background image

VH-3000

Respiratory Humidifier

2460

Manual

Read instruction carefully before use

Made In Taiwan

VA-1-VH-3000-EN.MDR

2021.12.23

Summary of Contents for VH-3000

Page 1: ...VH 3000 Respiratory Humidi er 2460 Manual Read instruction carefully before use Made In Taiwan VA 1 VH 3000 EN MDR 2021 12 23...

Page 2: ...Unlock 7 5 2 Standby 8 5 3 Heated Wire Mode 9 5 4 Non Heated Wire Mode 10 5 5 Silence 11 5 6 Power Switch 11 5 7 NO FLO 12 5 8 Hi Lo Alarm 12 5 9 Temperature Set 13 6 Installation 14 15 7 Use Instruc...

Page 3: ...oad Zhongshan Village Yangmei District Taoyuan City 32665 Taiwan TEL 886 3 475 9911 E mail sales vadimedical com tw FAX 886 3 475 9922 Website www vadimedical com tw MedNet EC REP IIb GmbH Borkstrasse...

Page 4: ...ower automatically The VH 3000 life cycle is 3 years Operating Environment Temperature 18 26 Humidity RH 15 95 Pressure 700 to 1060hPa Storage Temperature 25 40 Humidity RH 15 95 without condensation...

Page 5: ...Applied Part inhalation tube 2460 Caution Hot surface may exceed 75 Drip proof Alarm Caution Electrical Shock Hazard Refer to qualified service Personnel CE Mark Power On Off Standby Serial number UDI...

Page 6: ...e equipment It may a ect its performance 6 Ensure the temperature probe sensors correctly and securely tted Failure to do so may result in gas temperature in excess of 41 being delivered to the patien...

Page 7: ...Clean the parts with low ow running water using a soft brush to remove all foreign material for example soil or organic matter 3 Sterilization by autoclave 132 4minutes Chamber 1 Clean the parts with...

Page 8: ...4 Control Panel Airway Temp Set Airway Temp Stand by Heated Wire Mode Silence Power Switch Heated Wire Adaptor Temperature Probe Chamber Temp Active Alarm Unlock Chamber Temp Set 6...

Page 9: ...midi er The unlock button will automatically resume if no action is taken within 10 sec The following buttons are locked if the user doesn t activate the Unlock button 1 Standby 2 Heated Wire Non Heat...

Page 10: ...ated wire are kept for 15 minutes Push the button again to disable the function Notes 1 In standby mode it will not show No FLo 2 The humidi er will resume the period of standby mode if the power is p...

Page 11: ...n Heated Wire Mode The user can set the airway and chamber temp via adjustment of or button EX If airway temp set is 39 and chamber temp set is 2 the humidi er maintains 39 at airway and 37 at chamber...

Page 12: ...e the LED o the function is activated The user use or button on airway temp set to increase or decrease the Power level Power Level Heater Plate Power Level Heater plate Temperature 1 2 3 4 5 6 7 8 9...

Page 13: ...The button is used to turn o audio alarm for 120 sec When the airway temp exceeds 42 Silence is limited to 30 sec 5 6 Power Switch It is on o switch The LED is on while the humidi er is activated LED...

Page 14: ...increase in 30 Seconds the LED displays no Flo 5 8 Hi Lo Alarm If it is lower than 10 it displays Lo If it is higher than 70 it displays Hi Airway temp It is the temperature of Airway Airway temp Cha...

Page 15: ...The LED on Airway Temp Set should be kept the same as the previous one while being switched on The user can adjust it by and from 31 to 40 B Chamber Temp Set The user can adjust it by and from 5 to 2...

Page 16: ...Fig 1 6 2 Slide chamber onto the heater base Fig 2 Note Choose the appropriate chamber 6 3 Add distilled water into chamber Fig 3 or stab distilled water with the needle of IV SET and connect the tube...

Page 17: ...ber above the maximum ll level line Liquid could enter the breathing circuit if the chamber is over lled 2 When mounting a humidi er adjacent to patient ensure that the humidi er is always positioned...

Page 18: ...ature Probe VH 3000 Humidi er Airway Probe Fig 1 Heated Wire Breathing circuit System Ventilator Chamber VH 3000 Humidi er Fig 2 Non heated wire breathing circuit system 7 Use Instruction 7 1 Be sure...

Page 19: ...5 Set the temperature level refer to 5 1 5 3 5 9 7 6 If a heated wire circuit is being used check that the green indicator LED next to the button labeled heated wire is illuminated If not press the b...

Page 20: ...2 above or 4 lower than the temperature control setting The Airway Temp LED will ash and audible alarm 8 2 The alarm tolerance of Chamber Temp is above or lower than 4 the temperature control setting...

Page 21: ...8 3 If the exclamation ash and audible alarm please contact your local VADI distributor 8 4 The Temperature probe disconnect or malfunction the LED will ash and audible alarm 19...

Page 22: ...e temperature probe and the water in the circuits The alarm tolerance of Chamber Temp is above or lower than 4 the temperature control setting The Chamber Temp LED will ash and audible alarm Please ch...

Page 23: ...of the disposition of electronic device accessory waste 11 Clean The heated plate may be wiped with a damp cloth with using clean water or 75 ethyl alcohol 12 Maintenance Check the power cable for dam...

Page 24: ...ce is used with the consumables which are not manufactured by VADI Medical Technology Co Ltd Yangmei 4 Use any blanket to cover the breathing circuit is not allowed 5 NO MODIFICATION of this humidi er...

Page 25: ...adaptor inspiratory heated B VH30100L Power adaptor inspiratory heated C VH30200 Power adaptor dual heated D VH30200L Power adaptor dual heated E VH30C Connector 2P 3P F VH30T150 Temperature Probe 150...

Page 26: ...with port 110cm I 900130L Heated wire with port 130cm J 900150L Heated wire with port 150cm K 900110LE Heated wire without port 110cm L 900130LE Heated wire without port 130cm M 900150LE Heated wire...

Page 27: ...214000 Disposable adult circuit 150cm 22mmID dual heated w o chamber B G 215000 Disposable adult circuit 150cm 22mmID inspiratory heated w o chamber C G 216000 Disposable adult circuit 150cm 22mmID du...

Page 28: ...osable adult circuit 150cm 22mmID inspiratory heated w chamber E G 218000 Disposable infant circuit 150cm 12mmID dual heated w chamber F G 219000 Disposable infant circuit 150cm 12mmID inspiratory hea...

Page 29: ...iption Accessories Absorbent paper for G 314005 Disposable humidi er chamber Disposable humidi er chamber Disposable humidi er chamber Disposable humidi er chamber Disposable humidi er chamber Reusabl...

Reviews: