Luna II QX CPAP / Luna II QX Auto CPAP User Manual V1.1
5 / 40
6. Specifications
Device Size
Dimensions: 274mm×184 mm×115mm
Weight:1.9 kg
Water capacity: To maximum fill line 360mL
Product Use, Transport and Storage
Operation
Transport and Storage
Temperature: 5°C to 35°C (41˚F to 95˚F) -25°C to 70°C (-13˚F to 158˚F)
Humidity: 15% to 93% Non-condensing 15% to 93% Non-condensing
Atmospheric Pressure: 760
~
1060 cmH
2
O
760
~
1060 cmH
2
O
Heated Humidifier
Humidifier Settings: off, 1 to 5 (95˚F to 154.4˚F / 35°C to 68°C)
Humidifier Output: No less than 10 mg H
2
O/L
Environmental Conditions: Maximum airflow, 35°C, 15% relative humidity
Mode of Operation
Continuous
Work Mode
For LQX2000: CPAP
For LQX2A00: CPAP, Auto
SD Card
The SD card can record patient data and fault information
AC Power Consumption
100 - 240 V
~
, 50 / 60 Hz, Max 2 A
Device offer to USB Communications Port
5 V
2.0 A
Type of Protection Against Electric Shock
Class II Equipment
Degree of Protection Against Electric Shock
Type BF Applied Part
Degree of Protection Against Ingress of Water
IP22
Pressure Range
4 to 20 cmH
2
O (in 0.5 cmH
2
O increments), ≤30 cmH
2
O under single fault conditions.
Static Pressure Stability
±0.5 cmH
2
O