background image

 

Version 1.0A 

 

UM08 User’s Guide

 

DiGi Microscope

 

 

 

 

 

 

Summary of Contents for UM08

Page 1: ...Version 1 0A UM08 User s Guide DiGi Microscope...

Page 2: ......

Page 3: ...emble microscope with stand 2 1 3 IR Remote controller 4 1 4 Microscope menu functions 9 1 5 Microscope introduction 23 1 6 Microscope Focus 27 1 7 Microscope accessory 31 2 Printed Notice 32 2 1 Main...

Page 4: ...em Q ty Item Q ty DiGi Microscope 1 White Balance Card 1 HDMI Cable 1 IR Remote Control 1 Power adaptor 1 User s Guide 1 S107 Stand 1 Object Lens 4x 1 Stand assemble guide 1 Object Lens 10x Option 1 2...

Page 5: ...3 1 2 2 Start to use microscope Connect the HDMI cable b and adaptor c within the box to HDMI monitor and power See below a b c...

Page 6: ...cope Choose HDMI mode or Auto on HDMI monitor to output image 1 3 IR Remote controller The microscope operation and functions is controlled by the IR remote controller 1 Power Turn on off 2 HDMI No Fu...

Page 7: ...ity occurred during operation please reset the operation 7 Manual focus 7 1 Zoom in Zooms in speed increases when held 7 2 Zoom out Zooms out speed increases when held 7 3 Step zoom in zoom in by step...

Page 8: ...ing Refer to Section 1 4 11 Up Choose the previous item 12 Down Choose the next item 13 Left Choose the next value 14 Right Choose the last value 15 Video Ratio Digital zoom function can set x1 0 x2 0...

Page 9: ...ification The magnification size will change depending on the lens position screen size and image ratio 16 1 3 WD Working Distance 16 2 Area e 16 2 1 OZ Optical Zoom Ex OZ 500 means zoom in 500 means...

Page 10: ...r Reset 16 2 8 DZ Digital Zoom Ex DZ x4 0 means digital enlarge 4x 16 2 9 SHA Sharpness Ex SHA SET 5 means sharpness is 5 SHA SET 15 means the maximum 15 16 2 10 FRZ Image Freeze Ex FRZ ON means freez...

Page 11: ...e balance if the color seems off or incorrect 18 Image Freeze ON Off to freeze or unfreeze the image 19 Sharpness On Off to set sharpness at Minimum and Maximum level 1 4 Microscope menu functions The...

Page 12: ...ans more functions press Menu to enter next function or confirm means no more settings Return to return last page or press Left or Right to choose End to return preview window 1 4 1 EXPOSURE To set ex...

Page 13: ...AGC value is larger than 0 use to increase Senor to light s sensibility 4 Brightness Adjust image brightness 5 D WDR Digital wide dynamic range Digitally adjusting exposure in areas of the frame to ma...

Page 14: ...klight mode 1 OFF Disable backlight function 2 WDR Wide Dynamic Range WDR allows an imaging system to correct for the intense back light surrounding a subject and thus enhances the ability to distingu...

Page 15: ...ess level 3 2 AREA Choose the back light compensation area and size 3 3 DEFAULT Return to default setting 3 4 Return Return to last page or press left or right to choose End to return preview window 4...

Page 16: ...f the image is too bright 4 4 MODE Super back light compensation mode 4 5 ALL DAY Day mode 4 6 NIGHT Night mode 4 7 BLACK MASK Reverse the over exposed area 4 8 DEFAULT Return to factory default setti...

Page 17: ...ge 4 MANUAL Manually adjust blue and red values to find correct white color Refer to below fig 1 4 4 NR Noise Reduction NR To obtain high quality output image and increase file compression See below 1...

Page 18: ...r value gets higher sensitivity 3 1 2 RETURN Return to last page 3 2 LEVEL The threshold to judge noise 3 3 START AGC Start auto gain control When AGC value is higher than setting will start auto gain...

Page 19: ...few more functions in D EFFECT but 1 3 1 4 and 1 6 function can only turn on one item at one time Press Left or right to turn on or off See below functions 1 1 FREEZE Turn ON OFF to freeze image If Fr...

Page 20: ...3 1 PIP No function 1 3 2 D Zoom Set digital zoom times 1 3 3 PAN TILT To set the enlarge position 1 3 4 DEFAULT Return to default setting 1 3 5 RETURN Return to last page 1 4 SMART D ZOOM Assign 1 or...

Page 21: ...4 1 1 4 3 SENSITIVITY The sensitivity to detect whenever object moves The lower value gets the higher sensitivity 1 4 4 PIP No Function 1 4 5 D Zoom Set magnification times 1 4 6 TIME Set the enlargem...

Page 22: ...ted 4 areas are selectable 2 2 DISPLAY Display the selected area Position and area can be set in 2 1 2 3 SENSITIVITY The sensitivity to detect whenever object moves The lower value gets the higher sen...

Page 23: ...uring preview 4 2 1 LEVEL Set the threshold value to judge if it needs RETURN Return to last page 4 2 STATIC DPC Static Dead Pixel Correction scan and find the dead pixel then save the position for co...

Page 24: ...s not enough AGC will follow the threshold setting to compensate the dead pixel 4 2 5 RETURN Return to last page 4 2 2 to be corrected 5 VERSION Firmware version 6 RETURN Return to last page 1 4 6 ADJ...

Page 25: ...eft right button to choose CHANGE to return default 1 4 8 RETURN Exit from menu 1 5 Microscope introduction 1 5 1 IR Remote Area Indicator When both g and h LED indicator blinks the microscope is in s...

Page 26: ...blink whenever buttons are pressed If the indicator did not blink means the signal was not delivered and please repress again 2 HDMI indicator g When lit up the microscope is in HDMI mode If light bli...

Page 27: ...ed to connect the microscope to a screen When the device is in HDMI mode connect the microscope to the screen using the HDMI cable from the box For TV screen connections make sure to choose the homolo...

Page 28: ...tand and then screw the microscope onto the universal joint l on the gear box 1 5 4 LED Light socket 1 LED m Provides luminance for object lens 4x total 30 LEDs 2 Light socket n Provides luminance for...

Page 29: ...pe height then tuning focus by remote controller until clear On the contrary first to choose the image sensor position then adjusts the microscope height Focus distance distance from object lens to ob...

Page 30: ...ar distance control mode on the application program to focus the image 2 Fixed image sensor position The status bar will show on the screen please refer to Section 1 3 16 Use the remote controller to...

Page 31: ...have 15 focus sections The sensor position tells the magnification size Mag field of view FOV or working distance WD The below chart is measured on 4 3 ratio on 21 5 screen it s only a reference data...

Page 32: ...1 9 13 6 15 9 19 1 23 9 26 5 29 8 31 8 34 1 Focus section 10 11 12 13 14 15 16 17 18 Sensor position 1290 1250 1205 1150 1115 1080 1040 995 940 WD mm 75 0 69 5 64 0 58 5 55 5 52 5 49 5 47 0 44 0 FOV m...

Page 33: ...7 6 FOV mm 2 3 2 2 2 1 2 0 1 9 1 8 1 7 1 6 1 5 Mag X 207 216 227 238 251 265 280 298 318 Focus section 10 11 12 13 14 Sensor position 920 740 530 270 40 WD mm 7 3 7 0 6 7 6 4 6 2 FOV mm 1 4 1 3 1 2 1...

Page 34: ...lfunction if it encounters strong collision vibration or distortion 2 1 4 Turn the microscope off before cutting off the power supply do not forced cut off the power supply 2 1 5 Do not face the lens...

Page 35: ...iliary source White LED 30pcs DC port Adaptor DC Input HDMI port Output 1080P 1920 1080 Pixel image Focus control IR remote control Power supplier Adaptor Output DC 5 0V 2 0A Input AC 100 240V 50 60Hz...

Page 36: ...the device or the device is dropped or damaged please switch off and remove the power supply to avoid fire and electric shock Do not look steadily into Light sources after switching on the host as it...

Page 37: ...Confirmation Stamp is necessary for validation of the Warranty Distributor s seal shall include name of the shop telephone and address Please ask the distributor to fill in the name of the shop addres...

Page 38: ...r etc Services caused by any of the following situations in the period of the warranty shall not be provided free of charge 1 Improper use or disassembly repair or refitting 2 Any damages to the devic...

Page 39: ......

Page 40: ...HTTP WWW VITINY COM MicroLinks Technology Corp All rights reserved...

Reviews: