background image

1

Installation Guide

1-800-257-3872 

Toll Free

1-978-264-0292 

Fax

www.setra.com 

Web Site

Model GCT225

Ultra-High Purity

Pressure Transducer

ISO

Certified

9001:
     2000

Summary of Contents for GCT225

Page 1: ...1 Installation Guide 1 800 257 3872 Toll Free 1 978 264 0292 Fax www setra com Web Site Model GCT225 Ultra High Purity Pressure Transducer ISO Certified 9001 2000 ...

Page 2: ...the Surge Immunity requirements the following conditions must be followed during installation 1 Shielded cable must be used and the shield must be tied to earth ground not power supply ground on at least one end of the cable shield drain wire The shield must be maintained all the way from sensor to the power supply 2 If unshielded cable is used an earth grounded metal conduit fitting can be used t...

Page 3: ... Mounting Model GCT225 transducer can be installed with a male or female 4 Face Seal Fitting 1 4 NPT male or female fitting or a 1 4 tube stub When installing units with Face Seal Fittings a Align piping system to transducer connections b Hand tighten nuts c Torque nuts by placing wrenches on nuts only or on nut and wrench flats of transducer body for units fixed male face seals Never hold the uni...

Page 4: ...s Note 3 ATEX 94 9 EC Approved Special hazardous area instructions for GCT225 transducers with the N1 suffix with ATEX approval The GCT225 transducer with the N1 suffix is designed for use in hazardous areas indoor applications with ATEX rating Non incendive II 3G Eex nL IIC T4 X 30 C Ta 80 C Note Units ordered with the Mini Din Shell are not ATEX approved for use in hazardous locations The device...

Page 5: ...1 3 904 3 704 1 to 3 4 1 336 1 136 2 473 2 273 1 to 7 0 825 0 625 1 450 1 250 1 to 17 2 494 0 278 0 756 0 556 1 to 35 0 339 0 139 0 478 0 278 1 to 70 0 270 0 070 0 341 0 141 1 to 200 0 225 0 025 0 250 0 050 4 1 Voltage Output Zero Adjustment While monitoring the voltage between the positive output OUT and common COM and with the pressure port open to atmosphere or with zero pressure applied the ze...

Page 6: ...barometric pressure 4 4 Current Output Span Adjustment Span or full scale output adjustments should only be performed by using an accurate pressure standard electronic manometer digital pressure gauge etc with at least comparable accuracy to the GCT225 transducer 0 25 FS With full range pressure applied to the pressure port the span may be adjusted by turning the span potentiometer screw The span ...

Page 7: ...l GCT225 to be free from defects in materials and workmanship subject to the following terms and conditions Without charge SETRA will repair or replace products found to be defective in materials or workmanship within the warranty period provided that a the product has not been subjected to abuse neglect accident incorrect wiring not our own improper installation or servicing or use in violation o...

Page 8: ... FOR RETURN DESCRIPTION OF PROBLEM NOTES COMMENTS OR SPECIAL HANDLING Any product that has been used with Hazardous Materials must be 100 purged and should be accompanied by a MSDS sheet bagged sealed and tagged accordingly Setra will not accept delivery of any product exposed to chemicals radioactive agents or biological process without evidence of decontamination or laboratory analysis and proof...

Reviews: