9

3.3.1 Operating mode: electrical
3.3.2 Operating mode: electrical
zero-point center

Set the slide switch (4) to the “

setting.

Using the rotary switch (1) select
the corresponding measurement
range “V ...”.

The needle should now be posi-
tioned in the center of the scale.

Connect the multimeter and take
a reading from the lower scale.

3.4 Measuring alternating

voltage, directly up to 300 V

~

Set the slide switch (4) to “

setting.

Using the rotary switch (1) select
the corresponding measurement
range “V

~

”.

Connect the multimeter and take
a reading from the upper scale.

To reduce the effects of the fre-
quency, connect the socket “

⊥⊥

⊥⊥

directly to ground or to the point
with the lowest potential with re-
spect to ground.

3.5 Measuring AC voltage with

superimposed DC voltage

~

With the aid of a capacitor (recom-
mended: 4.7 µF/630 V) it is possi-

ble to isolate DC-voltage compo-
nents in an amplifier output stage,
for example. The resultant mea-
surement error is less than 0.2%
at a measurement frequency of
50 Hz.

Proceed with the measurement as
stated in point 3.4.

Measurement of the DC voltage
components is performed as de-
scribed under 3.3.

To avoid overloading, the set mea-
surement range must be greater
than the initially determined DC
voltage components.

Caution: before switching to a low-
er measurement range both volt-
age components must be checked.

3.6 Current measurement

When performing current mea-
surements the multimeter must
be connected in series with the
load in the circuit, which has the
lowest potential with respect to
ground.

3.6.1 DC measurement, directly

_

+

3.6.1.1 Operating mode: electrical
zero-point left

Switch the slide switch (4) to the

” setting.

Using the rotary switch (1), select
the corresponding measurement
range “A ...”.

Connect the multimeter take a
reading from the upper scale.

Содержание U17450

Страница 1: ...d if it is used in ac cordance with the instructions Do not operate or handle this unit incorrectly or inappropriately The device may only be used by persons who are aware of the haz ards of contact for voltages over 30 V rms and can undertake the appropriate safety precautions This also includes the appearance of unforeseen voltages e g in de fective units or charged capacitors In the case of vol...

Страница 2: ...ing coil movement which is not sensitive to external electromagnet ic fields The safety connection sock ets offer protection against acciden tal touch contact After approx 45 mins the battery is automatically switched off The power can be re established by turning off and on using the slide switch for the oper ating mode The robust plastic hous ing and the spring connected bear ing jewels of the m...

Страница 3: ...n The needle must be lo cated in the OFF position when the multimeter is in a hori zontal position If necessary make the corresponding adjustments with the adjustment screw 5 Check the electrical zero point Set the slide switch 4 into the position Select the measurement range the needle must be set to zero point in the center of the scale otherwise correct using the rotary knob 2 Check the battery...

Страница 4: ...ended 4 7 µF 630 V it is possi ble to isolate DC voltage compo nents in an amplifier output stage for example The resultant mea surement error is less than 0 2 at a measurement frequency of 50 Hz Proceed with the measurement as stated in point 3 4 Measurement of the DC voltage components is performed as de scribed under 3 3 To avoid overloading the set mea surement range must be greater than the i...

Страница 5: ...ake a reading from the lower scale 3 6 3 Measuring AC current Set the slide switch 4 to the setting Use the rotary switch 1 to select the corresponding measurement range A Connect the multimeter and take a reading from the upper scale 4 Maintenance 4 1 Cleaning Only use a paintbrush or soft tow el to clean the multimeter If stat ic electrical charge builds up on the view window this can be elim in...

Отзывы: