background image

      

 

February 2022

 

C6057701-01

 

Q-Series Quick Guide 

Please download the full instruction manual from the Ultrawave website 

Electrical Safety instructions 

The Q-Series Ultrasonic Baths are designed to be connected to a 230VAC fully 

earthed electrical supply via a three-pin plug. 
It  can  be  dangerous  to  operate  the  Ultrasonic  Bath  without  an  earth 

connected. 
The Q-Series machines are rated at up to 1050W maximum. 

The  detachable  mains  lead  is  rated  at  250V  6A. 

NEVER  FIT  A  LEAD  WITH  A 

LOWER RATING.

 

The  mains  plug  is  fitted  with  a  5A  fuse. 

NEVER  FIT  A  FUSE  OF  A  HIGHER 

RATING. 

The IEC electrical socket on the rear of the unit is fitted with 2x T5A (time-

lag) fuses. 

NEVER FIT A FUSE OF A HIGHER RATING.

 

 

The  detachable  mains  lead  is  the  disconnect  device  and  should  remain 

accessible while the machine is in use.

 

Ensure that any excess mains cable is stored neatly. 

General Safety Instructions 

Do not place fingers, hands, or other parts of the body in the liquid in the bath 

when it is sonicating.

 

Always ensure the liquid is above the level sensor when in operation. 
Do not place parts or containers directly on the bottom of the cleaning tank. 

Always  use  a  basket  or instrument  cassette  to  support  items  to  be  cleaned. 
Not doing so may damage the bath and invalidate your warranty. 
Never use toxic, flammable, acidic, caustic, or corrosive chemicals or liquids 

not recommended by Ultrawave. 
Ensure the electrical supply is disconnected before moving/emptying. 
Before  cleaning,  drain  the  bath.  Clean  by  wiping  with  a  clean  non-abrasive 

cloth. 
The user should familiarise themselves with the 

full instruction manual

 before 

operating the equipment. Safety may be impaired if they are not followed. 
Ultrawave will not be responsible for damage or injury caused by incorrect use 
of the equipment. 
The  user  should  apply  to  Ultrawave  or  its  agent  for  advice  on  cleaning 
techniques or detergents. 

 

 