![2in1 SMART Скачать руководство пользователя страница 35](http://html.mh-extra.com/html/2in1/smart/smart_user-manual_439656035.webp)
Performance Characteristics
34
Within run precision
(100 venous blood tests per glucose level)
Total precision
(100 control solution tests per glucose level)
Concentration
Standard
deviation
(mg/dL)
Coefficient of
variation (%)
30~50 mg/dL
2.0
4.2
51~110 mg/dL
3.4
3.6
111~150 mg/dL
4.4
3.2
151~250 mg/dL
6.1
2.5
251~400 mg/dL
6.0
1.6
Concentration
Standard
deviation
(mg/dL)
Coefficient of
variation (%)
30~50 mg/dL
1.9
3.9
90~130 mg/dL
4.1
3.8
280~420 mg/dL
9.2
2.6
Regression statistics
Slope
1.0133
Intercept
-1.4079 mg/dL
R
2
0.9968
No. of samples
100
Range tested
33.5~445 mg/dL