3B SCIENTIFIC PHYSICS 500 1003279 Скачать руководство пользователя страница 6

Object stage: 

x-y cross table, 155 x 145 mm

2

with object guide and coaxial adjustment knobs 
perpendicular to the object stage, adjustment 
range 50 x 76 mm

2

 

Illumination: 

Adjustable 6 V, 20 W halogen 

lamp incorporated into the base, universal 85 to 
265 V, 50/60 Hz power supply 

Condenser:

 Abbe condenser N.A.1.25 NA 0.65 

with iris diaphragm , filter holder and blue filter, 
focussed via rack and pinion drive 

Dimensions: 

306 x 190 x 407 mm³ approx. 

Weight: 

6.6 kg approx.

 

 
 

3. Unpacking and assembly 

The microscope is packed in a molded styro-
foam container.  

 

Take the container out of the carton remove 
the tape and carefully lift the top half off the 
container. Be careful not to let the optical 
items (objectives and eyepieces) drop down.  

 

To avoid condensation on the optical com-
ponents, leave the microscope in the original 
packing to allow it to adjust to room tem-
perature.  

 

Using both hands (one around the pillar and 
one around the base), lift the microscope 
from the container and put it on a stable 
desk.  

 

The objectives will be found within individual 
protective vials. Install the objectives into the 
microscope nosepiece from the lowest 
magnification to the highest, in a clockwise 
direction from the rear.  

 

Put the head onto the top of the stand and 
tighten the head-lock-screw. Insert the eye-
pieces into the tube. 

 
 

4. Operation 

4.1 General information 

 

Set the microscope on a level table. 

 

Place the object to be observed in the centre 
of the specimen stage and clamp it to the 
object guide. 

 

Connect the mains cable to the net and turn 
on the switch to get the object illuminated. 

 

Make certain that the specimen is centered 
over the opening in the stage.  

 

Adjust the interpupillary distance so that one 
circle of light can be seen. 

 

Make the necessary eyepiece dioptre ad-
justments to suit your eyes. 

 

To obtain a high contrast, adjust the back-
ground illumination by means of the iris dia-

phragm and the variable illumination control. 

 

Rotate the nosepiece until the objective with 
the lowest magnification is pointed at the 
specimen. There is a definite “click” when 
each objective is lined up properly. 

NOTE: 

It is best to begin with the lowest power 

objective. This is important to reveal general 
structural details with the largest field of view 
first. Than you may increase the magnification 
as needed to reveal small details. 

When 100x 

(oil) objective is chosen, objective oil must be 
dripped onto the slide.

 

To determine the magnification at which you are 
viewing a specimen, multiply the power of the 
eyepiece by the power of the objective. 

 

Adjust the holding brake to give a suitable 
degree of tightness in the focusing mecha-
nism. 

 

Adjust the coarse-focusing-knob which 
moves the stage up until the specimen is fo-
cused. Be careful that the objective does not 
make contact with the slide at any time. This 
may cause damage to the objective and/or 
crack your slide. 

 

Adjust the fine-focusing-knob to get the im-
age more sharp and more clear. 

 

Colour filters may be inserted into the filter 
holder for definition of specimen parts. 
Swing the filter holder out and insert colour 
filters. 

 

Use the knobs of the mechanical stage to 
move the slide side-, back- and forwards. 
The vernier provides accurate location of the 
specimen area. 

 

Always turn off the light immediately after 
use. 

 

Be careful not to spill any liquids on the mi-
croscope. 

 

Do not mishandle or impose unnecessary 
force on the microscope. 

 

Do not wipe the optics with your hands. 

 

Do not attempt to service the microscope 
yourself. 

 

4.2  Using the polarisation equipment 

 

Insert the analyser into the slot on the re-
volving nosepiece. 

 

Place the polarising filter on the rim aperture 
of the light source. 

 

Rotate the polariser until the planes of the 
polariser and the analyser are exactly 
crossed, so that one sees a black back-
ground. 

Any object with a doubly-refracting (birefringent) 
structure should now appear brightly illuminated 
against the dark background. If that does not 
occur, it is possible that the direction of light 

Содержание 500 1003279

Страница 1: ... Das binokulare Mikroskop ermöglicht die zwei dimensionale Betrachtung von Objekten dünne Schnitte von Pflanzen oder Tieren in 40 bis 1500 facher Vergrößerung Es ist mit einer Pola risationseinrichtung bestückt Stativ Robustes Ganzmetallstativ Stativarm fest mit Fuß verbunden Fokussierung über beidseitig am Stativ angebrachte koaxiale Stell knöpfe für Fein und Grobtrieb mit Kugellager und Feststel...

Страница 2: ...undbeleuchtung mittels der Irisblende und der regelbaren Beleuchtung einstellen Das Objektiv mit der kleinsten Vergrößerung in den Strahlengang drehen Ein Klick Ton zeigt die richtige Stellung an Hinweis Es ist am besten mit der kleinsten Vergrößerung zu beginnen um zuerst größere Strukturdetails zu erkennen Der Übergang zu einer stärkeren Vergrößerung zur Betrachtung feinerer Details erfolgt durc...

Страница 3: ...hen C B A Fig 1 Deckel des Lampenfachs A Rändelschraube B Lüftungsschlitze C Befestigungsschrauben der Lampenfassung 4 3 2 Sicherungswechsel Stromversorgung ausschalten und unbe dingt Netzstecker ziehen Sicherungshalter an der Rückseite des Mik roskops mit einem flachen Gegenstand z B Schraubenzieher herausschrauben Sicherung ersetzen und Halter wieder ein schrauben 5 Aufbewahrung Reinigung Entsor...

Страница 4: ......

Страница 5: ...pe allows two dimensional viewing of objects thin sections of plant or animal specimen in 40x to 1500x mag nification It is fitted with polarisation equipment Stand Robust all metal stand with arm perma nently connected to the base Focussing by means of separate knobs for coarse and fine adjustment located on either side of the stand and operated by rack and pinion drive with ball bearings and ret...

Страница 6: ...ragm and the variable illumination control Rotate the nosepiece until the objective with the lowest magnification is pointed at the specimen There is a definite click when each objective is lined up properly NOTE It is best to begin with the lowest power objective This is important to reveal general structural details with the largest field of view first Than you may increase the magnification as ...

Страница 7: ...alogen lamp and replace it with a new one Close the cover and secure it with the screw Push the lamp socket back into the original position and tighten screw C C B A Fig 1 Lamp socket cover A knurled screw B ven tilation slots C securing screws of lamp holder 4 3 2 Changing the fuse Turn off the power switch and unplug the mains plug Unscrew the fuse holder on the back of the stand base with a scr...

Страница 8: ......

Страница 9: ...et d observer des objets en deux dimensions coupes fines de plantes ou d animaux avec un agrandissement 40 à 1500x Il est équipé d un dispositif de pola risation Support Bâti support entièrement métallique Grande robustesse Mise au point rapide par pignon crémaillère à roulement à billes fine par vis micrométrique à commandes coaxiales de précision avec frein de serrage Butée de pro tection réglab...

Страница 10: ...par le plus petit agran dissement pour découvrir d abord les grands details de la structure Pour passer à un plus fort agrandissement pour voir des détails plus fins tourner le revolver jusqu à l objectif souhaité En cas d utilisation de l objectif 100x mettre de l huile sur le support de l objet La force d agrandissement résulte du produit du facteur d agrandissement de l oculaire et de l objecti...

Страница 11: ...ampe 4 3 2 Remplacement de fusible Couper l alimentation électrique et retirer impérativement la fiche secteur Dévisser le porte fusible à l arrière du mi croscope avec un objet plat par ex tourne vis Remplacer le fusible et visser de nouveau le porte fusible 5 Rangement nettoyage disposition Ranger le microscope à un endroit propre sec et exempt de poussière Si le microscope n est pas utilisé le ...

Страница 12: ......

Страница 13: ...are consente l osservazione bidimensionale di oggetti sezioni sottili di piante o animali con ingrandimento da 40 a 1500 volte Equipaggiato con un dispositivo di polarizzazione Stativo Robusto stativo completamente in me tallo braccio dello stativo fissato saldamente alla base messa a fuoco attraverso manopole coas siali per regolazione macrometrica e micrometri ca separate poste ai lati dello sta...

Страница 14: ...amente Regolare la distanza interoculare finché non sarà visibile un unico cerchio luminoso Adattare agli occhi il potere diottrico Per ottenere un contrasto elevato impostare l illuminazione posteriore attraverso il dia framma ad iride e l illuminazione regolabile Ruotare l obiettivo con l ingrandimento mi nimo fino a portarlo sul percorso dei raggi luminosi Il raggiungimento della corretta posiz...

Страница 15: ...lampada alogena e inserire quel la nuova Richiudere il coperchio e avvitarlo salda mente Riportare lo zoccolo della lampada in posi zione di partenza e serrare nuovamente le viti C C B A Fig 1 Coperchio del vano lampadina A vite a testa zigrinata B fessure di ventilazione C viti di fissaggio del portalampada 4 3 2 Sostituzione del fusibile Disconnettere l alimentazione elettrica ed estrarre assolu...

Страница 16: ......

Страница 17: ...roscopio binocular permite la observación bidimensional de objetos cortes delgados de tejido animal o vegetal ampliados entre 40 y 1500 veces Está dotado de un dispositivo de polarización Soporte Soporte de metal macizo brazo soporte de montaje fijo en el pedestal focalización por medio de ambos botones coaxiales para accionamiento burdo y fino con cojinete de bolas y freno de enclavamiento tope a...

Страница 18: ... hasta que sólo sea visible un círculo luminoso Se acomoda el número de dioptrias de los ojos Para lograr un contraste alto se ajusta la iluminación de fondo por medio del diafragma de iris y la iluminación regulable Gire el objetivo con el menor factor de ampliación hacia el haz de luz Un sonido de clic indica la posición correcta Nota Es mejor empezar con la ampliación más pequeña para en primer...

Страница 19: ...an los tornillos Se retorna el zócalo de la lámpara hacia la posición inicial y se vuelve a apretar el tornillo C C B A Fig 1 Tapa del cubículo de lámpara A Tornillo moleteado B Ranuras de ventilación C Tornillos se fijación del casquillo de la lámpara 4 3 2 Recambio de fusible Desconecte la alimentación de corriente Es imprescindible que también desconecte el enchufe de la red En la parte posteri...

Страница 20: ......

Страница 21: ...s técnicos O microscópio binocular possibilita a observação em duas dimensões de objetos finas lâminas cortadas de plantas ou animais com um aumento de 40 até 1500 vezes É equipado com um dispositivo de polarização Tripé Pé de apoio inteiramente metálico e robusto braço de apoio fixado no pé ajuste focal por meio de botões de ajuste fino e grosseiro coaxiais a ambos lados do pé de apoio movidos po...

Страница 22: ...tuar o ajuste por meio da íris e da iluminação ajustável Girar a objetiva com o menor grau de aumento até a linha do raio luminoso O som de um clique indicará que a posição está correta Observação é melhor começar a observação com a objetiva de menor grau de ampliação de forma a reconhecer primeiro os detalhes estruturais maiores A passagem para um grau maior de aumento obtêm se girando o revólver...

Страница 23: ...voltar a apertar o parafuso C C B A Fig 1 Tampa do compartimento da lâmpada A para fuso de dedo B fenda de ventilação C parafusos de fixação da rosca da lâmpada 4 3 2 Troca de fusível Desligue a alimentação elétrica e retire em todo caso o fio da tomada Desenrosque o suporte do fusível na parte posterior do microscópio com um objeto plano uma chave de fenda por exemplo Troque o fusível e volte a i...

Страница 24: ......

Отзывы: