background image

Digital Indicator 

SI 400 

 

User Manual 

 
 
 
 
 

 

 
 
 

 
 
 

Ver. 1.0 2014.08.25 

 

Summary of Contents for SI 400

Page 1: ...Digital Indicator SI 400 User Manual Ver 1 0 2014 08 25...

Page 2: ...13 5 1 TEST Weight Calibration Mode 13 5 2 Simulation Weight Calibration 17 5 3 F FUNCTION Setting 21 5 4 Test Mode 29 6 Interface 30 6 1 Serial Interface 31 6 2 ExternalInput 38 6 3 CurrentLoop 39 6...

Page 3: ...al may be changed as the version is upgraded without previous notice Inquiries If you have any kinds of inquiries for this model please contact your local agent or Head Office Head Office SEWHACNM CO...

Page 4: ...Don t install place where high voltage or heavy electric noise condition 4 When you connect with other devices please turn off the power of item 5 Avoid from water damage 6 For the improvement of fun...

Page 5: ...times sec MAX Environment Operating Temperature Range 10 40 14 104 Operation Humidity Range 40 85 RH Non condensing Function Calibration Mode Test Weight Calibration Mode Simulation Calibration Mode D...

Page 6: ...face RS 422 Option2 Serial Interface RS 485 Option3 Serial Interface RS 232 Option4 ETHERNET CARD Option5 Analog Output 0 20mA Option6 Analog Output 0 10V Option7 BCD OUT Option8 BIN IN Option9 SD Mem...

Page 7: ...external input 1 terminal is input ON IN2 When external input 2 terminal is input ON IN3 When external input 3 terminal is input ON IN4 When external input 4 terminal is input ON IN5 When external in...

Page 8: ...o Zero Number 1 Set the TARE Function Number 2 Set theTARE Reset number 3 Set the HOLD Function number 4 When HOLD function is set HOLD Reset number 5 Product No Setting number 6 Display the weighing...

Page 9: ...he current weight during 5 secs Print the Sub total out Print the Grand total out Delete the Sub total weight Delete the Grand total weight Max accumulated weighing count 999 999times Over 999 999time...

Page 10: ...D Load cell EXC EXC SIG SIG SHEILD 5 Analogue Output terminal Terminal 4 20mA Option 0 10V Option 6 Option serial interface terminal Terminal 1 2 3 4 RS 232C GND GND Rx Tx Option RS 422 TxD TxD RxD Rx...

Page 11: ...11 SI 400 DIGITAL INDICATOR 4 INSTALLATION 4 1 External Dimension Cutting Size 4 2 Installation Components SI 400 User Manual Cutting Size...

Page 12: ...temperature difference around Load cell it can cause wrong weight measurement 5 Don t do Welding job or Arc discharge around installation place But there is no choice please disconnect power cable an...

Page 13: ...ress CAL LOCK S W inside When CALIBR displays press key select WCAL and press key 5 1 2 Setting Capacity of weighing Scale If you want to set Max capacity as 1 000kg and the division is 0 1 100g then...

Page 14: ...e cannot be over 1 20 000 If this condition is not fulfilled Err 1 will be displayed and move back to capacity setting mode 5 1 4 Measuring the DEAD Weight of Weighing Scale After DIVI is displayed lo...

Page 15: ...ue will not be calculated Please remove the cause of the force or vibration and process it again 5 1 5 Calculating span value If the count is over input the weight of your Test Weight and press key If...

Page 16: ...20 secs After calculation span value will be displayed on the display Then press key This span value is not a weight value When CALEND is displayed and calibration is completed We recommend to procee...

Page 17: ...s calibration mode uses Load cells max capacity and Max Output Rate mV so the weight adjustment degree might be less than Test weight Calibration The guaranteed resolution of this Simulation Calibrati...

Page 18: ...load cells and each load cell s Max capacity is1 000kg Then total Max Capacity will be 4 000kg 1 000 x 4 and you have to input 4 000 5 2 3 Setting Digit Division value After CAPA displayed Check Max C...

Page 19: ...tage mV After mV V is displayed Check the Rated output value of Load cell Refer to the load cell label or Test Report And Press key to save and move to next step When DEAD displays Press key then indi...

Page 20: ...the rated output will be same as single load cell s Because plural load cells are connected with parallel connection the sum of rated output voltage is same as single load cell s rated output Due to s...

Page 21: ...SETUP is displayed press key 1 Function No or input the function No by number key 0 9 after select Function No area by pressing key 2 Input the Set value by number key 0 9 after select Set value area...

Page 22: ...Grand total 108 Buzzer sound External input detection 00 00 Buzzer sound 01 No Buzzer sound 109 Function Clear key Activation display selection 00 00 F or C key press show on the screen 01 F or C key...

Page 23: ...range 02 Auto At the steady condition 215 Auto Tare reset Time 00 00 Disuse 01 09 Use Unit 1 sec 216 Hold Mode 00 00 Sample Hold 01 Peak Hold 02 Average Hold 217 Hold Delay Time 00 00 Disuse 01 10 Use...

Page 24: ...Paritybit Even 03 Databit 7 Stopbit 1 Paritybit Odd 04 Databit 7 Stopbit 1 Paritybit Even 302 Serial Communication Speed 02 00 2 400bps 01 4 800bps 02 9 600bps 03 14 400bps 04 19 200bps 05 28 800bps 0...

Page 25: ...s 06 38 400bps 07 57 600bps 08 76 800bps 09 1115 200bps 310 Data transmission mode Option Port 02 00 Simplex Stream Mode 01 Duplex Command Mode 02 Print Mode 311 Check Sum detection selection under co...

Page 26: ...Paper Withdraw after Grand total printing out 00 00 10 Unit 1line add 357 After Sub Print Sub data Deleted 00 00 Not Delete 01 Delete 358 Deleting Grand total data after printing out 00 00 Delete 01 N...

Page 27: ...teady weight Steady weight 04 Manual Auto At every steady states Current weight Current weight Steady weight Steady weight 05 Manual Auto At the first steady states Current weight Current weight Stead...

Page 28: ...ion mode Division HF09 Check the division which is set under calibration mode span value HF10 Check the weight of test weight which is used for your last calibration Zero Analog HF11 Check the analog...

Page 29: ...TEST mode please remove operating devices Test Mode Analog Value key 4times Analog Deviation Check key 4times Key Input Check key 4times Display Check key 4times External Input key 4times Analog Outpu...

Page 30: ...30 SI 400 DIGITAL INDICATOR 6 INTERFACE 6 1 Serial Interface 6 1 1 Standard serial interface terminal 1 RS 232 6 1 2 Option serial interface terminal 1 RS 232...

Page 31: ...31 SI 400 DIGITAL INDICATOR 2 RS 422 3 RS 485 Maximum 32units of Indicators connectable Maximum 32units of Indicators connectable...

Page 32: ...STEADY GS when setting TARE US UNSTEADY 2 Data Format2 ID Number Data Transference Refer F function 101 305 01 22byte Header1 Header2 OL OVER LOAD NT NET WEIGHT Tare is not set ST STEADY GS when setti...

Page 33: ...eader1 Header2 OL OVER LOAD NT NET WEIGHT Tare is not set ST STEADY GS when setting TARE US UNSTEADY LAMP DISPLAY Bit 0 Bit 1 Bit 2 Bit 3 Bit 4 Bit 5 Bit 6 Bit 7 ZERO 1 TARE Gross Weight Print HOLD 1...

Page 34: ...h Data length Error 6 1 3 Read Command Subject Command Length of transmission data 305 312 00 01 03 04 305 312 02 Current Weight STX ID RCWT ETX 22 byte 22 byte Current data STX ID RCWD ETX 44 byte 48...

Page 35: ...8 byte Hold STX ID WHOL ETX 8 byte Hold Reset STX ID WHRS ETX 8 byte Print STX ID WPRT ETX 8 byte Sub total Print STX ID WSPR ETX 8 byte Grand total Print STX ID WGPR ETX 8 byte Sub total Delete STX I...

Page 36: ...ers 04 0x04 Read Input Registers 06 0x06 Write Single Registers 16 0x10 Write Multiple Registers CRC Check Method is CRC 16 Add ress Leng th Fea ture Description 1 2 RO Capacity 3 2 RO Real weight 5 2...

Page 37: ...ndicated by 32bit 0 1 2 3 4 5 4 7 Steady Zero Tare Hold TxD RxD Printer Dead Battery 8 9 10 11 12 13 14 15 IN1 IN2 IN3 IN4 IN5 IN6 3 Error Register Address 21 Length 2 Error data is indicated by 32bit...

Page 38: ...6 2 External Input Each External Input funtion setting is F Funtion 233 238 possible 6 2 1 External Input configuration 6 2 2 External Input connector connection TERMINAL I1 I2 I3 I4 I5 I6 IC INPUT IN...

Page 39: ...loop is suitable for middle distance transmission because stronger than RS 232C against electric noise About 100M Maximum communication speed is 9 600 6 4 1 Current loop circuit composition 6 4 2 Con...

Page 40: ...gue output value is adjusted with DIGITAL MULTI METER 3 How to adjust analog output value Step1 Connect Digital multi meter to the Indicator A out terminal Step2 Enter F function HF14 Minimum Analog O...

Page 41: ...t analog output value is 0V weight zero 10V Full capacity 2 The analogue output value is adjusted with DIGITAL MULTI METER 3 How to adjust analog output value Step1 Connect Digital multi meter to the...

Page 42: ...10V 2 HF13 Analog output setting should be changed also 6 7 Print Interface It can be connected with all kinds of Serial interface printer but the printing format is already programmed and fixed with...

Page 43: ...IN IN card Changing Product number 6 8 1 BIN IN card circuit composition 6 8 2 BIN IN card connection PIN No 1 14 2 15 3 16 4 17 5 Role IN1 IN2 IN3 IN4 IN5 IN6 IN7 IN8 IN COM Function 404 00 1 2 4 8 1...

Page 44: ...19 4X10000 28 2X1 2 8X10000 11 4X1 20 1X100000 29 8X1 3 2X100000 12 1X10 21 4X100000 30 2X10 4 8X100000 13 4X10 22 COM 32 8X10 5 Net weight HIGH 31 1X100 23 Total weight LOW 31 2X100 6 GND 1 14 4X100...

Page 45: ...11 SD memory card Weighing data will be saved to SD memory card depends on your function 103 6 11 1 Saving format File name YYMMDD CSV Ex 140728 CSV DATE TIME ID PART SERIAL WEIGHT UNIT 2014 07 18 12...

Page 46: ...nect SD card to PC and select FORMAT from PC system folder Select FAT32 from file system 6 12 Option card combination Maximum 2EA of option card can be instlled Below combination is available SERIAL 2...

Page 47: ...solate Resistance is more than 100M Weight Value is increased regular rate but not return to Zero 1 Load cell Error 2 Load cell connection Error 1 Check Load cell connection 2 Measure Load cell Resist...

Page 48: ...ty Eerr06 Amp Gain is too big Check standard weight s weight with set value If there is difference between set value and real weight please re input the value set value is too small Eerr07 Amp Gain is...

Page 49: ...r s A D with the other indicator 4 Check the power and connection of terminal UNpa55 1 Power is ON when some materials are on weighing part Under Normal Mode if there are more than 20 loading of Max c...

Page 50: ...m s permission Any kinds of Mal function Defection or External damage caused by operator Any kinds of Mal function Defection caused by using spare part from Non Authorized Distributor or Agent Any kin...

Reviews: