11/2012, JK 

 

 

USER INSTRUCTIONS - TP7C & TP9 

    

 
 

 
INTRODUCTION 
 

This manual describes the basic function, use and safety instructions for a model TP7C or TP9 portable digital thermometer 
instrument. These ThermoProbe instruments are intended for use in both hazardous (flammable) and non-hazardous areas under dry 
conditions at ambient temperatures between -20 to 40

°

C. The instruments are not intended for use in permanent outdoor installations 

and are not intended or tested for icing conditions.

 

Additional means of protection should be used where the equipment may be 

exposed to excessive external stresses (e.g. vibration, heat, impact, etc.).

 

 
 

SAFETY INFORMATION BEFORE USE 
 

ThermoProbe thermometers are designed for safe operation in hazardous locations (Potentially Flammable or Explosive). The user 
must have a working knowledge of appropriate safety instructions.

 

 

a) The inspector must have a thorough knowledge of the products to be measured and must know of the safety precautions to be 
taken when working with the material to be measured. 
b) The instrument shall be checked concerning severe defects; check that instrument is complete (including grounding/bonding 
cable), has good batteries, etc. If necessary, check measurement accuracy. If any defects are found, the instrument should not be used 
until repairs have been made. 
c) The instrument, especially cable and probe, should be clean both for safety reasons and ease of use. 
d) The physical measurement location should be evaluated for primary and secondary risks. 
e) Power source must be removed before performing any maintenance.  
f) Exchange of components other than the batteries may compromise ATEX/IECEx certification and shall only be undertaken by 
Thermoprobe or one of its qualified service providers. See also “Authorized Repair” section. 
g) To reduce the risk of fire or explosion, this device must be bonded to the tank according to clause 6.3.2 e), IEC/EN 60079-14 
before and during introduction into the tank and shall remain bonded until complete withdrawal from the tank. 
h) The device must remain bonded to ground/earth using the provided connection whenever a hazardous atmosphere could be present 
as well as during situations where electrostatic charging can occur such as the

 

unwinding/winding of the thermometer cable or filling 

or emptying of the tank. 

 

CAUTION:  In the event that any part of the instrument should become electrostatically charged in a potentially hazardous location, 
follow company policies for testing and clearing the area of any hazardous gases before attempting to bond the instrument to earth 
ground.  If this is not possible allow sufficient time for the instrument to naturally dissipate any charges before attempting to bond to 
earth ground.  Given the atmosphere, this could take several hours.

 

 
 

GUIDANCE NOTE 

Problems with aggressive substances and environments: Be aware of aggressive substances and that extra protection may be needed. 
Caustic soda, highly basic and acidic substances will erode aluminum and copper ground clip and wire. The Sensor-Cable assembly 
has external surfaces of stainless steel and fluoropolymer material.  Exposure to Excessive heat can melt the plastic components of 
the instrument. 
 

 
 
 
 

Summary of Contents for TP7C

Page 1: ...ry risks e Power source must be removed before performing any maintenance f Exchange of components other than the batteries may compromise ATEX IECEx certification and shall only be undertaken by Thermoprobe or one of its qualified service providers See also Authorized Repair section g To reduce the risk of fire or explosion this device must be bonded to the tank according to clause 6 3 2 e IEC EN...

Page 2: ...rdous area to a level below that which will ignite the gases In order to have a fire or explosion fuel oxygen and a source of ignition must be present An intrinsically safe system assumes the fuel and oxygen is present in the atmosphere but the system is designed so the electrical energy or thermal energy of a particular instrument loop can never be great enough to cause ignition TP7C TP9 Hazardou...

Page 3: ...onic AA LR6 Alkaline LR6XWA GP Gold Peak AA LR6 Alkaline GP15A MEASURING PROCEDURES See www thermoprobe net for video on proper use of this instrument Contact your distributor or ThermoProbe Inc for names of training facilities Refer to American Petroleum Institute measurement standard Chapter 7 2 When the temperature has stabilized both Up and Down arrows will be displayed If you wish to log the ...

Page 4: ...te Lo bAtt on the display before resuming normal display functions When the batteries are low the backlight will not operate in order to conserve power while the user completes his operations Replace batteries as soon as possible in a safe location after Lo bAtt is noticed this will ensure backlight operation and avoid possible malfunctioning Do not attempt to calibrate the instrument if the Lo bA...

Page 5: ...ained from ThermoProbe Inc or one of its authorized distributors with part specifications as follows a First follow REPLACING BATTERY instructions a through c to remove batteries b On the circuit board push the terminals clamps down and remove the wires noting the wire lead color code arrangement White positive sensor wire Black negative sensor wire and sheild c Set the cover and circuit board asi...

Page 6: ...ndations Calibration must not be performed in any environment considered to be hazardous Equipment needed Ice Bath or other low temperature bath with certified reference thermometer Warm to hot fluid bath between 20 C approx 68 F or higher up to 90 C approx 194 F with certified reference thermometer see Note Optional high temperature oil bath at about 150 C 300 F and certified reference thermomete...

Page 7: ...ust the device to the actual temperature The On button will decrement the readings and the f Button will increment the reading Once the device temperature matches the actual temperature press the Calibration Button to save the setting The display will scroll SAUE or CAnCEL If the SAUE option is chosen the blinking letter will change to represent the next temperature level A b If the CAnCEL option ...