background image

 

 

 

Sales 

News: 

TNA0928 

       Date: 

07/29/2009 

Portable Calibration Kit (M1233SR) 

 

Why Calibrate? 

The relationship between oxygen concentration and the electromotive force of a cell is 
theoretical. Generally, a sensor shows a slight deviation from the theoretical value 
versus the actual value.  In this situation, it is necessary to conduct a calibration in order 
to ensure that an accurate calibration curve is obtained, and the deviation from the 
theoretical cell electromotive force is corrected. 
 
A gas with a know oxygen concentration is used for calibration. In this instance 
Yokogawa recommends using a Zero gas of 0.95 to 1.0% oxygen (O2) with a balance 
of nitrogen (N2). The recommended Span, or Reference Gas, should be 20.95 to 21% 
oxygen with a balance of nitrogen. This can be accomplished using clean dry instrument 
air.  The M1233SR Portable Calibration Kit provides one 105L bottle of each of these 
gases in a convenient carrying case. 
 

How to use the M1233SR Portable Calibration Kit  

 CAUTION – The contents of the bottled gases are under pressure. Please use caution when 

handling the calibration bottles, and installing the pressure regulators. 

 

 

1.  Remove the Cal (1% O

2

) gas bottle from the case. 

2.  Install the pressure regulator, with the bottle opening facing 

away from you or other personnel. Verify that the o-ring on 
the regulator is not nicked, cracked or cut prior to installation.  

 

 

3.  Attach the 3/16in clear tubing onto the regulator nipple.  

 
 
 

4.  Insert the brass ¼ in NPT fitting to the ZR22G “CAL IN” port 

on the bottom of the ZR22G gas manifold.   

    

5.  Attach the 3/16 clear tubing to the brass ¼ in NPT fitting. 

 
 
 
 
  
 
 
 
   
 
 
 
 
 
 

 
 
 
 
 
 

Reviews: