background image

www.tsi.com 

AeroTrak

®

 

Handheld Airborne 
Particle Counter 

Model 9306 

Operation Manual 

P/N 6004215, Revision M 
September 2022 

Summary of Contents for AEROTRAK 9306

Page 1: ...www tsi com AeroTrak Handheld Airborne Particle Counter Model 9306 Operation Manual P N 6004215 Revision M September 2022 ...

Page 2: ...tes product enhancements and new products By registering your instrument TSI will be able to send this important information to you http register tsi com As part of the registration process you will be asked for your comments on TSI products and services TSI s customer feedback program gives customers like you a way to tell us how we are doing ...

Page 3: ...l SHIP TO MAIL TO TSI Incorporated 500 Cardigan Road Shoreview MN 55126 3996 USA U S Technical Support 800 680 1220 651 490 2860 Fax 651 490 3824 E mail address answers tsi com Website http www tsi com INTERNATIONAL Technical Support 001 651 490 2860 Fax 001 651 490 3824 ...

Page 4: ...ld Airborne Particle Counter Model 9306 Operation Manual P N 6004215 Revision Date Revision Date A July 2010 G January 2014 B September 2010 H February 2016 C February 2011 J August 2016 D March 2012 K March 2018 E November 2012 L December 2019 F October 2013 M September 2022 ...

Page 5: ...recommended cleaning g This warranty is VOID if the product has been misused neglected subjected to accidental or intentional damage or is not properly installed maintained or cleaned according to the requirements of the manual Unless specifically authorized in a separate writing by Seller Seller makes no warranty with respect to and shall have no liability in connection with goods which are incor...

Page 6: ...ustomers our service policy is designed to give prompt attention to any problems If any mal function is discovered please contact your nearest sales office or representative or call TSI s Customer Service department at 1 800 680 1220 USA or 001 651 490 2860 International Trademarks TSI and the TSI logo are registered trademarks of TSI Incorporated in the United States and may be protected under ot...

Page 7: ...m Ion Battery 2 3 To Use AC Power 2 4 Using with a Printer 2 4 Installing an Isokinetic Inlet 2 5 Installing a Temperature Relative Humidity Probe 2 5 CHAPTER 3 Operation 3 1 Screen Layout and Functionality 3 2 Software Input Panel Keyboard or Keypad 3 2 Main Tab 3 3 Zoomed Data Screen 3 7 Setup Tab 3 8 Zone Setup Screen 3 9 Recipes Setup Screen 3 14 System Setup Screen 3 19 Change Power On Passwo...

Page 8: ...oftware 4 2 Ethernet Communications 4 2 CHAPTER 5 Maintenance 5 1 Maintenance Schedule 5 1 Zero Check 5 1 Cleaning the Instrument Enclosure 5 1 CHAPTER 6 Troubleshooting 6 1 CHAPTER 7 Contacting Customer Service 7 1 Technical Contacts 7 1 International Contacts 7 1 Service 7 1 Technical Support 7 2 Returning the AeroTrak Handheld Airborne Particle Counter for Service 7 3 APPENDIX A Specifications ...

Page 9: ...held Airborne Particle Counter is a Class I laser based instrument During normal operation you WILL NOT be exposed to laser radiation Precaution should be taken to avoid exposure to hazardous radiation in the form of intense focused visible light Exposure to this light may cause blindness Take these precautions DO NOT remove any parts from the particle counter unless you are specifically told to d...

Page 10: ...ached to the outside of the particle counter housing and to the optics housing on the inside of the instrument 1 Serial number label back panel 2 Laser radiation label internal 3 Calibration Label side panel 4 Laser radiation symbol label internal 5 European symbol for non disposable item Item must be recycled ...

Page 11: ...sult in serious injury to you or cause damage to the instrument Follow the procedures prescribed Caution or Warning Symbols The following symbols may accompany cautions and warnings to indicate the nature and consequences of hazards Warns that uninsulated voltage within the instrument may have sufficient magnitude to cause electric shock Therefore it is dangerous to make contact with any part insi...

Page 12: ...r to submit suggestions please contact Customer Service TSI Incorporated 500 Cardigan Road Shoreview MN 55126 U S A Fax 651 490 3824 USA Fax 001 651 490 3824 International Telephone 1 800 680 1220 USA or 651 490 2860 International 001 651 490 2811 E mail Address technical services tsi com Web site www tsi com ...

Page 13: ...00 data sets can be downloaded for analysis and reporting using the TrakPro Lite Secure Data Download Software included with the device Model Size Range 9306 03 0 3 0 5 0 7 1 0 2 0 5 0 μm 9306 04 0 3 0 5 1 0 3 0 5 0 10 0 μm 9306 V2 0 3 to 10 μm user selectable factory calibrated at 0 3 0 5 1 0 3 0 5 0 10 0 μm Typical applications for this particle counter include clean room monitoring research exp...

Page 14: ... TSI immediately if items are missing or broken Model 9306 AeroTrak Airborne Particle Counter Parts List Qty Item Description Part Model Reference Picture 1 AeroTrak Airborne Particle Counter 9306 03 9306 04 9306 V2 1 Power Supply with universal plugs 801694 1 Isokinetic inlet 700003 AL 1 Battery pack 700032 1 Computer cable 2 m USB A to B 700033 1 Stylus N A 1 HEPA zero filter assembly 700005 1 T...

Page 15: ...Introduction and Unpacking 1 3 Qty Item Description Part Model Reference Picture 1 Operation Manual 6004215 Available on TSI website tsi com 1 Quick Start Guide 6004216 1 Calibration certificate N A ...

Page 16: ...6 AeroTrak Airborne Particle Counter Optional Accessories Item Description Part Model Reference Picture External battery charger with AC adapter and power cord 700025 External Printer 700085 Carry case 700083 Temperature humidity probe 700084 Stainless Steel Isokinetic inlet 700004 Isokinetic probe used with tubing 700001 AL 700002 SS 0 1 cfm Barb Inlet Fitting 700020 Tubing Superthane 1 8 inch ID...

Page 17: ...e Instrument Stand and Stylus Providing Power Installing an Isokinetic Inlet Installing a Temperature Relative Humidity Probe I n s t r u m e n t D e s c r i p t i o n The Model 9306 has many features to make measurements convenient They are described in detail below Isokinetic Inlet Touch Screen USB Host and Device Ports Stand Power Button Start Sample Stylus Power Entry Ethernet Port ...

Page 18: ...and grasp it by the large finger hole and pull it out until it locks into place Be careful not to overextend the stand To store the stand out of the way when not in use simply push the stand back until it snaps into place The Model 9306 is also equipped with a plastic stylus for use with the touch screen interface The stylus locks into place in the case near the top of the unit when not in use ...

Page 19: ... of data To Install the Lithium Ion Battery 1 Remove the battery cover from the back of the instrument by lightly depressing the textured tab on the cover located on the lower left 2 Place the lithium ion battery into the battery compartment and slide it forward toward the top of the unit until it locks into place 3 Replace the battery cover and slide it in place until you hear a click W A R N I N...

Page 20: ... TSI logo a brief start up sequence begins as the Windows CE operating system boots up U s i n g w i t h a P r i n t e r A hard copy of a sample set can be printed from the instrument using the optional TSI Model 700085 thermal printer see Chapter 3 Operation Only the TSI Model 700085 printer is compatible with the Model 9306 The printer may be used on its internal battery or an AC adapter A custo...

Page 21: ... m p e r a t u r e R e l a t i v e H u m i d i t y P r o b e To install the optional temperature relative humidity probe 1 Align the probe so the pins slide into the sockets of the base connector 2 Align the locking collar on the probe so it will slide over the alignment pins on the base connector 3 Press down and turn the locking collar to lock in the probe 4 Temperature and relative humidity are...

Page 22: ...2 6 AeroTrak Handheld Airborne Particle Counter This page intentionally left blank ...

Page 23: ...damage the screen overlay To turn on the instrument press the on off button located in the center of the front of the instrument After a splash screen displays the TSI logo a brief start up sequence begins as the Windows CE operating system boots up The instrument is ready for operation when the Main tab shown below appears If an optional temperature humidity probe is attached those values will be...

Page 24: ...on To enter information tap on the screen and an on screen keyboard appears Software Input Panel Keyboard or Keypad 1 Throughout the setup screens a keyboard or keypad appears on the screen when text or numbers may be entered 2 When you enter information using the keyboard press either the Enter or Esc keys when you are done When you enter data using the keypad the data is entered when you press O...

Page 25: ... instrument NOTICE During Start Delay Delay and Hold Times Hold this is only an indicator of flow On During Sample Time Time this is an indicator of flow within specified tolerances Insufficient flow through the instrument If this icon is shown it can be pressed and a more detailed description of the flow error will be shown Refer to the trouble shooting section for appropriate actions If a flow e...

Page 26: ...th the AeroTrak particle counter The front panel GUI is inoperable when the software is operational Once the software is exited normal front panel operation will resume Press and hold the Zone icon to display a summary of information for the current Zone Tap the Location icon to step through the list of Locations for the Zone ...

Page 27: ...en immediately after the Start button is pressed the Delay time begins a countdown 2 During sampling and between cycles after the Start Delay has been displayed the Hold Delay is displayed and then begins a countdown Recs The total number of records in the database 10000 maximum number of records Manual Automatic Beep Mode Indicator refers to the Data Count Mode see section below Start Stop button...

Page 28: ...nName i e the zone name followed by a colon followed by the location name The barcode can then be scanned and the specified zone and location will be automatically selected in the list boxes There are free online barcode generators available i e http barcode tec it com en Select linear Code 128 to generate the barcodes Zebra and the stylized Zebra head are trademarks of Zebra Technology Corp regis...

Page 29: ... display to switch back to the Main tab display The display shows Temperature Relative humidity Bin sizes Particle count concentration Field Description Location Label that displays information about the currently selected location Manual Automatic Beep Mode Indicator refers to the Data Count Mode see section below Press the Start Stop button to begin sampling in the configured mode Temperature an...

Page 30: ...vidual settings Environment Setup Sets which units are preferred for displaying environmental measurements taken using optional measurement probes System Setup Change Power On Password Setup Password System Configuration Print Settings Print Schedule and Clear Samples Information View instrument information such as firmware version model serial number last calibration date USB IP Address etc Devic...

Page 31: ...use for the samples taken in the Zone Options include ISO 14644 1 EU GMP Fed Std 209E F Fed Std 209E and None Use None for taking measurements that are not associated with standards classification Class The Class selected for the classification of the Zone Options vary by the Standard selected Status The occupancy status of the Zone Options vary based on the selected standard but include At Rest O...

Page 32: ...with it cannot be deleted The data associated with the zone must be deleted from the instrument before the zone can be deleted To Add a Zone To add a zone press Add The Definition screen is displayed 1 Enter a name for the zone and select the Standard Class Status Airflow and Largest Particle Size to Consider options from the drop down boxes Input the Area using the keypad and select either ft2 or...

Page 33: ...meters of the recipe tabs Recipe Timing Channels T and Channels V NOTICES You can also create recipes from the Setup tab by selecting the Recipe icon but if you create a new recipe here information you have already entered for the zone is prepopulated into the required fields If you edit an existing recipe your changes will affect all zones using that recipe Be certain that is what you want to do ...

Page 34: ...ion from the zone You can also move the location name up or down in the list The Auto feature will generate the number of locations required by the chosen standard This utilizes the room area entered in the zone definition The locations can then be renamed or the default naming convention can be maintained 5 You can take as many samples as possible up to 10 000 under a specific location attached t...

Page 35: ...can select a different recipe for the zone or you can create a new recipe or edit an existing recipe For information about recipes see the Recipes section in Setup NOTICE If you edit a recipe the changes will affect all zones using that recipe Be certain that is what you want to do 7 When you have made all the changes press Save ...

Page 36: ...d NOTICE The Delete button will be grayed out and unavailable if the recipe has samples The steps for adding or editing a recipe are identical Press either the Add button or the Edit button and proceed as follows 1 Press Add or Edit button to display the Recipe tab 2 On the Recipe tab enter a name or edit the name of the recipe For a new recipe a default name will appear but you can type over it a...

Page 37: ... To alarm on viable particulates be sure that all total particulate alarms are disabled and configure an alarm in the viable count channels It is configured in the same manner described above If multiple alarms are configured the AeroTrak particle counter will emit beeps only on a single channel The AeroTrak particle counter searches for the alarm to trigger starting with the smallest to largest c...

Page 38: ...times are 10 seconds or less Flow error alarms may not occur is sample time is less than 10 seconds To ensure proper instrument status and flow alarm operation use a sample time of 15 seconds or larger NOTICE With firmware version 2 4 and higher you may select a date to print reports and or export sample data There is no need to create new zones to select data by date There is no need to print all...

Page 39: ...e until the Start Stop button is pressed again Volume Volume sets a limit of air volume to collect for each sample Select the volume unit then enter a volume value The Sample Time will be updated automatically to the nearest second adequate to provide the desired volume 7 Press the Channels tab 8 This tab can be used to view or set the particle size for each channel not supported in all models ena...

Page 40: ...ial Δ particle count or concentration is selected the total number of counts is the number of particles between enabled bin sizes When particle concentration is cumulative Σ the total number of counts includes all particles larger than the bin size In Differential Display Alarm mode there are two constraints If alarming is desired all channels must be enabled If channel selection is desired then a...

Page 41: ...e instrument will operate If a Power On password has been previously set that password must be entered before choosing a new Power On password A blank in the password field is considered no password so if set as a new password the instrument will not prompt for a password on system startup NOTICE Keep the password in a safe place It is difficult to reset the password and requires contacting the fa...

Page 42: ...g password if one has already been set or leave blank New Password Enter a new password The password can be any length and use any characters Confirm New Password Retype the new password then press OK A confirmation message appears if the password is changed NOTICE Leave both New Password and Confirm New Password fields blank to turn off password protection Contact TSI if the password is forgotten...

Page 43: ... You can still use the existing zones locations recipes create reports and transfer data You would need the password to change any of these functions Tap the Setup Password icon to display the on screen keyboard and enter the required information Field Description Old Password Enter your existing password if one has already been set or leave blank New Password Enter a new password The password can...

Page 44: ...m enlarge both cumulative Σ and differential Δ counts allow the screen a moment to update Tap on the screen again to return to normal view Store Partial Samples When selected if the instrument stops during a sampling period the database will record the partial sample data Alarm Volume Level Controls the alarm volume setting Particle Density Enter the particle density value that will be used to cal...

Page 45: ...particle counter used to collect the data will be printed Separator Indicates a line separator will be printed after the Model Name and Serial Number in the header of all printouts Differential Indicates that the differential value of the data will be printed Cumulative Indicates that the cumulative value of the data will be printed Last Calibration The date and time the instrument was last calibr...

Page 46: ... in English even when the selected language is not English The default zones and locations will print in English If zones or locations have been created with names in other languages the names are not translated to English Field Description Automatic Printing Enables automatic printing when checked On Sample Print data whenever a sample completes On Alarm Print data when an alarm condition occurs ...

Page 47: ...ELECTED ON THE CLEAR SAMPLES SCREEN ALL SAMPLE RECORDS WILL BE DELETED FROM THE INSTRUMENT THERE IS NO WAY TO RECOVER THEM ONCE THEY HAVE BEEN DELETED Information Screen Use the Information screen to view the following on the system Model Number Serial Number Copyright Manufacture Date Last Calibrated Calibration Due Firmware Version Ethernet IP Address USB IP Address Date Format Time Format Press...

Page 48: ...and Time screen to set the current date time and the time format Press OK when finished Select options using the arrows or tapping on the screen which displays the keypad Field Description Date Press the down arrow to display a calendar then select the date from the calendar Time Select the time component you want to change hours minutes seconds and then use the left and right arrows to adjust to ...

Page 49: ...ignment Press this item to reset the screen alignment Follow the directions on the alignment screen NOTICE The touch screen display is aligned at the factory and typically will stay aligned for the life of the instrument Only perform this alignment if tapping on the on screen controls seems to give poor results ...

Page 50: ... in one hop time to live 1 implying that all hosts in a subnet are connected to the same link Default Gateway A node on the computer network that serves as an access point to another network and is chosen when the IP address does not belong to any other entities in the Routing Table Use DHCP Dynamic Host Configuration Protocol When checked this protocol is used to automatically obtain the informat...

Page 51: ...alogs and the regional format for numbers Field Description Language Select the language for the display options are German English Spanish French Italian Chinese simplified and Japanese Formats Select the format that is commonly used to display real numbers and the date and time in your region ...

Page 52: ...splay on the Main and Data Tabs and in the printouts Field Description F Display temperature in degrees Fahrenheit C Display temperature in degrees Celsius Data Tab Use the Data tab to preview data that has been collected To scroll though the records use the elevator slide on the right The record number is displayed at the bottom of the tab As each record displays its data and relevant parameters ...

Page 53: ...micrograms per cubic meter mass concentration Size Channel size Differential mode Cumulative mode Export the data to a flash drive See Export Data Screen below Print data to the optional printer See Print Data below Zone Z Zone where the data was collected Location L Location where the data was collected Sample Duration of the sampling period Date Date on which the data was collected Time Time at ...

Page 54: ...ed file This file is intended to be used with TrakPro Lite Secure software and maintains CFR 21 Part 11 compliance The file has the extension file name_sec xml Unsecured file This file is intended for user input into Microsoft Excel for graphing and data manipulation purposes and has the extension file name xml Both If using both file types both file formats can be exported Please note that the da...

Page 55: ...t OK 3 Select Sample Data by Zone or by Sample index range with option of limiting samples to a date range a Check Use Start Date checkbox and select a start date to exclude data that was collected prior to that start date b Check Use End Date checkbox and select an end date to exclude data that was collected after that end date Date is displayed as month day year ...

Page 56: ...radio button and then select the lower and upper sample index numbers To select data for export by zone instead tap the Select by Zone button at the bottom 6 Once the records or the Zones have been selected press Export to begin exporting Status screens allow viewing the progress of the export or C A U T I O N DO NOT remove the external drive during the export process If the thumb drive is removed...

Page 57: ...mit data selection to a date range Check the Use Start Date checkbox and select a start date to exclude samples collected before that date Check the Use End Date checkbox and select an end date to exclude samples collected after that date Check the Recipe checkbox to include recipe details for each recipe used in the report To Print Data 1 Click the Printer icon on the Data tab The Print by Zone s...

Page 58: ...The Select by Record screen above allows you to select data by range of sample index number a Tap the Select Range radio button and select the lower and upper sample index numbers b To select data by zone instead tap the Select by Zone button at the bottom 6 To filter the data by start and or end date check the appropriate checkboxes and select the date 7 Check Recipe to include the details of eac...

Page 59: ...sample and Y is the total number of samples in a set being printed not the total number of samples on the instrument Reports Tab Use the Reports tab screen to select various standard reports for viewing and printing 1 Select a zone to be included in the report You have the option to restrict data to a date range a Check the Use Start Date checkbox and enter a start date to exclude data collected b...

Page 60: ...lude recipe details on the report 3 When you press the Generate button you are given the option to enter a title for the report 4 Enter a title or leave the text box blank 5 Press the Enter key on the visual keyboard or press the Close button to close the form and generate the report ...

Page 61: ...specified date Use End Date Exclude samples collected after a specified date Recipe Check to include the recipe details on the report Exclude Samples button Press Exclude Samples button to display Exclude Samples form which allows you to exclude samples from report if allowed by selected standard IDs of samples you exclude on that form will be displayed in list box to right of this button Generate...

Page 62: ... in right column You can exclude up to five samples and replace them with other samples for ISO 14644 1 2015 If the standard allows you to exclude only one sample e g ISO 14644 1 1999 only the first text box in the left column will be enabled and the replacement text boxes will all be disabled 2 Press the Save button to exclude the specified samples from the report 3 Press the Clear button to eras...

Page 63: ...conjunction with TrakPro Lite Secure software When enabling data integrity mode on the instrument only the data can be viewed The lock icon on the main page indicates the unit is in data integrity mode but no user is logged in To allow sampling the unit will need to be unlocked using TrakPro Lite Secure software During this process TrakPro Lite Secure software will place the user ID and password i...

Page 64: ...tion is first initiated the instrument will prompt for a password NOTICE The password is the same password used when you logged into TrakPro Lite Secure software If an incorrect password is entered you will not be able to start a sample A password needs to be entered when the unit first starts or when the first sample is taken after disconnecting from TrakPro Lite Secure software or after a timeou...

Page 65: ...the instrument defaults to no zone and no location the start sample icon will no longer be available A zone and location will need to be selected to enable the start sample icon NOTICE In rare events a new sample will be started that requires the instrument to overwrite existing sample data The screen shown at right will display and the instrument will not start the sample ...

Page 66: ...3 44 AeroTrak Handheld Airborne Particle Counter This page intentionally left blank ...

Page 67: ...ollow the instructions in the operation section of this manual The data is downloaded in XML format that can be opened in Microsoft Excel spreadsheet program version 2003 or greater The data files can also be opened in the latest versions of OpenOffice application U S B C o m p u t e r C o m m u n i c a t i o n The Model 9306 AeroTrak Handheld Airborne Particle Counter is equipped with a USB compa...

Page 68: ...3 See the TrakPro Lite Secure version 3 0 or later Software User s Guide P N 6004404 for installation instructions N O T I C E Make sure the particle counter is connected before you run the software E t h e r n e t C o m m u n i c a t i o n s An Ethernet port is provided for use with TSI Facility Monitoring Software FMS Refer to the FMS Software documentation and the TSI service and installation m...

Page 69: ... TSI recommends annual factory cleaning and calibration for the AeroTrak Airborne Particle Counter See Chapter 7 Contacting Customer Service for service calibration Recommended Field Maintenance Schedule Item Frequency Zero check Daily or according to application Factory cleaning and calibration Annually Cleaning the instrument enclosure As needed Z e r o C h e c k The zero check ensures that the ...

Page 70: ...5 2 AeroTrak Handheld Airborne Particle Counter This page intentionally left blank ...

Page 71: ...o count specification 1 particle 5 mins Particles are in the instrument chamber Run the instrument for one half hour with a filter and then recheck the zero count HEPA filter is not connected properly and room air is leaking into the HEPA filter assembly Check that the HEPA filter has been tightly connected to the inlet Check that rubber O ring black on the inlet is in place Residual particles fro...

Page 72: ...al support TEMPERATURE HUMIDITY PROBE ERROR Temperature RH probe was not recognized Detach and reconnect probe If problem persists contact TSI technical support FLOW ERROR Instrument was unable to control flow rate if any tubing is connected to particle counter Restart measurement Pressure drop across inlet may be too large Lessen pressure drop across inlet by using larger diameter tubing less tub...

Page 73: ...1 800 680 1220 USA or 651 490 2860 or e mail technical services tsi com If the AeroTrak Model 9306 does not operate properly or if you are returning the instrument for service visit our website at tsi com service or contact TSI Customer Service at 1 800 680 1220 USA or 651 490 2860 International Contacts Service TSI Instruments Singapore Pte Ltd 150 Kampong Ampat 05 05 KA Centre Singapore 368324 T...

Page 74: ...jing Co Ltd Unit 1201 Pan Pacific Plaza No 12 A Zhongguancun South Avenue Haidian District Beijing 100181 CHINA Telephone 86 10 8219 7688 Fax 86 10 8219 7699 E mail tsibeijing tsi com TSI GmbH Neuköllner Strasse 4 52068 Aachen GERMANY Telephone 49 241 52303 0 E mail tsigmbh tsi com TSI Instruments Ltd Stirling Road Cressex Business Park High Wycombe Buckinghamshire HP12 3ST UNITED KINGDOM Telephon...

Page 75: ...return instructions Customer Service will need this information when you call The instrument model number The instrument serial number A purchase order number unless under warranty A billing address A shipping address Use the original packing material to return the instrument to TSI If you no longer have the original packing material seal off any ports to prevent debris from entering the instrumen...

Page 76: ...7 4 AeroTrak Handheld Airborne Particle Counter This page intentionally left blank ...

Page 77: ...ration system Calibration Frequency Recommended minimum once per year Sample Probe Tubing Isokinetic sampling probe Sampling Modes Manual automatic beep cumulative differential count or concentration Sampling Time 1 second to 99 hours Sampling Frequency 1 to 9999 cycles or continuous Exhaust Internally filtered Vacuum Source Internal pump Communication Mode Modbus TCP over Ethernet or USB Data Sto...

Page 78: ...ns 41 to 95 F 5 C to 35 C 20 to 95 non condensing relative humidity Storage Conditions 32 to 122 F 0 C to 50 C Up to 98 non condensing relative humidity Included Accessories Power supply power cord battery isokinetic inlet stylus purge filter TrakPro Lite data download software operational manual available on tsi com website computer cable calibration certificate and Quick Start Guide Optional Acc...

Page 79: ...61000 3 3 Voltage Fluctuations EN61000 4 2 Electrostatic Discharge Immunity EN61000 4 3 Electromagnetic Field Immunity EN61000 4 4 Burst Immunity EN61000 4 6 Conducted PS Immunity EN61000 4 5 Surge Immunity EN61000 4 8 Rated Power Frequency Field Immunity EN61000 4 11 Voltage Dips Short Interruptions Immunity RoHS Marking Yes D i m e n s i o n a l D i a g r a m ...

Page 80: ...A 4 AeroTrak Handheld Airborne Particle Counter This page intentionally left blank ...

Page 81: ......

Page 82: ... more information USA Tel 1 800 680 1220 UK Tel 44 149 4 459200 France Tel 33 1 41 19 21 99 Germany Tel 49 241 523030 India Tel 91 80 67877200 China Tel 86 10 8219 7688 Singapore Tel 65 6595 6388 P N 6004215 Rev M 2022 TSI Incorporated Printed in U S A ...

Reviews: