57
Performance Characteristics
56
Regression statistics
Model
PG110 (2in1. MICRO
Blood Glucose
Monitoring System)
Test Method
Electrochemical
sensor
Calibration Method
Plasma
Sample
Whole blood,
capillary
Sample Size
0.5
μ
L
Test Time
5 seconds
Memory
No limit
Result Range
10~600mg/dL
(0.6~33.3mmol/L)
Hematocrit
20-60%
Operating
Temperature
10~40°C (50~104°F)
Operating
Relative Humidity
10-90%
Altitude
Up to 3,048 meters
(10,000 feets)
Power Source
From the smart
phone
Size
41.5 mm x 33.6 mm
x 9.0 mm
Weight
5.9 g (0.21 oz)
Within run precision
(100 venous blood tests per glucose level)
Total precision
(100 control solution tests per glucose level)
Slope
0.987
Intercept
0.469 mg/dL
R
2
0.989
No. of samples
100
Range tested
35.6~422.4 mg/dL
Mean glucose
Standard
deviation
(mg/dL)
Coefficient
of variation
(%)
41 mg/dL (2.3 mmol/L)
1.2
2.8
99 mg/dL (5.5 mmol/L)
2.7
2.7
131 mg/dL (7.3 mmol/L)
2.8
2.1
231 mg/dL (12.8 mmol/L)
3.7
1.6
373 mg/dL (20.7 mmol/L)
8.0
2.1
Mean glucose
Standard
deviation
(mg/dL)
Coefficient
of variation
(%)
43 mg/dL (2.4 mmol/L)
1.0
2.2
104 mg/dL (5.8 mmol/L)
2.4
2.3
297 mg/dL (16.5 mmol/L)
8.9
3.0
specifications