Service Manual 

Fire

CR

+

 

 Computed Radiography Scanner 

 

 

Fire

CR

+

 

 |  Installation 

5

 

 

Installation 

Table Top 

1.

 

Install on a flat and stable surface.  

 

 

2.

 

Adjust the feet underneath the scanner to ensure stability. 

 

 

3.

 

Allow 50cm (20 inches) of free space at the front for cassette insertion and 20cm (8 inches) in 
the back for cables and power switch access. 

 

 

 

Содержание FireCR+

Страница 1: ...ications which must be met by 3D Imaging Simulations Corp All information contained herein is intended for guidance purposes only and characteristics of the products and services described in this publication can be changed at any time without notice Products and services may not be available in your local area Please contact your local sales representative for availability information 3D Imaging ...

Страница 2: ...31 2299 Website www 3DISCimaging com E mail info 3DISCimaging com 3DISC Americas 22560 Glenn Dr Suite 116 Sterling VA 20164 USA Tel 1 703 430 6080 E mail info 3DISCimaging com 3DISC Europe Gydevang 39 41 3450 Alleroed Denmark Tel 45 88 276 650 E mail info 3DISCimaging com ...

Страница 3: ...ation Geometry 9 Calibration Procedure 10 Reader Diagnostics 12 Scanner Control 12 Diagnostics Window 13 Parts Removal Replacement 14 Top Cover 14 Internal Cover 16 Main Board 17 Core Board 18 BLDC Board 19 Cassette Lock Motor Board 21 Optic Plate 22 Aligner Motor 23 Y Axis Motor 25 Fiber Bundle 27 Internal Frame 30 RFID Board 32 Power Board 33 Troubleshooting 35 Image Quality 35 Mechanical Noise ...

Страница 4: ...sk Space Network 1Gbps Ethernet Video 32 bit Color Display Video Resolution 1280 x 1024 Item Qty FireCR Computed Radiography Scanner 1 Fire Kit Optional Product Identification Card 1 14 x 17 Cassette 1 10 x 12 Cassette 1 14 x 17 Imaging Plate Installed in the cassette 1 10 x 12 Imaging Plate Installed in the cassette 1 IP Extractor 1 USB 2 0 Interface Cable 1 Power Adapter 1 Power Cord 1 Transport...

Страница 5: ...tion 5 Installation Table Top 1 Install on a flat and stable surface 2 Adjust the feet underneath the scanner to ensure stability 3 Allow 50cm 20 inches of free space at the front for cassette insertion and 20cm 8 inches in the back for cables and power switch access ...

Страница 6: ...omputed Radiography Scanner 6 Installation FireCR Wall Mount 1 Turn reader on its side Note Do not turn upside down 2 Remove adjustable feet 3 Attach brackets to the scanner Note Washers go between the brackets and the bolts ...

Страница 7: ...r FireCR Installation 7 4 Secure wall mount bracket to wall Note Make sure that the wall bracket is mounted level 5 Carefully attach the scanner to the wall bracket Note Make sure all four bracket hooks are securely attached to the wall mount bracket ...

Страница 8: ...ections 1 Connect power supply and USB cable Cassette and IP Visual Inspection 1 Remove the IPs from the cassettes by hooking and pulling the plate using the IP Extractor 2 Visually inspect for defects air bubbles scratches dust etc 3 Inspect cassette housing for physical damage ...

Страница 9: ...Service Manual FireCR Computed Radiography Scanner FireCR Calibration 9 Calibration Calibration Geometry Note Exposure must always cover the entire cassette ...

Страница 10: ...lank button Note Use the same cassette for the entire calibration Step 4 5 6 Scan Low Medium High Dose Expose the entire cassette at the recommended values and if the value is out of range adjust mAs until the numbers are green Note mAs and calibration values are directly proportional so a 5 increase in mAs will equal a 5 increase in calibration value etc Step 7 Calibration Press and wait until th...

Страница 11: ...each cassette size to verify an artifact free calibration 14x17 Cassette Calibration 10x12 Cassette Calibration File Description File Description Cal0 tif Scan Blank Cal4 tif Scan Blank Cal1 tif Scan Low Dose Cal5 tif Scan Low Dose Cal2 tif Scan Mid Dose Cal6 tif Scan Mid Dose Cal3 tif Scan High Dose Cal7 tif Scan High Dose Section dat Cal0 Cal3 compressed together Section2 dat Cal4 Cal7 compresse...

Страница 12: ... Run scan 4 Stop scan 5 Select the resolution of scan 6 Laser on off 7 Eraser on off 8 Stage control current should be 40 40 9 Beam aligner control 10 PMT Photo Multiplier Tube gain 11 ADC Analog to Digital Converter offset 12 Diagnostics window on off 13 Save values highlighted in orange 14 Send a report to technical support 15 Connect to remote support ...

Страница 13: ...gnostics Window Gauge 1 Rotating mirror RPM Has to be 1800RPM 10 Gauge 2 Right PSD peak Gauge 3 Right PSD balance Gauge 4 Left PSD peak Gauge 5 Left PSD balance Gauge 6 Internal frame temperature Gauge 7 Mainboard temperature Gauge 8 Eraser temperature Gauge 9 Laser module temperature ...

Страница 14: ...ography Scanner 14 Parts Removal Replacement FireCR Parts Removal Replacement Top Cover 1 Turn the power off 2 Unplug the USB cable and power cord 3 Remove 4 front screws and 3 rear screws from the base plate 2 5mm hex 4 Pull back cover ...

Страница 15: ... Manual FireCR Computed Radiography Scanner FireCR Parts Removal Replacement 15 5 Slowly lift cover from the back Note Before removing the cover disconnect the LED connector underneath 6 Refit in reverse order ...

Страница 16: ...nt FireCR Internal Cover 1 Remove Top Cover Pg 14 2 Remove the 6 screws on top of the Internal Cover 2 Phillips Note Some scanners have 10 screws 3 Disconnect the cassette lock cables from the cassette lock board and feed them through the hole in the middle of the Internal Cover ...

Страница 17: ...r front When refitting the Internal Cover make sure that the cables on both sides go through the upper grooves Refit it from front to back and make sure that the end of the Internal Cover goes over the entire internal frame and not into the space between the internal frame and the back cover Main Board 1 Remove Top Cover Pg 14 2 Remove Internal Cover Pg 16 3 Disconnect all connections to the board...

Страница 18: ...ake sure to connect the cables correctly when reassembling The cables all have tags corresponding to descriptions on the connectors 6 Once you ve finished the replacement of the main board replace the internal and Top Cover Core Board 1 Remove Top Cover Pg 14 2 Remove Internal Cover Pg 16 3 Remove the Loctite on the nuts 4 Unscrew the 4 nuts on the core board ...

Страница 19: ...bon cables and wire connector 6 Replace the old core board with a new core board 7 Reinstall the nuts to secure the core board 8 Connect the 2 ribbon cables and wire connector 9 Put Loctite on the nuts BLDC Board 1 Remove Top Cover Pg 14 2 Remove Internal Cover Pg 16 3 Unplug 2 wire connectors from BLDC Board ...

Страница 20: ...Service Manual FireCR Computed Radiography Scanner 20 Parts Removal Replacement FireCR 4 Remove 4 screws 2 5mm hex 5 Reinstall in reverse order ...

Страница 21: ...nal Cover Pg 16 3 The Cassette Lock Motor Board is located front section of the Fire CR 4 Remove the three wire harnesses that connect to the board 5 Remove four screws holding the board down and lift up Note use a No 2 screwdriver bit 6 Reinstall in reverse order 7 Reconnect the motor harnesses to their marked locations on the board L Left and R Right ...

Страница 22: ...ve Top Cover Pg 14 2 Remove Internal Cover Pg 16 3 Remove 2 screws from each of the three spring loaded brackets 2 5mm hex Note Keep pressure over the spring while removing and re installing screws 4 Unplug the 3 wire connectors 1 to BLDC 2 to Main Board 5 Lift out Optic Plate 6 Reinstall in reverse order ...

Страница 23: ...uted Radiography Scanner FireCR Parts Removal Replacement 23 Aligner Motor 1 Remove Top Cover Pg 14 2 Remove Internal Cover Pg 16 3 Remove Optic Plate Pg 22 4 Remove internal frame from aluminum housing Pg 30 5 Remove back cover ...

Страница 24: ...ireCR 6 Remove 2 screws from aligner motor 2 5mm hex 7 Cut cable ties holding wiring harness to cross bar 8 Remove 3 BLDC bracket screws 2 5mm hex Note Wiring harness routes under BLDC bracket and along the side of scanner 9 Remove 2 cable clamps along the side of scanner 2 Phillips ...

Страница 25: ...FireCR Parts Removal Replacement 25 10 Unplug wire connector from Main Board 11 Process is the same for other aligner motor 12 Reinstall in reverse order Y Axis Motor 1 Remove Top Cover Pg 14 2 Remove Internal Cover Pg 16 3 Remove BLDC Bracket 2 5mm hex ...

Страница 26: ...5 Remove 2 cable clamps along the side of the scanner 2 Phillips 6 Unplug wire connector from Main Board 7 Loosen set screws on belt pulley on motor and axle 8 Slide both pulleys toward side of scanner until motor pulley slips off end of shaft 9 Remove 2 screws from Y Axis bracket 3mm hex 10 Lift out Y axis motor 11 Reinstall in opposite order ...

Страница 27: ... Scanner FireCR Parts Removal Replacement 27 Fiber Bundle 1 Remove Top Cover Pg 14 2 Remove Internal Cover Pg 16 3 Remove the 6 screws from the fiber bundle cover 2 5mm hex 4 Lift off cover 5 Remove the two screws from the PMT housing 2 5mm Hex ...

Страница 28: ...Service Manual FireCR Computed Radiography Scanner 28 Parts Removal Replacement FireCR 6 Loosen PMT connectors 7 Remove the two screws at the edge of the eraser 2 Phillips 8 Pull back eraser ...

Страница 29: ...fiber bundle 2 5mm hex 10 Lift out fiber bundle Note Be careful not to scratch the fibers while pulling out the fiber bundle and note the location of the dowel pin on each side 11 Reattach in reverse order Note When reattaching the fiber bundle make sure that the fiber bundle rests on the dowel pins before refitting the screws ...

Страница 30: ...er 30 Parts Removal Replacement FireCR Internal Frame 1 Remove Top Cover Pg 14 2 Loosen the reset button by removing the two screws attaching it to the base plate 2 5mm hex 3 Remove six 3mm hex screws connecting the internal frame and the base plate ...

Страница 31: ... Remove the internal frame by lifting on the edge of each side Note Do not lift on anything but the Internal Cover as pulling on components will damage them 6 When reattaching the internal frame note that there are two pins at the bottom of the internal frame that fit into the base plate Make sure that these pins fit before refitting the screws to the bottom of the base plate ...

Страница 32: ...y Scanner 32 Parts Removal Replacement FireCR RFID Board 1 Remove the Top Cover Pg 14 2 Remove Internal Frame Pg 30 3 The RFID board is located in the front mouth of the Fire CR 4 Disconnect all three harnesses going to the cassette lock motor ...

Страница 33: ...ng to the RFID Board 7 Remove the six screws holding the RFID Board to the plexi guides 2 Phillips Note make sure to not lose or drop the spacers 8 Reinstall in opposite order Power Board 1 Remove Top Cover Pg 14 2 Remove Internal Cover Pg 16 3 Disconnect all the cables that are connected to the Power Board 4 Use the tweezers to disconnect or connect all the cables which are not reachable by hand ...

Страница 34: ...er 34 Parts Removal Replacement FireCR 5 Loosen the 3 screws that hold down the Power Board 2 5mm hex 6 Remove Power Board 7 Connect all the cables two switch cables USB cable and power cable onto a new Power Board using the tweezers 8 Replace screws ...

Страница 35: ...em is likely to be found Clean the fiber ends the bounce mirror 45 degree mirror the F theta lens the rotating mirror and the laser and make sure that no cables are blocking the path of light Vertical stripes in the image many lines This artifact is more a pattern of stripes than several individual stripes and is caused by either an improper calibration an unstable reader or an unaligned optics pl...

Страница 36: ...posed An overexposure can be determined by an exposure index at 2500 or higher Noisy Images There are two main causes of noise 1 Image artifacts related to a Mainboard and PMT 2 The noisy artifacts in the mainboard is that both image and noisy artifacts displays on the screen so that user can recognize both image and noisy on the image This Noisy artifacts is not in high noise but it can reduce th...

Страница 37: ...ther way to check this issue is to go the Control panel Click Show diagnostics You should have some type of waveform in the Right Peak and Left Peak field If there is no waveform flatline it means that there is no single from the system so that the either the BLDC board or the Optics board assembly will be the cause Figures 4 6 show in this case you can also check it on the Control Panel see the S...

Страница 38: ...mirror Figure 6 Example of defective optics module unstable Laser power Figure 7 Example of horizontal lines from a faulty Y axis motor or an incorrect stage alignment To correct this issue please follow the mechanical troubleshooting manual readjust the Stage assembly including Timing belt or replace the defective Y axis motor ...

Страница 39: ...1 Remove Top Cover Pg 14 2 Remove the Internal Cover Pg 16 and locate the Y axis motor 3 Loosen the two Allen screws on the Y axis motor s base 4 Pry the Y axis motor forward keeping it straight until the tension on the belt is fairly tight 5 Tighten the screws 6 Reinstall in reverse order Unaligned stage 1 Remove Top Cover Pg 14 2 Remove the Internal Cover Pg 16 3 Loosen the two Allen head screws...

Страница 40: ...Service Manual FireCR Computed Radiography Scanner 40 Troubleshooting FireCR ...

Отзывы: