background image

User’s manual for 

VC99 3 6/7 DMM   

1.SUMMARY 

This  digital  multimeter  is  a  steady  performance, 

battery-driven  3  6/7  digital  multimeter.  It  uses  the  LCD 

with  23mm-high  figure  to  make  the  reading  clear  and 

make operation more convenient. 

The  digital  multimeter  has  the  function  of  measuring 

DCV,  ACV,  DCA,  ACA,  resistance,  capacitance, 

frequency, temperature, duty cycle, transistor and diode, 

continuity  performance  test  .The  meter  can  provide 

functions  including  analog  bar  and  unit  symbol  display, 

data holding, relative value measuring, maximal/minimal 

value 

measuring, 

auto/manual 

range 

switching 

(RANGE),  auto  power  off  and  warning  functions  .it 

adopts double integral A/D converter as its core. It is an 

ideal tool for labs, factories and radio-technology. 

 

2.SAFTY NOTES

 

The  instrument  is  designed  according  to  IEC1010 

standard  (safety  standard  issued  by  International 

Electro technical Committee). Please read the following 

before operation. 

1-1.Check the connection and insulation of test leads to 

avoid electric shock. 

1-2.Do not input a limited voltage higher than DC 1000V 

or AC 750V when measuring ranges. 

1-3.When  measuring  voltage  higher  than  DC  60V,  AC 

40V, please be careful. 

1-4.  Select  correct  function  and  range  to  avoid  fault 

operation. 

1-5.Please  move  the  test  leads  away  from  test  points 

when switching the function. 

1-6.Please 

don’t  input  voltage  value  when  measuring 

current. 

1-7.Please d

on’t modify the circuit. 

1-8.Introduction for safety symbol: 

 

”  exists  high  voltage;  “

”  GND  “

”  dual 

insulation 

” must refer to manual “

” Low battery. 

 

3. SPECIFICATION 

1. General features 

1-1. Displaying: LCD; 

1-2.Max  display:  6000(3  6/7)  digits  automatic  polarity, 

unit symbol and 61 section analog display; 

1-3.  Measurement  method:  double  integral  A/D 

conversion; 

1-4. Sampling rate: approx.3 times/sec. 

1-5.  Over-range  display: 

“OL”  displayed  in  the  highest 

digit.   

1-6. Low battery display

” 

1-7.  Working  environment:  (0

40)

,  relative  humidity: 

<80%; 

1-8.Store  environment:  (-10

40)

,  relative  humidity: 

<80%; 

1-9.Power supply

2pcs 1.5V battery

“AAA” battery

);

 

1-10. Dimension: 

185×93×35mm (length*width*height); 

1-11. Weight: approx..290g

(including    battery)

 

1-12.Accessories: test  leads, user manual, temperature 

probe, holster, gift box, and 2*1.5V battery. 

 

 

 

2. TECHNICAL FEATURES 

2-1.  Accuracy: 

±  (a%  ×  reading  data  +  digits), 

environment  temperature  at  (23±5)

,  relative  humidity 

<75%, One year guarantee since production date. 

2-2.DC Voltage

DCV

 

Range 

Accuracy 

Resolution 

600mV 

±(0.5%+3) 

0.1mV 

6V 

1mV 

60V 

10mV 

600V 

100mV 

1000V 

±(0.8%+3) 

1V 

Input impedance: approx

. 10MΩ 

Overload protection

1000V DC or 750V AC. 

2-3.AC Voltage

ACV

 

Range 

Accuracy 

Resolution 

6V 

±(0.8%+5) 

1mV 

60V 

10mV 

600V 

100mV 

750V 

±(1.0%+5) 

1V 

Input impedance: approx

. 10MΩ 

Overload protection

1000V DC or 750V AC. 

Frequency response:

40

200

Hz under 750V ,other 

range:(40

400

Hz. 

Sine wave RMS(Average value response). 

2-4.DC Current

DCA

 

Range 

Accuracy 

Resolution 

600uA 

±(1.0%+3) 

0.1μA 

6000uA 

1μA 

60mA 

10μA 

600mA 

100μA 

6A 

±(2.0%+5) 

1mA 

20A 

10mA 

MAX 

measurement 

voltage 

drop:

full 

range 

mA :300mV

A:100mV

 

Max input current: 20A (within 15 seconds) 

Over load protection: 500mA/ 250V and 13A/250V 

fast molten fuse. 

 