26
Control Solution Testing
STEP 6
The test result is complete when you hear
beeps. Your control solution test result is
shown on the display screen.
STEP 7
Compare the result displayed on the meter
to the control solution range printed on the
test strip vial. If the results you get are not
within this range, the meter and strips may
not be working properly. Repeat the control
solution test.
STEP 8
To turn your meter off, remove the used test
strip.
Control Test Range
0.6-33.3 mmol/dL
not following the instructions on this USER GUIDE,
expired, contaminated or watered-down control solution,
expired or damaged test strip,
control solution test done outside 20°C to 25°C (68°F to 77°F),
or a problem with the meter.
Out of range results may be caused by:
If you continue to get control solution results that fall outside the
range printed on the test strip vial, do not use the meter, the test
strips, or the control solution. Call the local distributor.