background image

Condenser:

 Abbe condenser N.A.1.25 NA 0.65 

with iris diaphragm , filter holder and blue filter, 
focussed via rack and pinion drive 

Dimensions: 

240 x 190 x 425 mm³ approx. 

Weight: 

6 kg approx. 

 
 

3. Unpacking and assembly 

The microscope is packed in a molded styro-
foam container.  

 

Take the container out of the carton remove 
the tape and carefully lift the top half off the 
container. Be careful not to let the optical 
items (objectives and eyepieces) drop down.  

 

To avoid condensation on the optical com-
ponents, leave the microscope in the original 
packing to allow it to adjust to room tem-
perature. 

 

Using both hands (one around the pillar and 
one around the base), lift the microscope 
from the container and put it on a stable desk. 

 

The objectives will be found within individual 
protective vials. Install the objectives into the 
microscope nosepiece from the lowest 
magnification to the highest, in a clockwise 
direction from the rear.  

 

Put the head onto the top of the stand and 
tighten the head-lock-screw. Insert the eye-
pieces into the tube. 

 

Insert the analyser into the slot on the re-
volving nosepiece. 

 

Insert the condensor with polariser and tigh-
ten the lock-screw.  

 
 

4. Operation 

4.1 General information 

 

Set the microscope on a level table. 

 

Place the object to be observed in the centre 
of the specimen stage. Use the clips to fas-
ten it into place. 

 

Connect the mains cable to the net and turn 
on the switch to get the object illuminated. 

 

Make certain that the specimen is centered 
over the opening in the stage.  

 

Adjust the interpupillary distance so that one 
circle of light can be seen. 

 

Make the necessary eyepiece dioptre ad-
justments to suit your eyes. 

 

To obtain a high contrast, adjust the back-
ground illumination by means of the iris dia-
phragm and the variable illumination control. 

 

Rotate the nosepiece until the objective with 
the lowest magnification is pointed at the 

specimen. There is a definite “click” when 
each objective is lined up properly. 

NOTE: 

It is best to begin with the lowest power 

objective. This is important to reveal general 
structural details with the largest field of view 
first. Than you may increase the magnification 
as needed to reveal small details.  

To determine the magnification at which you are 
viewing a specimen, multiply the power of the 
eyepiece by the power of the objective. 

 

Adjust the holding brake to give a suitable 
degree of tightness in the focusing mecha-
nism. 

 

Adjust the coarse-focusing-knob which 
moves the stage up until the specimen is fo-
cused. Be careful that the objective does not 
make contact with the slide at any time. This 
may cause damage to the objective and/or 
crack your slide. 

 

Adjust the fine-focusing-knob to get the im-
age more sharp and more clear. 

 

Colour filters may be inserted into the filter 
holder for definition of specimen parts. 
Swing the filter holder out and insert colour 
filters. 

 

Always turn off the light immediately after 
use. 

 

Be careful not to spill any liquids on the mi-
croscope. 

 

Do not mishandle or impose unnecessary 
force on the microscope. 

 

Do not wipe the optics with your hands. 

 

Do not attempt to service the microscope 
yourself. 

 

4.2  Using the polarisation equipment 

 

Insert the analyser into the slot on the re-
volving nosepiece. 

 

Rotate the polariser until the planes of the 
polariser and the analyser are exactly crossed, 
so that one sees a black background. 

Any object with a doubly-refracting (birefringent) 
structure should now appear brightly illuminated 
against the dark background. If that does not 
occur, it is possible that the direction of light 
vibration of the object coincides with the polari-
sation direction. Whether or not that is the case 
can be tested by rotating the polariser or the 
specimen itself. 

A birefringent object, when rotated continuously, 
shows up brightly after each 90° rotation and is 
dark between these positions. In contrast, ob-
jects that are isotropic and not birefringent re-
main dark in all positions. 

 
 

Содержание 1012404

Страница 1: ...Tieren in 40 bis 400 facher Vergrößerung in polarisiertem Licht Stativ Robustes Ganzmetallstativ Stativarm fest mit Fuß verbunden Fokussierung über beidseitig am Stativ angebrachte koaxiale Stellknöpfe für Fein und Grobtrieb mit Kugellager und Feststell bremse einstellbarer Anschlag zum Schutz der Objektträger und Objektive Tubus Binokularer Schrägeinblick 30 Kopf um 360 drehbar Augenabstand zwisc...

Страница 2: ...eginnen um zuerst größere Strukturdetails zu erkennen Der Übergang zu einer stärkeren Vergrößerung zur Betrachtung feinerer Details erfolgt durch Drehen des Revol vers bis zum gewünschten Objektiv Die Stärke der Vergrößerung ergibt sich aus dem Produkt des Vergrößerungsfaktors des Okulars und des Objektivs Mit der Feststellbremse geeignete Span nung des Fokusiersystems einstellen Mit dem Triebknop...

Страница 3: ...ps mit einem flachen Gegenstand z B Schraubenzieher herausschrauben Sicherung ersetzen und Halter wieder ein schrauben 5 Aufbewahrung Reinigung Entsorgung Das Mikroskop an einem sauberen trocke nen und staubfreien Platz aufbewahren Bei Nicht Benutzung das Mikroskop immer mit der Staubschutzhülle abdecken Das Mikroskop keinen Temperaturen unter 0 C und über 40 C sowie keiner relativen Luftfeuchtigk...

Страница 4: ......

Страница 5: ...in 40x to 400x magnification in polarised light Stand Robust all metal stand with arm perma nently connected to the base Focussing by means of separate knobs for coarse and fine adjustment located on either side of the stand and operated by rack and pinion drive with ball bearings and retaining lever adjustable stopper for protecting the object slides and objective Tube Binocular head 30 viewing a...

Страница 6: ...s pointed at the specimen There is a definite click when each objective is lined up properly NOTE It is best to begin with the lowest power objective This is important to reveal general structural details with the largest field of view first Than you may increase the magnification as needed to reveal small details To determine the magnification at which you are viewing a specimen multiply the powe...

Страница 7: ... A knurled screw B ven tilation slots C securing screws of lamp holder 4 3 2 Changing the fuse Turn off the power switch and unplug the mains plug Unscrew the fuse holder on the back of the stand base with a screwdriver Replace the fuse and reinsert the holder in its socket 5 Storage cleaning and disposal Keep the microscope in a clean dry and dust free place When not in use always cover the micro...

Страница 8: ......

Страница 9: ...scope polarisant binoculaire permet d observer des objets en deux dimensions cou pes fines d échantillons de roche des plantes ou d animaux avec un agrandissement 40 à 400x de la lumière polarisée Support Bâti support entièrement métallique Grande robustesse Mise au point rapide par pignon crémaillère à roulement à billes fine par vis micrométrique à commandes coaxiales de précision avec frein de ...

Страница 10: ...c le plus petit agrandis sement dans le faisceau lumineux Un clic signale la bonne position Remarque commencer par le plus petit agran dissement pour découvrir d abord les grands details de la structure Pour passer à un plus fort agrandissement pour voir des détails plus fins tourner le revolver jusqu à l objectif souhaité La force d agrandissement résulte du produit du facteur d agrandissement de...

Страница 11: ... 4 3 2 Remplacement de fusible Couper l alimentation électrique et retirer impérativement la fiche secteur Dévisser le porte fusible à l arrière du mi croscope avec un objet plat par ex tourne vis Remplacer le fusible et visser de nouveau le porte fusible 5 Rangement nettoyage disposition Ranger le microscope à un endroit propre sec et exempt de poussière Si le microscope n est pas utilisé le reco...

Страница 12: ......

Страница 13: ...i di campioni di rocce piante o animali con ingrandimento da 40 a 400 volte in luce polarizzata Stativo Robusto stativo completamente in me tallo braccio dello stativo fissato saldamente alla base messa a fuoco attraverso manopole coas siali per regolazione macrometrica e micrometri ca separate poste ai lati dello stativo con cusci netto a sfere e freno di arresto battuta regolabile per la protezi...

Страница 14: ...ide e l illuminazione regolabile Ruotare l obiettivo con l ingrandimento mi nimo fino a portarlo sul percorso dei raggi luminosi Il raggiungimento della corretta posizione viene segnalato dallo scatto dell obiettivo Nota È opportuno cominciare con l ingrandimento minimo per poter riconoscere dapprima i dettagli macroscopici delle strutture Il passaggio a fattori di ingrandimento maggiori avviene a...

Страница 15: ...del portalampada 4 3 2 Sostituzione del fusibile Disconnettere l alimentazione elettrica ed estrarre assolutamente la spina Svitare il portafusibili sul lato posteriore del microscopio con un oggetto piatto ad es un cacciavite Sostituire il fusibile e riavvitare il supporto 5 Conservazione pulizia smaltimento Conservare il microscopio in un luogo pulito asciutto e privo di polvere Durante il perio...

Страница 16: ......

Страница 17: ...cos El microscopio de polarización binocular permite la observación bidimensional de objetos cortes delgados de muestras de rocas tejido animal o vegetal ampliados entre 40 y 400 veces con luz polarizada Soporte Soporte de metal macizo brazo soporte de montaje fijo en el pedestal focalización por medio de ambos botones coaxiales para accionamiento burdo y fino con cojinete de bolas y freno de encl...

Страница 18: ...ble un círculo luminoso Se acomoda el número de dioptrias de los ojos Para lograr un contraste alto se ajusta la iluminación de fondo por medio del diafragma de iris y la iluminación regulable Gire el objetivo con el menor factor de ampliación hacia el haz de luz Un sonido de clic indica la posición correcta Nota Es mejor empezar con la ampliación más pequeña para en primer lugar reconocer los det...

Страница 19: ...ornillos se fijación del casquillo de la lámpara 4 3 2 Recambio de fusible Desconecte la alimentación de corriente Es imprescindible que también desconecte el enchufe de la red En la parte posterior del microscopio desatornille el portafusibles con un objeto plano p ej Un destornillador Reemplace el fusible y vuelva a atornillar el portafusibles 5 Mantenimiento limpieza desecho El microscopio debe...

Страница 20: ......

Страница 21: ...o após o utilização do microscópio 2 Descrição dados técnicos O microscópio de polarização binocular possibilita a observação em duas dimensões de objetos finas lâminas cortadas de amostras de rocha plantas ou animais com um aumento de 40 até 400 vezes em luz polarizada Tripé Pé de apoio inteiramente metálico e robusto braço de apoio fixado no pé ajuste focal por meio de botões de ajuste fino e gr...

Страница 22: ...au de dioptria aos olhos Para se obter uma iluminação de fundo de alto contraste deve se efetuar o ajuste por meio da íris e da iluminação ajustável Girar a objetiva com o menor grau de aumento até a linha do raio luminoso O som de um clique indicará que a posição está correta Observação é melhor começar a observação com a objetiva de menor grau de ampliação de forma a reconhecer primeiro os detal...

Страница 23: ...afusos de fixação da rosca da lâmpada 4 3 2 Troca de fusível Desligue a alimentação elétrica e retire em todo caso o fio da tomada Desenrosque o suporte do fusível na parte posterior do microscópio com um objeto plano uma chave de fenda por exemplo Troque o fusível e volte a instalar o suporte 5 Armazenamento limpeza eliminação Armazenar o microscópio num lugar limpo seco e sem poeira Quando não u...

Страница 24: ......

Отзывы: