background image

c.  The low air flow or low battery alarms trigger:

The air flow alarm activates when air flow falls below 6 CFM (170 lpm) for greater than 15-30 seconds. The
3M™ Adflo™ PAPR will shut down 2 minutes after continuous alarm activation.

The battery alarm - during approximately the last 15 to 30 minutes of charge, the bottom bar on
motor/blower battery LED will flash continuously one second ON, one second OFF. The audible alarm will
beep four times in two seconds, repeating every 30 seconds. The final 15 to 60 seconds the alarm will fast
beep every 1/2 second. 3M™ Adflo™ PAPR will shut down 30 seconds after the quick beep starts.

d.  In the event only an audible or only a visual alarm triggers, the user should still immediately leave the

contaminated area.

e.  Breathing becomes difficult.
f.  You feel dizzy or your vision is impaired.
g.  You taste or smell contaminants.
h.  Your face, eyes, nose or mouth become(s) irritated.
i. 

You suspect that the concentration of contaminants may have reached levels at which this respirator may no
longer provide adequate protection.

5.  Do not remove the respirator or reach your hand into the headgear in areas where the air is contaminated.
6.  Follow exiting and decontamination procedures specific to your workplace for turning off the motor/blower and

removing the respirator system.

 WARNING

The 3M™ Adflo™ PAPR is not intrinsically safe. Do not use in flammable or explosive atmospheres, doing so 

may 

result in serious injury or death.

Do not enter a contaminated area until properly donning the respirator system. Do not remove or turn off the respirator 
before leaving the contaminated area. Doing so 

may result in sickness or death.

Do not use the 3M™ Adflo™ PAPR system with a heavily loaded spark arrestor or without the spark arrestor screen 
properly installed. Exposing a heavily loaded spark arrestor, or filter assembly to direct contact with sparks or molten 
metal spatter may damage the filter, allowing unfiltered air into the breathing zone,

which may result in sickness or 

death, and may cause the filter or blower housing to ignite, resulting in serious injury, sickness or death.

CLEANING AND STORAGE

 WARNING

Dispose of lithium ion battery packs according to local environmental regulations. Do not crush, disassemble, dispose 
of in standard waste bins, in a fire or send for incineration. Failure to properly dispose of battery packs may lead to 
environmental contamination, fire or explosion. 

The 3M™ Adflo™ PAPR Assembly should be cleaned and inspected after each use and prior to storage.

NOTE:

 To avoid damage, do not use solvents for cleaning. Do not immerse in water or spray directly with water.

Cleaning

Detach the battery pack, breathing tube and headgear from the motor/blower. Inspect all parts for damage or other signs 
of excessive wear. Replace all damaged parts prior to storage or next use.
1.  Motor/blower and battery pack: Clean the outer surfaces of the 3M™ Adflo™ PAPR Assembly and battery pack with a soft

cloth dampened in a solution of water and mild, pH neutral detergent. Do not immerse the motor/blower or battery pack in
water. Do not use solvents or abrasive cleaners. Do not attempt to clean the interior of the motor/blower with compressed
air or vacuum. Dust accumulation in the motor/blower could indicate possible filter leakage or improper storage and must
be addressed prior to next use. Ensure the electrical contacts of the motor/blower and battery pack are dry.

2.  Breathing tube: Clean the exterior and connection sites on the breathing tube with the water and detergent solution.

Optional breathing tube covers can also be used to facilitate cleaning. Ensure the breathing tube is completely dry
before using or storing. The L-122SG and SG-50W breathing tube interiors contain sound absorbing material. They
cannot be immersed in liquids for cleaning and must be replaced if wet. Wiping the exterior is insufficient.

3.  HE filter: Open the filter cover and inspect the HE filter, pre-filter and spark arrestor. Replace if excessively dirty, wet or

damaged. The HE and pre-filters cannot be cleaned. The spark arrestor can be cleaned using a clean, soft cloth dipped
in a solution of water and a mild pH neutral detergent. Completely dry the spark arrestor with a clean cloth. If the spark
arrestor cannot be cleaned or is damaged, replace with a new spark arrestor.

Inspection

Follow inspection procedures listed in this 

User Instructions.

Motor/Blower Storage

The 3M™ Adflo™ PAPR system should be stored at a temperature between -4°F to 131°F (-20°C to 55°C). The motor/
blower should be stored in a clean environment out of direct sunlight.

Battery Pack Storage 

To help maximize battery pack service life:

Store the battery pack at -4°F (-20°C) to 115°F (45°C). R.H. < 85%. Storage of the battery pack outside of this range
will shorten service life of the battery.

The 3M™ Adflo™ PAPR Battery Pack may remain on the charger for 3 to 4 days after completion of charging, but for
maximum battery life, disconnect the charger after a full charge has been received.

Battery packs for long-term storage should be disconnected from the motor/blower.

Battery packs not intended for immediate or regular use should be charged upon receipt and placed on regular
recharge schedule (every 3 months of non-use). Prolonged storage of battery packs without regular recharge may
damage the battery pack cells.

SPECIFICATIONS

Technical

Air flow

Greater than 6 CFM (170 lpm)

Weight (approximate)
1.  Assembly (motor/blower; HE filter; belt;

battery pack and SG-30W breathing tube)

2. OV/AG 

Cartridge

3.  Nuisance Odor Pad Assembly

1.  3.6 lbs (1.65 kg)

2.  1.72 lbs. (0.78 kg)
3.  0.320 lbs (0.145kg)

Motor/Blower Assembly
1. Operating 

Temperature

2.  Storage Temperature (RH <90%)

3.  Operating Altitude and Limitation

4. Noise 

level

1.  23ºF to 131ºF (-5º to 55ºC)
2.  -4°F to 131°F (–20°C to 55°C)
If 3M™ Adflo™ PAPR system is stored at temperatures below 
32°F (0°C), the battery pack must be allowed to warm up to 
achieve optimal battery performance. 
3. Altitude*:

Less than 2500 feet (800 meters) MSL –HE filter and
OV/AG Cartridge, HE filter and nuisance odor pad assembly,
or HE filter alone may be used.

2500 to 3500 feet (800 to 1200 meters) MSL – HE filter
alone may be used.

Greater than 3500 feet (1200 meters) MSL – not
recommended. Use the high altitude version of the 3M™
Adflo™ PAPR (35-3099-01) or a different 3M™ PAPR.

* If you have questions regarding these products contact 3M
Technical Service.
4.  Noise levels – less than 80 dBA at user’s ear excluding external

noise.

Belt Size (Approximate)
Belt Extender

29.5 to 45.25 inches (75 to 115 cm)
Approximately 26 inches (66 cm)

Battery Pack/Charger
1. Chemistry
2. Run 

Time

3. Charging 

Temperatures

4. Storage 

Temperatures

5.  Battery Pack Service Life

1.  Lithium Ion (Li)
2. •  10-12 

hours*

5.5 hours with OV/AG cartridge*

*New battery pack and HE filter, battery life will be reduced if using
at altitudes greater than 3500 feet.
3.  32°F to 104°F (0°C to 40°C). If the charger’s LED blinks rapidly

during charging, the battery pack is either too hot or too cold to
charge.

4.  -4°F (-20°C) to 115°F (45°C). R.H. < 85%. Storage of the

battery pack outside of this range will shorten service life of the
battery.

5.  Approximately 500 charges.

6. Charger 

Codes

6. Charger 

Code 

Table:

Mode

Description

LED Indication

Standby

Pre-Charge

f

Waiting 

tempera

or correct 

ture

Rapid charge

charge

Maintenance 

Ready

Error

Yellow Permanent on

Yellow flashing slow

Y

alterna

ellow and g

ting

reen 

Green flashing fast

Green flashing slow

Green permanent on

Yellow flashing fast

Yellow

Green

Yellow

Green

Yellow

Green

Yellow

Green

Yellow

Green

Yellow

Green

Yellow

Green

Motor/Blower Alarms
Low air flow

Low battery pack voltage

Activates when air flow falls below 6 CFM (170 lpm) for greater 
than 15-30 seconds. 3M™ Adflo™ PAPR will shut down 2 mins. 
after continuous alarm activation.
During approximately the last 15 to 30 minutes of charge, the 
bottom bar on motor/blower battery LED will flash continuously one 
second ON, one second OFF. The audible alarm will beep four times 
in two seconds, repeating every 30 seconds. The final 15 to 60 
seconds the alarm will fast beep every 1/2 second. 3M™ Adflo™ 
PAPR will shut down 30 seconds after quick beep starts.

Hazardous Location (Intrinsic Safety)

The 3M™ Adflo™ PAPR Assembly is not approved for use in 
hazardous or intrinsically safe locations.

Shelf Life – HE Filter, Nuisance Odor Pad 
Assembly and OV/AG Cartridge 

5 years from date of manufacture.

Assigned Protection Factor

Refer to the 

User Instructions

 for the specific head gear to be used to determine the assigned protection factor (APF) for 

the 3M™ Adflo™ PAPR system. Consult 3M™ Technical Data Bulletin #175 for additional information on APFs and 
supporting test data.

Service Life
Chemical Cartridges 

The useful service life of the 3M™ Adflo™ PAPR Organic Vapor/Acid Gas Cartridge, 15-0399-99 or 15-0499-99, will 
depend on the specific type, volatility and concentration of the contaminants and environmental conditions such as 
humidity and temperature. Replace cartridges in accordance with an established change schedule, regulations, or 
sooner if smell, taste, or irritation from contaminants are detected.

Nuisance Odor Pad Assembly

The useful service life of the 3M™ Adflo™ PAPR Nuisance Odor Pad will depend on the environmental conditions and 
concentration of the contaminants and environmental conditions such as humidity and temperature. The nuisance odor 
pad should only be used for concentrations of organic vapors not exceeding the OSHA permissible exposure limit (PEL) or 
applicable government occupational exposure limits, whichever is lower. Replace if smell or taste is detected. 

Particle Filter 

The 3M™ Adflo™ PAPR Motor/Blower Assembly is equipped with a five-step particle filter-loading indicator. There are three 
green and two red indicators. The illumination of each indicator light signifies increased filter clogging and a reduction in 
battery run time. When the red indicator(s) are illuminated, battery run time will be significantly reduced and HE particle filter 
and pre-filter replacement is recommended. When using a chemical cartridge or nuisance odor pad assembly with a fresh 
particle filter, the first two LEDs of the scale will be illuminated due to the additional resistance to air flow. 
The HE filter should also be changed if it becomes wet or damaged. The pre-filter should be changed at least as frequently 
as the HE filter. The pre-filter and HE filter should be changed if the battery pack operating time becomes shortened or if 
the low air flow alarm is activated. 

LISTING OF COMPONENTS, ACCESSORIES AND REPLACEMENT PARTS

Part Number

Description

Battery Packs and Chargers

35-1099-07

Battery Pack, Li

35-0099-08

Lithium Ion Battery Smart Charger

Belts

15-0099-16

Belt, Leather

15-0099-06

Belt Tongue Extender Replacement, Leather

35-0099-14

Belt, Nylon webbing with quick connection

TR-325

Belt, Versaflo, Standard

TR-326

Belt, Versaflo, High Durability

BPK-01

Backpack

Breathing Tubes

15-0099-10
L-122SG
SG-30W
SG-40W

Breathing tube - 9000 HWR, 9000 Flexview and ClearVisor only
Breathing tube - L-SG series headgear only
Breathing tube - Self adjusting - 9100 series and M-100/300 headgear
Breathing tube - Heavy duty - 9100 series and M-100/300-series headgear

SG-50W

Breathing tube – Sound dampening – 9100 series helmets. Required when used with 
3M™ Adflo™ PAPR with HE Filter and OV/AG Cartridge and Welding Helmet 9100-Air, 
9100 MP, and must be used with M-400 headgear when used with any 3M™ Adflo™ 
PAPR configuration.

Filters and Cartridges – Refer to the NIOSH approval label for proper filter/cartridge/headgear combinations

15-0099-99X06

Pre-filter, 6-pack

15-0099-99X12

Pre-filter, 12-pack

15-0299-99X02

High Efficiency Particulate Filter, 2/case

15-0299-99X06

High Efficiency Particulate Filter, 6/ case

15-0299-99X36

High Efficiency Particulate Filter, 36/case

Содержание Adflo 35-5401-01

Страница 1: ...ithout altitude compensation electronics Below 2500 feet 800 meters mean sea level MSL the 3M Adflo PAPR can be used with the 3M Adflo PAPR HE Particulate Filter by itself or in combination with the 3...

Страница 2: ...attached properly the 3M Adflo PAPR will sit tight against the belt and will stay in place Backpack The BPK 01 backpack may be used in place of the belt Fig 9A 1 With either end of the motor blower t...

Страница 3: ...55 C If 3M Adflo PAPR system is stored at temperatures below 32 F 0 C the battery pack must be allowed to warm up to achieve optimal battery performance 3 Altitude Less than 2500 feet 800 meters MSL H...

Страница 4: ...motor 1 Check that the battery contact is not bent or broken 2 Fully charge the battery pack 3 and 4 Contact 3M Warranty and Repair Low air flow per indicator but no alarm s 1 Damaged circuit board 2...

Страница 5: ...omfort requirements of the user 3M Vortemp Heating Assembly V 200 The 3M Vortemp Heating Assembly is designed to provide those 3M headgear which are approved for use with it a continuous air ow rangin...

Страница 6: ...ly with OSHA standard 29 CFR 1910 134 which states that Airline couplings shall be incompatible with outlets for other gas systems to prevent inadvertent servicing of airline respirators with nonrespi...

Страница 7: ...aist 4 Connect the supplied air hose W 3020 between the air pump and the 3M Low Pressure Connector Assembly Adjust the pressure to within the acceptable range as per the Special or Critical Use Instru...

Отзывы: