FUNCTION SETTINGS 

1.

 

Press [ON/OFF] to turn on the scale, the display will show flash digits and     

then --------,

press 

[PRINT]

, the display will show XXXX BAU value ( press 

[UNIT]

 to select among 1200,2400,4800,9600), our default setting is 9600. 

Press 

[ZERO] 

to enter Parity mode, the display will show none 

PAR

,press [UNIT] to 

select odd,even,none (odd stands for 7 data bits with odd parity, even stands for 7 

data bits with even parity,none stands for 8 data bit without parity)our default setting 

is none. 

Press 

[ZERO] 

to enter Print mode, the display will show X mod, press [UNIT]to 

select 0,1,2,3 (0 stands for serial data output disabled,1 stands for demand mode by 

press 

[PRINT],2 

stands for automatic print when scale is stable, 3 stands for continuous 

print, our default setting is 1.press [ZERO] again to return to weighing mode. 

2. Press [ON/OFF] to turn on the scale, the display will show flash digits and 

then ----------,

 press 

[UNIT]

 to enter auto off function, the display will show OFF

 

PWR

, press

 [UNIT]

 to select among 3,5,10,off ( 3 stands for 3minutes, 5 stands for 

5minutes, 10 stands for10 minutes, off stands for inactivate auto off    function ) 

press 

[ZERO]

 to select backlight, the display will show X BL ,press

 

[UNIT] to select 

value (

X

 stands for 0,1,2) , 0 stands for inactivate backlight function, 1 stands for 

the backlight is always make function. 2 satnds for auto backlight . 

Press 

[ZERO]

 to enter unit selecting function, the display will show ON X ( X stands 

for the units

  :

g,oz, lb,twt…tola)  press 

[UNIT]

 to select ON (activate the selected 

unit) or OFF ( inactivate the selected unit), press [ZERO] again to return to weighing 

mode. 

3. Switch the power to 

ON

  and than press [ON/OFF]to turn on the scale

The 

display shows F-E-9 to 0 and than 

- - - - - - -

press[PCS]key to enter ZERO track 

function setting

display shows x.xd ZER

press [UNIT] key to select ZERO track 

value

0.0d-0.5d-1.0d-1.5d-2.0d-3.0d

; press [ZERO]key to enter ZERO setting

display shows x.xd RTZ, press [UNIT] key to select ZERO value

0.5d-1.0d-2.0d-

3.0d

; press [ZERO] key to enter single segment setting or double segment setting , 

press [UNIT] key to switch between them ,SIN dsp means single segment setting 

(default setting),db 1 dsp means double segment setting; press [ZERO] key to enter 

FIL setting, press [UNIT] key to select filter rangn amond (0-1-2-3-4),”2”is default 

setting; press [ZERO] key to select zero range when turn on the scale, press [UNIT] 

key to select percent (2 PZR-5 PZR -10 PZR -20 PZR); press [ZERO] key to set zero 

range, press [UNIT] key to select percent (5 KZR or10 KZR); press [ZERO] key to 

set TARE range, press [UNIT] key select percent (50 KTR or100 KTR); press [ZERO] 

key to enter overload protection setting, press [UNIT] key to select overload 

protection value (9d OVR-1 OVR-2 OVR-10 OVR); press [ZERO] key to set beep 

function, press [UNIT] key to choose turn on or turn off the beep, “ON”means 

activate the beep, “OFF” means inactivate the beep; press [ZERO] key to unite the 

function of TARE and ZREO, press [UNIT]    to switch between them.“ON”means 

unite function activate, “OFF”means unite function inactivate. press [ZERO]key 

again to return weighing mode.. 

4. Press 

[ZERO]

 to return to weighing mode

 
 

 
 
 

 

COMPACT SCALE

/

BALANCE 

 
 

INSTRUCTION MANUAL 

Manual_FR-H_V1 ENG 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for FR-H-100

Page 1: ...ect ON activate the selected unit or OFF inactivate the selected unit press ZERO again to return to weighing mode 3 Switch the power to ON and than press ON OFF to turn on the scale The display shows F E 9 to 0 and than press PCS key to enter ZERO track function setting display shows x xd ZER press UNIT key to select ZERO track value 0 0d 0 5d 1 0d 1 5d 2 0d 3 0d press ZERO key to enter ZERO setti...

Page 2: ...Discharge the static electricity As example the method is to use the electrification prevention spray to spray it to both side of weighing platform 1 SPECIFICATIONS Model number Capacity Division FR H 100 100g 0 001g 0 005g FR H 200 200g 0 005g FR H 300 300g 0 005g 0 01g FR H 600 600g 0 01g 0 005g FR H 1000 1000g 0 01g 0 02g FR H 3000 3000g 0 05g 0 1g FR H 6000 6000g 0 1g 0 2g FR H 10000 10000g 0 ...

Page 3: ...ch can be setted according to your requirement 3 Calibration by the weight When you selected the calibration weight then place the calibration weight on platform Then press ZERO key the display return to the weight value Now the calibration is complete 4 Press ON OFF turn off the power press ON OFF turn on the power again place the weight on the platform to make sure whether weighing is correct if...

Page 4: ... accurately Error messages EEEE Overload Low battery Calibration may be required before weighing Read CALIBRATION first and if necessary calibrate your scale balance for accurate weighing Weighing procedures 1 Press ON OFF to turn on the balance When power is turned on all display segments appear on LCD for a few seconds and finally 0 will appear on the display 2 Select the weighing unit with UNIT...

Page 5: ...Bit 8 7 7 Parity None Even Odd Stop Bit 1 Code ASCII Data Format W WEIGHT M MESSAGE 1 2 3 4 5 6 7 8 9 1 0 1 1 1 2 1 3 1 4 1 5 1 6 1 7 1 8 1 9 2 0 HEA D DATA UNIT CR W M g C R L F Example 100 00g W 100 00g 500 00ct W 500 00ct Err 0 M over ...