background image

2B Technologies   

2100Central Ave Suite 105, Boulder, Colorado 80301 USA                                                Model 306 Quick Start Guide Rev B 

Phone 303-273-0559 Fax 303-277-1812 http://www.twobtech.com 

4.  With your Model 306 attached to your ozone monitor via the connection tube, you may now 

begin generating ozone. Press the select switch on the front of the instrument and scroll to 
“O3”  in  the  menu.  Set  the  desired  amount  of  ozone  and  press  the  select  switch  again  to 
begin generating ozone. 
 

            

            

 

 

5.  Allow the bar located on the LCD to max out before you begin logging your measurements. 

 

 

 

6.  At this point your Model 306 is fully operational and may be used for calibration. 

 

7.  Please 

contact 

2B 

Technologies 

at 

(303)-273-0559 

or 

via 

the 

web 

at 

http://twobtech.com/techsupport.htm

 if you have any problems. 

 

 

Model 306 Installation Guidelines 

 

Typical Installation: Connection to a Device Drawing Flow from the Model 306 

The  Model  306  is  frequently  used  to  calibrate  an  ozone  monitor  or  to  provide  a  known  ozone 
concentration to another kind of equipment that draws flow from the calibrator.  The Model 306 is 
designed for this type of installation.  Inside the Model 306, an air pump maintains a total flow of ~ 
3 L/min through the ozone generator.  Any portion of the total flow that is not being drawn into the 
user’s  equipment  is vented through an internal ozone scrubber  and vented inside the  enclosure 
(Figure 1). 
 
 
 
 
 
 
 
 
 
 
 

Figure 1.

 Schematic of the Model 306 showing the internal overflow scrubber 

 and vent for any unused portion of the total flow rate. 

Model 306 Ozone Calibration Source 

 

Internal 

Overflo

Scrubb

er 

Vent 

Ozone 

Generato

 

Air 

Pump 

 

User’s 

Equipme

nt 

Отзывы: