background image

               For further information about the functioning of 

 

the clinical AED Plus, see the AED Plus Operator’s Guide 

 

and Administrator’s Guide.       

 

 

               The Remote Controller lights indicate the current training 

settings while the Trainer2 simulates the actual AED Plus behavior. 

                The Remote Controller will turn itself off after 10 seconds.  

The Remote Controller will remember the last scenario setting which  

reappears the next time the  View/Enter Button is pressed.   

 

 

NOTE 

SETUP and TEST 

SELECT SCENARIO SETTINGS 

Scenario 3: SHOCK - NO SHOCK 

Scenario 2: NO SHOCK - SHOCK 

Scenario 1: SHOCK - SHOCK 

Manual Scenario Mode 

CONTROLS / INDICATORS 

Handheld Remote Controller: 

PUSH HARDER prompt 

(only when CPR Metronome is beeping). 

NO SHOCK  advised on next heart analysis 

(manual mode only). 

Pads attached/not attached to victim’s bare chest 

(toggle-switch). 

View/Enter scenario 

Show current scenario selection (press once) 

Start scenario (press again) 

Move Up  to next scenario. 

Scenario 4: NO SHOCK - NO SHOCK 

Set-up Controls: 

 

Training Controls: 

GOOD COMPRESSIONS prompt 

(only when CPR Metronome is beeping). 

Pause  (resume) current scenario. 

End current scenario. 

NOTE 

Install 4 “D” alkaline batteries in the AED Plus Trainer2.   

Trainer2 unit: 

Action: 

Press  On/Off  Button to turn Trainer2 on.   

Green light turns on, indicating Trainer2 is  

ready for use.   

Remote Controller: 

Install 2 “AA” batteries in handheld remote.  

Turn the Remote Controller on by pushing the 

View/Enter  Button.   

Point Remote Controller at unit and press 

View/Enter  Button again.  Green light on Remote 

Controller will show Manual Scenario Mode is 

selected. 

Manual Scenario Mode 
View/Enter Button 

Trainer2 unit: 

Action: 

Turn Trainer2 unit off (hold down  On/Off  Button).   

Setup test is now complete.  Trainer2 unit and  

Remote Controller are ready for use.   

Trainer2 unit: 

Action: 

Press the On/Off Button to turn on the Trainer2.  

Action: 

Remote Controller: 

Action: 

Use Move Up  Button to select desired scenario.  

Use CPR Metronome On/Off Button to turn 

Real CPR Help

®

 (beeping) on or off for this 

scenario.

 

Press the View/Enter Button once to transmit the  

selected Training Scenario to the Trainer2 unit.  

CPR PROMPTS 

Remote Controller: 

Action: 

The Instructor uses the Remote Controller to control prompts during a 

Training Scenario’s CPR period.  The instructor can use the following buttons 

to simulate the use of an AED Plus defibrillator with Real CPR Help.

 

Remote Controller: 

Action: 

In all scenarios (manual and automatic), the rescue 

can be paused by pressing the   Pause/Resume 

Button once. Press it a second time to resume the  

scenario.  

Trainer2 unit: 

Action: 

The Push Harder and Good Compressions  

Buttons are only valid during the CPR intervall.  

The red light on the Trainer2’s status window emits  

one long flash if the prompt is currently valid. 

If the Push Harder or Good Compressions  

Buttons are pushed when the Trainer2 is not  

beeping, the unit’s red light emits three quick 

flashes, indicating that the prompt is currently 

invalid. 

In all scenarios (manual and automatic), the rescue 

can be ended by pressing the End Button once. 

NOTE 

On 
Off 

On 
Off 

On 
Off 

 

 

The buttons above the line on the handheld Remote Controller are Set-up  

Controls, which the instructor uses to select a training scenario (an LED lights  

next to the selected scenario).  The buttons below the line are Training Controls,  

which the instructor uses to introduce events into the training scenario.   

• 
• 

• 

• 
• 

 

  

Good Compressions Button 

 

                Accurately imitating the student’s AED Plus unit requires  

 instructors be familiar with the protocol currently followed by the  
AED Plus for  PUSH HARDER  and  GOOD COMPRESSIONS  prompts. 

NOTE 

Pressing this button during the CPR period causes  

the Trainer2 to issue the GOOD COMPRESSIONS voice  

prompt.  Intructors should press this button to provide  

students with positive feedback when they respond  

to the PUSH HARDER prompt by improving the depth  

of chest compressions  

 

 
 

 

 

 

Push Harder Button 
Pressing this button during the CPR period causes  

the Trainer2 to issue the PUSH HARDER voice prompt.   

Instructors should press this button when a student's 

chest compressions are too shallow.  

 

 

SCENARIO CONTROLS 

Remote Controller: 

Action: 

VF (Ventricular Fibrillation-shock advised) is the  

default simulated heart rhythm in Manual Scenario.    

Analysis - VF 

For the next analysis to result in NO SHOCK ADVISED,  

push the No Shock button.  

NOTE 

                Once the No Shock button has been pushed, the next  

simulated heart analysis will advise NO Shock. This setting cannot be  

reversed for the next analysis. 

In all scenarios the  Pads  Button must be pressed  

once to proceed with analysis.  If pressed again, it  

will begin to prompt for attaching pads. 

 

 

Press to turn on Trainer2.

Press again to start current scenario

(as selected using Remote Controller).

Press and hold down to turn off Trainer2. 

ZOLL AED Plus

®

 

Trainer2:

Green light remains on = Trainer2 is on and ready for use. 

Green light off = Trainer2 is turned off or has no battery power. 

Red Light steady = Last command from Remote Controller accepted. 

Red Light 3 flashes = Last command from Remote Controller is invalid. 

On 
Off 

CPR Metronome On/Off Button 

Enables the CPR metronome to beep for the selected 

Training Scenario.  The LED above the button lights to 

indicate that the CPR metronome is enabled.  You should 

enable the CPR metronome whenever CPR-D padz

®

are connected to the Trainer2.

When a CPR period occurs in the Training Scenario, the 

CPR metronome begins to beep at 60 compressions 

per minute (CPM), then automatically accelerates to 

100 CPM after a few seconds.  Metronome beeps are 

independent of the student's actual compression rate

and continue until either the CPR interval ends or the

the instructor ends the scenario.

 

 

• 

Press the View/Enter Button once to turn on the  

Remote Controller. 

 

 

 

 

  

CPR Prompts 

3a 

 

PAUSE/END SCENARIO 

3b 

 

 

 

 

 

Reviews: