Precision AC Video Camera 

0700010000
0700010001
0700410105

Содержание 0700010000

Страница 1: ...Precision AC Video Camera 0700010000 0700010001 0700410105...

Страница 2: ......

Страница 3: ...n 17 Operating the Camera with a Light Source 17 Powering the Console On Off 18 Controlling Remote Video Accessories 18 Using the Camera Head Buttons 19 Using the Touchscreen Interface 20 Menu Screen...

Страница 4: ...Expected Service Life 40 Disposal and Recycling Information 40 Diagrams 41 Technical Specifications 43 Electromagnetic Compatibility 45 Symbol Definitions 49...

Страница 5: ...amounts of heat exceeding 41 C 106 F at the scope tip the scope light post the light cable tip and or near the light cable adapter Higher levels of brightness from the light source result in higher l...

Страница 6: ...nce of flammable or explosive gases 10 Always treat the camera system with care The camera system contains sensitive parts that are precisely aligned and may suffer damage if dropped or mistreated 11...

Страница 7: ...d or electric shock 18 Attempt no internal repairs or adjustments not specifically detailed in this operating manual 19 Ensure that readjustments modifications and or repairs are carried out by person...

Страница 8: ...ide a high quality image while remaining lightweight and easy to use A coupler is also available for attaching a scope to the camera head The available models for each part are listed below Consoles 0...

Страница 9: ...copic anterior spinal fusion anterior cruciate ligament reconstruction knee arthroscopy shoulder arthroscopy small joint arthroscopy decompression fixation wedge resection lung biopsy pleural biopsy d...

Страница 10: ...e as well as allowing the selection of surgical specialty settings that optimize camera performance for specific surgical procedures The touchscreen also allows activation of remote outputs which are...

Страница 11: ...tablet control 2 Remote Out 1 Connects to a video accessory remote input 3 Remote Out 2 Connects to a video accessory remote input 4 S Video Out Analog video output 5 DVI Out 1 Digital video output 6...

Страница 12: ...tion 4 AUTOCLAVE 1 2 3 1 Soaking Cap Protects the cable connector during cleaning disinfection and sterilization 2 Cable Connector Connects the camera head to the camera console 3 Camera Cable The cam...

Страница 13: ...he camera It provides a focusing ring to adjust image sharpness Refer to Stryker user guide P26183 for complete coupler instructions 1 2 3 4 1 Rear Adapter Threads onto the camera head 2 Focusing Ring...

Страница 14: ...ing a hospital grade power cord Loss of AC power will cause the camera to shut down and the surgical image to be lost Only connect items to the Precision AC Camera that have been specified for use wit...

Страница 15: ...e labeled Y C The DVI connectors can also output analog SXGA signals through a DVI I to VGA adapter When using any device with unterminated analog video inputs connect a cable from the VIDEO OUT of th...

Страница 16: ...12 Precision AC Wiring Diagram SDC3 Precision AC Camera L9000 Light Source VisionPro 26 LED Surgical Display underside shown 3 4 2 5 1...

Страница 17: ...from the cable connector on the camera head 2 Align the arrow on the cable connector with the arrow above the camera connector port on the front console panel 3 Push in the connector until it locks in...

Страница 18: ...t forms a tight seal and skip to step 3 When attaching or removing the coupler grip only the rear adapter as twisting other parts of the coupler with force may result in mechanical damage Do not overt...

Страница 19: ...there are no rough surfaces sharp edges or protrusions Remove the coupler dust cap if it is present Depress the endobody clamp 1 and insert a scope into the endobody 2 Release the endobody clamp to s...

Страница 20: ...our fibers should be connected to the camera console in the labeled order RGBC The fibers should be connected to the monitor in one of two configurations CBGR reverse order or BGRC R B switched Note T...

Страница 21: ...nitor before adjusting the brightness level of the light source Adjust the brightness level of the light source to the minimum brightness necessary to adequately illuminate the surgical site In additi...

Страница 22: ...refer to the Setting Up the Camera Head section to connect the camera head Controlling Remote Video Accessories The Precision AC Camera can remotely control up to two functions of a video accessory su...

Страница 23: ...ed to a Stryker digital capture console this will capture a photo Press the P button for more than two seconds to select Remote 2 Two beeps will sound When the camera is connected to a Stryker digital...

Страница 24: ...Hysteroscopy Laparoscopy Laser Microscope Standard 2 Press and hold the WB button for two seconds to activate white balance See Performing the White Balance Test below for more detail A checkmark wil...

Страница 25: ...roscopy ENT Skull or Standard surgical specialty is selected When DRE mode is turned on pressing the Up and Down buttons on the camera head for less than two seconds increases and decreases the DRE le...

Страница 26: ...camera head W button or WB on the touchscreen until the monitor displays the message WHITE BALANCE IN PROGRESS 5 Continue pointing the scope at the white surface until the video monitor displays the m...

Страница 27: ...urred during console startup Turn off the console wait 3 seconds and turn it back on Restart Camera Console message Color bar background Camera head temporarily shut down due to overcurrent Turn off t...

Страница 28: ...ting Picture is too dark Increase the camera light level with the camera head Increase the light source output Check the fiberoptic light cable for excessive broken fibers Picture is too bright Decrea...

Страница 29: ...econnect it Make sure the scope is seated correctly in the coupler Variability in color reproduction between different light sources or peripherals Perform the white balance test See the Performing th...

Страница 30: ...completely dry before reassembling them or fogging may result Blurry picture Ensure the coupler or C mount scope is in focus On the Home screen ensure the surgical specialty is not set to FLEXI SCOPE...

Страница 31: ...he AC power source before cleaning Observe the following cautions to avoid damaging the console Do not sterilize the console Do not immerse the console in any liquid Do not allow liquid to drip onto t...

Страница 32: ...ic Detergent Step 3 optional Disinfection Step 4 required Sterilization Warnings This device must be cleaned and sterilized prior to the first use and after every subsequent use Separate the camera he...

Страница 33: ...ra head to cool before connecting it to the console Connecting the camera head while it is hot may result in system error Limitations on Reprocessing Do not cross sterilize the device Using multiple s...

Страница 34: ...with drying Non enzymatic detergent4 Used in cleaning solution to remove surgical debris Automated washer For using the automated cleaning procedure Disinfection Washer disinfector To perform thermal...

Страница 35: ...contaminated surgical instruments and devices Cleaning Manual Cleaning 1 Soak Ensure the soaking cap is installed Refer to the Installing the Soaking Cap section for more detail about installing the...

Страница 36: ...d use an appropriately sized syringe to inject any mated surfaces and crevices with at least 50 mL of the prepared detergent Soak the device for a minimum of 15 minutes 5 Brush Thoroughly brush the ex...

Страница 37: ...ns Fully immerse the device and use an appropriate sized syringe to inject any mated surfaces and crevices with at least 50 mL of the prepared detergent Soak the device for a minimum of 15 minutes 2 B...

Страница 38: ...ent Type Enzymatic Detergent Wash 1 Recirculation Time 2 Minutes Water Temperature 66 C Set Point Detergent Type Non Enzymatic Detergent Rinse 1 Recirculation Time 2 Minutes Water Temperature Hot Tap...

Страница 39: ...reas Disinfection Optional Thermal Disinfection The device must be sterilized after disinfection Failure to sterilize the device before reuse presents an acute infection control risk to the patient Fo...

Страница 40: ...detail about installing the cap 2 If using a sterilization tray optional follow any additional instructions provided with the tray Use only trays that are compatible with steam sterilization 3 Double...

Страница 41: ...use Drying time depends on several variables including altitude humidity type of wrap preconditioning size of chamber mass of load material of load and placement in the chamber Users must verify that...

Страница 42: ...ompatible with Sterrad systems Using an incompatible tray may result in incomplete device sterilization Consult the instructions that came with your sterilization tray to determine which sterilization...

Страница 43: ...t results follow the instructions provided by the drape manufacturer Storage Never store the device in a non ventilated humid environment such as a carrying case This can present an infection control...

Страница 44: ...has an expected service life of 2 years except with repeated use of an automated washer the expected service life is 1 year Note If the camera head is cleaned with the automated cleaning method steam...

Страница 45: ...10 11 2 3 4 5 6 7 Item Material Qty Comments 1 PC Board 1 2 PC Board 1 3 Touchscreen LCD 1 4 Power Supply 1 5 PC Board 1 6 PC Board 1 Located behind LCD 7 PC Board 1 8 PC Board 1 0700010001 only 9 AC...

Страница 46: ...a Head 1 2 Item Material Qty Comments 1 Cable 1 2 Camera Head Enclosure PC Boards 1 The camera head enclosure that contains PC Boards is sealed and cannot be dismantled without special equipment and t...

Страница 47: ...mini DIN Digital Fiber HD HDTV R G B Clk Connector Four Lucent fiber connectors with 1 25 mm ferrules Mounting Endoscope eyepiece used with C mount coupler C mount camera head used with C mount scopes...

Страница 48: ...ry Equipment Precision AC consoles Ingress Protection IPX7 Protected against the effects of temporary immersion in water Precision AC Camera Head Complies with Laser Product Standards Class 1 Laser Pr...

Страница 49: ...he Precision AC Camera Guidance and Manufacturer s Declaration Electromagnetic Emissions The Precision AC Camera is intended for use in the electromagnetic environment specified below The customer or...

Страница 50: ...mon mode 0 5 1kV differential mode 0 5 1 2kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on p...

Страница 51: ...ion distance calculated from the equation applicable to the frequency of the transmitter Recommended Separation Distance d 1 17 P d 1 17 P 80 MHz to 800 MHz d 2 23 P 800 MHz to 2 5 GHz where P is the...

Страница 52: ...mmunications Equipment and the Precision AC Camera The Precision AC Camera is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The user of the camera...

Страница 53: ...for use Caution consult instructions for use Federal law USA restricts this device to use by or on order of a physician Device is shipped non sterile and must be sterilized before use Date of manufac...

Страница 54: ...ce if this appears on the display monitor Equipotentiality Alternating current Fuse rating Device recycling code applicable in China This product contains electrical waste or electronic equipment It m...

Страница 55: ......

Страница 56: ...atents www stryker com patents Stryker Corporation or its divisions or other corporate affiliated entities own use or have applied for the following trademarks or service marks SIDNE Ideal Eyes and th...