use, scientific 
knowledge, current recommendation practices, 
applicable compliance requirements, inventory 
levels, and the risk to the patient if the medical 
device is not sterile. 

 

 
 

 

       

 

W

W

W

WARNING

ARNING

ARNING

ARNING

 

 

Do not remove or change placement 
of RRBI once it is placed into a well. 
Doing so may result in loss of results 
and the RRBI test may be invalid. 

Содержание Attest 290

Страница 1: ...3M Attest 290 Auto reader Operator s Manual Read and understand all safety information before using this product ...

Страница 2: ......

Страница 3: ...tTM 1292 Rapid Readout Biological Indicators at 60o C for a final negative fluorescence reading at 1 hour for 1291 or 3 hours for 1292 The 3MTM AttestTM 290 Auto reader is also designed to allow for further incubation of the 3MTM AttestTM 1291 RRBI and or 3MTM AttestTM 1292 RRBI for a final negative visual pH color change of the growth media at 24 hours for 1291 and 48 hours for 1292 A positive fl...

Страница 4: ...inside the plastic vial of the RRBI To avoid the risk of injury from flying debris due to a ruptured RRBI which if not avoided could result in serious injury or death Allow the RRBI to cool for the recommended time period before crushing Crushing or excessive handling of the biological indicator before cooling may cause the glass ampule to burst Wear safety glasses and gloves when removing the RRB...

Страница 5: ...he Attest 290 Auto reader complies with IEC EN 61326 1 as confirmed in the Certificate of Compliance generated by 3M The Attest 290 Auto reader complies with the EMC requirements of the CE mark EMC Directive 89 336 EEC The Attest 290 Auto reader complies with the Australian EMC requirements as confirmed in the Supplier s Declaration of Conformity that is linked to the C tick mark Note This equipme...

Страница 6: ...w this procedure for every RRBI to be read Step 1 While wearing gloves and safety glasses CLOSE RRBI cap by pressing down Step 2 CRUSH the glass ampule containing growth media in the designated crushing well built into the incubator Step 3 Hold the RRBI by the cap and TAP bottom of the vial on a tabletop until media wets spore strip at bottom of vial Do not tap vial on unit Step 4 Open the cover a...

Страница 7: ...ation period A positive result will automatically be shown as soon as it is detected The full incubation period is required before a negative result is shown The Attest 1291 RRBI requires 1 hour of incubation and the Attest 1292 RRBI requires 3 hours of incubation before a negative result will be shown The negative result will be shown automatically after completion of the required incubation time...

Страница 8: ...dicating an acceptable sterilization process After either the green or red light comes on the RRBI can be removed from the well The RRBI should not be removed or touched when the yellow light is on The positive unprocessed control RRBI must provide a positive fluorescent result red light Processed RRBI results are not valid until the positive control reads fluorescent positive red light With the p...

Страница 9: ...l will turn off indicating that the alarm is again active Time Remaining When the time remaining button is pressed the incubation time remaining for each individual incubation reader well that contains a RRBI will be displayed All three lights for each individual incubation reader well will illuminate as the time remaining is displayed The process begins with the incubation reader well containing ...

Страница 10: ...d before incubation Replace RRBI in proper well within 10 seconds complete All three lights for affected RRBI will light and alarm will sound if active E1 Temperature Control Error Call for repair or replacement This error will occur if system is no longer able to control the temperature of the incubator block 60ºC 2ºC E3 Memory Error Call for repair or replacement E4 Analog Digital Conversion Err...

Страница 11: ...Other Negative Control Vial Vial not crushed Crush vial before inserting into well Media did not wet the spore strip Tap vial on table until media wets spore strip Control vial was sterilized Use a non sterilized vial as a control Check chemical indicator on label Other Unexplained positive Adhesive residue on vial Do not place tape or labels on vial Test vial is absorbing fluorescent Place the vi...

Страница 12: ...ing 275 4E 01 St Paul MN 55144 1000 1 800 228 3957 3M Health Care D 41460 Neuss Germany Recycled paper 40 pre consumer 10 post consumer Litho in U S A with 3M film proofing systems and offset plates Attest is a registered trademark of 3M 3M 2003 78 8078 8977 5 Rev D Lit Code 70 2009 4089 1 ...

Отзывы: