Page 2: ...logging 5 3 3 2 The menu Start type 6 3 3 3 The menu Interval 6 3 3 4 The menu Countdown 6 3 3 5 The menu Show log data 6 3 3 6 The menu Delete log data 7 3 3 7 The menu item Back 7 3 4 The menu Devi...

Page 3: ...ement There are several methods to start a measurement Start on pressing a button start after a settable count down time expired or start on movement of the PATBOX By means of a NFC connection log dat...

Page 4: ...he actual measured temperature The actual measured pressure The indicated measurement values are not stored With button you proceed to the main menu With buttons you proceed to the indication of store...

Page 5: ...art type Section 3 3 2 During logging the display will be deactivated 3 2 1 Check stop a running logging To check weather a logging is in process press or If this is the case Logging will be indicated...

Page 6: ...menu Interval In this menu you can select the time interval min 100 milliseconds max 24 hours for storing measurement values Set the desired time interval in hours minutes seconds 100 milliseconds as...

Page 7: ...em Device info After selecting this menu item the software version and serial number will be indicated 3 5 The menu item NFC enable After selecting this menu item the device will try to establish a Ne...

Page 8: ...charging The charging indication will be indicated some time after putting the PATBOX on the charging pad The date and time setting can be lost After charging these settings should be checked and corr...

Page 9: ...m Show Chart Select the serial number of the desired PATBOX if there are data of more than one device Select the desire data for indication Note If you select the desired data for about 2 seconds you...

Page 10: ...Logging interval Settable from 0 1 second to 24 hours in steps of 0 1 s Memory Flash chip Capacity 8 MB Number of records approx 600 000 Communication NFC Near Field Communication Housing Shock resis...

Page 11: ...must accept any interference received including interference that may cause un desired operation NOTE Changes or modifications made to this equipment not expressly approved by manufacturer name may v...