1-CUBE GMA Скачать руководство пользователя страница 2

 

Finally REMEMBER TO CLEAN the GMA after each session

.  

1) 

Connect the inlet hose to a clean cold water supply

 (Max. pressure for GMA: 4 bar g).  

2) 

Flush through by opening both inlet and outlet valves

, with water flowing to drain until the 

chamber is clean. 

REMEMBER to

 unlock and 

pump the gas release plunger several times

 to 

avoid the piston becoming stuck with dry beer. Relock the plunger. 

If the plunger does become stuck, unscrew it and soak it in warm water until it frees. 
 

3) 

Disconnect the water supply

, with the valves still open. Lift the inlet hose and lay the 

instrument on its side so the 

water drains out via the outlet valve

. Finally, turn the instrument 

upside-down to drain the last few drops.  

4)  Close the valve gently and store the unit upright until it is next required.  

Normally cold water is sufficient for cleaning. In severe cases warm water (max. 25°C) or dilute 0.5 to 
max. 1% acetic or peracetic acid or hydrogen peroxide can be used at ambient temp. The GMA has 
brass valves, so do not leave them soaking or in contact with sanitizer for longer than is essential.  
Finally, washout thoroughly with clean water. 
 

Max. Temperature: 25°C. Max. Pressure for GMA: 4 bar g.

 (notice the gauge limit!)  

 
GMA construction materials: Chamber: PVC-U, Pump: Stainless Steel, Valves: Brass, Hoses: Silicone 
rubber 

 
Application Information  

Just  as  with  other  more  expensive  instruments  in  the  market,  the  dissolved  CO

2

  result  is  a  calculated  value 

based  on  the  measured  pressure  and  temperature  values.  The  calculation  formula  describes  detailed 
observations  and  measurements  of  a  physical  behaviour.  It  was  first  defined  in  1803  by  Dr William  Henry,  as 
‘Henry’s  Law  of  Partial  Pressures’.  The  mathematics  are  very  complex.  Over  the  years  numerous  learned 
people  have  developed  and  improved  various  different  formulae.  Nevertheless,  none  can  give  us  an  absolute 
and  100%  true  value.  Variations  between  different  instruments  and  different  ‘measuring’  methods  are  normal 
and must be expected. 

For  best  results,  a  good  understanding  of  the  physical  behaviour  and  a  consistent  operating  procedure  are 
essential. With this, reliable temperature and pressure measurements should be achievable. Attention should be 
given to achieving both temperature stability and true partial pressure equilibrium, which is not as quick or easy 
as one might think. For example, the temperature of the equipment, the operating environment or the operator’s 
hands  will  adversely  affect  the  beer  temperature  and  thereby  the  pressure.  They  both  influence  the  gas 
solubility.  Ideally  and  before  testing  starts,  the  instrument  should  be  at  exactly  the  same  temperature  as  the 
beer.  Even  the  shortest  exposure  of  the  beer  to  air  will  reduce  the  dissolved  CO

2

  content,  this  is  especially 

significant when testing kegs. The smallest of opening in pipes and seals will allow CO

2

 gas out and O

2

 & N

2

 in, 

even though a beer leak might not be seen

.   

 

 

 

 

 

 

 

 

We hope you enjoy making great beer, helped a little by your new 1-CUBE instrument. 

There is more equipment for brewers on our websites. 

 

 

 

 

 

www.1-cube.com 
[email protected]

 

 

 

 

Отзывы: