INTENDED USE

The 3M Model 701 Megohmmeter is designed to measure
the resistance of static control work surfaces and flooring.
This is to be done in a manner as described in this operator's
manual. Measurements described in this manual refer to the
ESD Association Standards S4.1 (Worksurfaces-Resistive
Characterization) and S7.1 (Floor Materials-Resistive
Characterization of Materials). These standards are
available from the ESD Association (315-339-6937). 
Refer to the operator’s manual for complete use of this
instrument. Use of other components may cause improper
performance and or an unsafe condition.

SAFETY SYMBOLS

This marking adjacent to another marking
or a terminal or operating device
indicates that the operator must refer to
an explanation in the Operating
Instructions to avoid damage to the
equipment and/or personal injury.

WARNING:

Indicates a potentially

hazardous situation which, if not avoided,
could result in death or serious injury.

CAUTION:

Indicates a potentially

hazardous situation which, if not avoided,
may result in minor or moderate injury. It
also may be used to alert against unsafe
practices that may cause property
damage.

SAFETY INFORMATION

The Operator’s Manual contains cautions and warnings
alerting the user to hazardous operating and servicing
conditions. This information is flagged by CAUTION or
WARNING headings throughout this publication, where
applicable, and is defined in this manual under SAFETY
SYMBOLS. To ensure the safety of operating and servicing
personnel and to retain the operating integrity of this
Instrument, these instructions must be adhered to.

INTRODUCTION

DESCRIPTION

The Model 701 Test Kit for Static Control Surfaces includes
a hand-held megohmmeter (here-in after referred to as the
“Instrument” or “Model 701”) and accessories. The kit has
been specifically designed for evaluating the resistive
characteristics of static control surface materials and testing
installed surfaces as specified in ESD Association Standards.

The Instrument has four test functions which allow the user
to measure the resistance of a static control surface at either
of two test voltages, verify proper Instrument performance,
and check the electrical continuity of the test set-up.

ACCESSORIES

All accessories required for operating the Model 701 are
included in the test kit. Available replacement parts are
listed on page 6.

UNPACKING AND INSPECTION

Examine the shipping carton for obvious signs of damage.
If damage is suspected, open the carton and inspect the
Instrument for possible damage. If damage is noted, notify
the carrier and supplier immediately. If the Instrument
appears to be in good condition, read this Operator’s Manual
in its entirety. Check that all items are included with the Kit,
then run a series of familiarity tests as instructed in this
manual.

CONTROLS, CONNECTORS, 
AND INDICATORS

GENERAL

Before attempting any operation of the Instrument, become
familiar with each control. A thorough understanding of
how the Instrument operates will avoid undue mistakes and
prolong its useful life. 

Figure 3-1. Controls, Connectors, and
Indicators

1. Main Selector Switch:

Main function selector switch allows selection of desired test
function.

2. TEST Button:

This button engages the Instrument and activates the test function
selected.

3. OHMS Scale:

This scale is used in conjunction with the SURFACE TEST functions
to indicate the amount of resistance measured.

4. CONTINUITY Scale:

This scale indicates the total resistance of the test set-up (meter,
leads, weights, and calibration plate) in the CONTINUITY TEST
function.

5. BATTERY Scale:

This scale indicates the charge level of the main battery of the
Instrument in the BATTERY TEST function by verifying the open
circuit output test voltage.

6. Test Jacks:

These jacks are used to connect the Instrument to the test leads.

1

• Turn off the Instrument before connecting or 

disconnecting test leads, or before moving test weights.

• Do not use this Instrument with any accessories not 

specifically designed to be used with this product.

• Do not use this Instrument to measure live circuits.

• The circuitry enclosed in the Model 701 produces high 

voltages. Make sure that the main selector switch is in 
the OFF position before removing the back cover.

• Product intended for indoor use only.

• The test weights included in this kit are heavy. Exercise 

care in handling.

• Improper battery installation will damage this 

instrument.

Содержание 2290

Страница 1: ...Operator s Manual 701 Test Kit for Static Control Surfaces Megohmmeter ...

Страница 2: ... either of two test voltages verify proper Instrument performance and check the electrical continuity of the test set up ACCESSORIES All accessories required for operating the Model 701 are included in the test kit Available replacement parts are listed on page 6 UNPACKING AND INSPECTION Examine the shipping carton for obvious signs of damage If damage is suspected open the carton and inspect the ...

Страница 3: ...onds and then read the resistance from the OHMS scale After all readings have been completed return the Main Selector Switch to the OFF position RESISTIVE CHARACTERIZATION OF STATIC CONTROL WORK SURFACES GENERAL This section provides a summary of the types of surface measurements specified and described by ESD S4 1 Measurements are performed for three reasons 1 Periodic performance testing of inst...

Страница 4: ...ings are above 1x109 ohms the static control surface may be dirty Clean the surface using the manufacturer s recommended cleaning procedure followed by two clear water rinses and a rinse with 70 Isopropyl alcohol Allow the surface to dry overnight and retest If the resistance reading is infinite there is an interruption open in the ground connection repair and retest 2 QUALIFICATION OF INSTALLED S...

Страница 5: ... building ground or other suitable ground If you have questions concerning the correct ground refer to EOS ESD Standard 6 0 and or contact a qualified electrician The Resistance to Ground test is the most important test that can be done on a static control surface because it verifies that the surface is working correctly and will drain a static charge in a reasonable time This test involves measur...

Страница 6: ...of five tests per 5 000 square feet 464 5 m2 of floor material whichever is greater A minimum of three of the five tests should be conducted in those areas that are subject to wear or have chemical or water spillage or are visibly dirty MAINTENANCE GENERAL This section details the maintenance which may be performed by the user Performance by the user of any maintenance not specified here will void...

Страница 7: ...lative humidity up to 90 for altitudes up to 2000m unless otherwise noted Pollution degree 2 class 3 equipment Resistance Ranges Test Measurement Open Circuit Internal Function Range Voltage Resistance Continuity 0 to 10 M 10V 7V 500 10V Surface 105 to 1011 10V 7V 2 100V Surface 105 to 1011 100V 7V 2 Measurement Accuracy 5 2 of arc Measured internally at voltage source Battery Test Internal test o...

Страница 8: ...prohibited by law 3M will not be liable for any loss or damage arising from this 3M product whether direct indirect special incidental or consequential regardless of the legal theory asserted Shipping of Warranty Returns Pack the Instrument carefully and ship it prepaid and insured to the proper destination noted below DO NOT RETURN ANY ITEM WITHOUT HAVING FIRST RECEIVED RETURN AUTHORIZATION FROM ...